Fingertip pulse oximeters - easy to use ,small in volume,low power consumption,convenient operation

 

Features
➢Power consumption of the product is low and the two originally equipped AAA batteries can be operated continuously for 20 hours.

➢The product will automatically be powered off when no signal is in the product within 5 seconds.

➢Operation of the product is simple and convenient.
➢The product is small in volume, light in weight (total weight is about 29g) and convenient in carrying.


 

Small and portable the size of a matchbox, easy to carry, anytime, anywhere, just test

High precision one-key measurement, just press the key to start the measurement.

Four directions display the reading is more comfortable

Automatically detect shutdown save energy

Soft finger pad One-touch finger moulding, good fit, soft and comfortable

 

 

 

Major Applications and Scope of Application
 

The pulse oxygen saturation is the percentage of HbO2 in the total Hb in the blood, so-called the O2 concentration in the blood. It is an important bio-parameter for the respiration .For the purpose of measuring the SpO2 more easily and accurately, our company developed the Pulse Oximeter. At the same time, the device can measure the pulse rate simultaneously.
The Pulse Oximeter features in small volume, low power consumption, convenient operation and being portable. It is only necessary for patient yp put one of his fingers into a fingertip photoelectric sensor for diagnosis, and a display screen will directly show measured value of Hemoglobin Saturation.

 

The Pulse Oximeter can be used to measure human Hemoglobin Saturation and pulse rate through finger, and indicate the pulse intensity by the bar-display. The product is suitable for use in family, hospital (Ordinary sickroom), Oxygen Bar, social medical organizations and also the measure of saturation oxygen and pulse rate.

 

 

Function Specification

Display InformationDisplay Mode
The Pulse Oxygen Saturation(SpO2)LED
Pulse Rate(PR)LED
Pulse Intensity (bar -graph)LED
Pulse waveLED
SpO2 Parameter Specification
Measuring range0%~-100%, (the resolution is 1%).
Accuracy70%~100%±2%, Below 70% unspecified.
Optical Sensor

Red light (wavelength is 660nm)

Infrared (wavelength is 880nm)

Pulse Parameter Specification
Measuring range30bpm~250bpm (the resolution is 1 bpm)
Accuracy±2bpm or±2% select larger
Pulse Intensity
RangeContinuous bar-graph display, the higher display indicate the strong pulse
Battery Requirement
1.5V (AAA size) alkaline batteries x 2 or rechargeable battery
Battery Useful Life
Two batteries can work continually for 20 hours
Dimensions and Weight 
Dimensions57(L) x 31(W) x 32(H) mm .
WeightAbout 50g (with the batteries)

 

Troubleshooting

TroublePossible ReasonSolution

The SpO2 and Pulse Rate can not be displayed normally

 

1. The finger is not properly positioned.

2. The patient's SpO2 is too low to be detected.

 

1. Place the finger properly and try again.

2. Try again; Go to a hospital for a diagnosis if you are sure the device works all right.

The SpO2 and Pulse

Rate are not

displayed stably

 

1. The finger is not placed inside deep enough.

2. The finger is shaking or the patient is moving.

1. Place the finger properly and try again.

2. Let the patient keep calm

 

The device can not

be turned on

 

1. The batteries are drained or almost drained.

2. The batteries are not inserted properly.

3. The malfunction of the device.

1. Change batteries.

2. Reinstall batteries.

3. Please contact the local service center.

 

The display is off

suddenly

 

1. The device will power off

automatically when it gets no signal within 5 seconds.

2. The batteries are almost drained.

1. Normal.

2. Change batteries.

 

 

Operation Instructions

1.Install a pair of AAA batteries into the battery compartment.
2.Plug a finger into a rubber hole of the Oximter.
3.Press button on the front panel to start.
4.Do not shake your finger or move your body when the Oximeter is working
5.Press the button if you want to change display direction
6.Read the relevant value from the display screen
7.The device will automatically power off in 8 seconds if there is no finger signal.

 

 

 

 

 

 

Hot Products

Fingertip pulse oximeters - easy to use ,small in volume,low power consumption,convenient operation ...
CE CMS50D finger pulse oximeter - easy to use ,small in volume,low power consumption,convenient ...
Fingertip pulse oximeters - easy to use ,small in volume,low power consumption,convenient operation ...
Product Name Medical Blood Oxygen Saturation Pulse Monitoring Home Heartbeat Heart Rate Detector ...
Product Name Portable LED Screen ABS Plastic Household Blood Oxygen Saturation Monitor Fingertip ...
Product Name Large OLED Display Portable Finger Pulse Oximeter Blood Oxygen Saturation Monitor SPO2 ...