GE 3SC RS Ultrasound Transducers Repair Damaged Crystals

Country/Region china
City & Province guangzhou guangdong
Categories Solar Chargers
InquireNow

Product Details

 

Repair GE 3SC-RS Crystal/Crystal Replacement/Damaged Crystals

 

 

Crystal failure reason


Most internal ultrasound probe uses a piezoelectric crystal. After long-term use, it can be causing the crystal of natural aging, and external force such as the use of non-standard or fall off, touch hit will make probe crystal damage, after crystal damaged, B ultrasonic detector detected in the image will appear signal attenuation, the image has dark, black bars, interference, defects such as blind area, serious the images will be "blind", which will cause doctors can not diagnose, affecting the quality of diagnosis.

 

Solution

 

This kind of failure is caused by ultrasound probe wafer short circuit. The short circuit of the chip can be judged by measuring the power supply voltage of the chip, if a chip or chip set of power supply voltage is lowerthan others, can judge the chip or chip set has a short circuit. Replace the corresponding chip or chip group, makes the probe to restore the previous feature, the interference disappears.

 

 

 

Other hot GE models we can repair:

 

GEC1-5-D
GEC1-5-RS
GE4C
GE4C-RS
GE4C-A
GE4C-D
GE4C-SC
GEE8C/E8C-RS
GEE8CS
GERIC5-9-D
GE3CRF-D
GEC1-6-D
GERAB4-8-RS
GERAB2-5-RS
GE6S-RS
GE12L-RS
GE3SC-RS
GEL6-12-RS
GE11L
GE11L-D
GE12L-RS
GE8C
GE3S
GE3S-RS
 

 

 

 

Common ultrasonic Probe Damage(Convex, Linear, Sector, Endocavity probes)

 

Common ultrasonic probe damageSolutions
Lens damage, wear, holes, swelling, delaminationLens replacement
Strain relief damage, separationStrain replacement
Nosepiece and probe separation and cracks

Cosmetic repairs

Cable cutsCable patches, possible cable replacement
Connector housing electrical damageMajor and minor electrical repairs, pin module replacement

 

 

 
Probe Troubleshooting Summary
 
1.Check whether multiple probes are the same phenomenon, sometimes it is necessary to plug again
2.Lens and crystal failure are sometimes similar, both are shadowing on the image
3.If there is a single probe problem, replace the socket to have a check.
4.The high frequency probe (small parts probe) is sensitive. If only this probe doesn’t work, may be caused by machine failure.

 

 

Daily and Long-Term Storage of ultrasound transducer

 

1. Always store transducers in the transducer holders on the side of your system or on a securely mounted wall rack when you are not using them.

2. Ensure the transducer holders are clean before storing transducers

3. When storing transducers, use the cable-management clips, if available, to secure the transducer cable

4. Avoid storing transducers in areas of temperature extremes or in direct sunlight.

5. Store transducers separately from other instruments to avoid inadvertent transducer damage.

6. Before storing transducers, make sure they are thoroughly dry.

 



Hot Products

Repair Toshiba PLT-805AT Pin/Repair Connector/Broken Pins/Locking Shafts​ Probe connectors are the ...
Repair Toshiba PVU-375BT Lens/Lens Replacement/Repair Service The lens of failures Reasons An ...
Repair GE 3SC-RS Crystal/Crystal Replacement/Damaged Crystals Crystal failure reason Most internal ...
Repair 4D V6-2 IU22/IE33 Cable/Connecting Cable Leakage/Damaged Cable Cable leakage failure reason ...
GE 3D 4D Probe Oil Leakage Repair Service Repair Oil-Leaking Rab4-8-D Ultrasonic Sensor​ We provide ...
Ultrasound Repair Service For Probes Aloka UST-5548 Lens Replacement Replace Lens for probes We ...