Repair 3D/4D GE RAB4-8-D Crystal/Crystal Replacement/Broken Probes

 

 

Crystal damage -Weak, dead, or damaged crystals

 

Very often the cause of crystal damage in ultrasound probes is the dropping of probe on the floor or hitting of probe with gel bottle or other object. Once a probe is dropped its piezoelectric crystals may stop sending or receiving signals, resulting in scans with weak or dead elements. This leads to black lines in an image, these lines are called dropouts. Weak elements cause the faint little lines that won’t obscure much. These are the first signs that the elements are getting progressively damaged and need to be replaced. Instead of faint lines if the lines are thick and pronounced, the elements are dead. If there are thick lines in the scanned image, important things can be missed out during diagnosis. You may be harming the patient due to misdiagnosis.

 

Solution

 

Repair/Replace the Crystal

 

 

 

 

 

Other hot GE models we can repair:

 

GEC1-5-D
GEC1-5-RS
GE4C
GE4C-RS
GE4C-A
GE4C-D
GE4C-SC
GEE8C/E8C-RS
GEE8CS
GERIC5-9-D
GE3CRF-D
GEC1-6-D
GERAB4-8-RS
GERAB2-5-RS
GE6S-RS
GE12L-RS
GE3SC-RS
GEL6-12-RS
GE11L
GE11L-D
GE12L-RS
GE8C
GE3S
GE3S-RS
 

 

 

 

Common ultrasonic Probe Damage(Convex, Linear, Sector, Endocavity probes)

 

Common ultrasonic probe damageSolutions
Lens damage, wear, holes, swelling, delaminationLens replacement
Strain relief damage, separationStrain replacement
Nosepiece and probe separation and cracks

Cosmetic repairs

Cable cutsCable patches, possible cable replacement
Connector housing electrical damageMajor and minor electrical repairs, pin module replacement

 

 

 
Probe Troubleshooting Summary
 
1.Check whether multiple probes are the same phenomenon, sometimes it is necessary to plug again
2.Lens and crystal failure are sometimes similar, both are shadowing on the image
3.If there is a single probe problem, replace the socket to have a check.
4.The high frequency probe (small parts probe) is sensitive. If only this probe doesn’t work, may be caused by machine failure.

 

 

Easy Ultrasound Transducer Maintenance Tips & Damage Prevention

 

Shipping

If possible, use the box/case/foam from the manufacturer to protect the ultrasound probe during the shipping process.

 

Storage

Do not store in clamp control housing, which puts unneeded stress on the cord and strain relief. Instead, keep unused ultrasound probes in a cool, dry place — ideally in a case or box.

 

Use

Check the ultrasound transducer before each use to make sure all elements are properly functioning.

 

Cleaning

Before submerging the probe, make sure the strain relief and cord are undamaged. Don’t leave the ultrasound probe in contact with cleaning solutions for extended periods.



Hot Products

Repair 4D Samsung V5-9 Strain Relief Among the more common issues is a crack or separation in the ...
Repair 4D Siemens 7CF2 Acuson Lens Cap/Fluid And Oil Leaks/Leaking Oil Fluid and oil leaks Solution ...
Repair 3D/4D GE RAB4-8-D Crystal/Crystal Replacement/Broken Probes Crystal damage -Weak, dead, or ...
Repair Toshiba PLT-805AT Pin/Repair Connector/Broken Pins/Locking Shafts​ Probe connectors are the ...
Repair Toshiba PVU-375BT Lens/Lens Replacement/Repair Service The lens of failures Reasons An ...
Repair GE 3SC-RS Crystal/Crystal Replacement/Damaged Crystals Crystal failure reason Most internal ...