Samsung V5-9 Ultrasound Probe Repair Oil and Sealing Problem Oil leakage

 

Probe Troubleshooting Summary

 

1. Check whether multiple probes are the same phenomenon, sometimes it is necessary to plug again

2. Lens and crystal failure are sometimes similar, both are shadowing on the image

3. If there is a single probe problem, replace the socket to have a check.

4. The high frequency probe (small parts probe) is sensitive. If only this probe doesn’t work, may be caused by machine failure.

 

 


 

 


 

Basic Ultrasound Transducer Types
1Linear Transducers
2convex or curved array Transducers
3Sector Transducers
4Endocavity Transducers
5Tee Transducers
63d/4d Transducers
7Phased Array Transducers

 

 

 

Knowledge point

 

 

One may be facing several types of problems while using the ultrasound machine probes. They may be one or more of the following:

 

1. Poor quality image: Poor image may be due to

    *Cracked or air bubbles on lens

    *Connector issue

    *Crystal damage

    *Tears on the sheath

 

2. Physical damage or broken housing: Physical damage may be due improper handling, dropping of probe, poor storage etc.

 

3. Noise: Noise may be due to broken cable, damaged / broken pins in connector or electrical interference.

    *Breakage of probe connector pins may be caused by improper insertion or connection.

    *Damage to cable occurs due to cables being ridden by some wheels or closure of drawers without proper placement of the probe inside, etc.

 

4. No image: If there is image being received, it could be a result of probe cable – connector breakage, probe compatibility issue or software issue.

Hot Products

Repair GE 3S-RS​ Pin/Repair Connector/Broken Pins/Locking Shafts​ Probe connectors are the piece of ...
Samsung V5-9 Ultrasound Probe Repair Oil and Sealing Problem Oil leakage Probe Troubleshooting ...
Repair GE 4C-RS Ultrasound Probe Repair Crystal Replacement Imagine Problem Crystal damage -Weak, ...
Damaged Cable Repair Strain Relief Replacement Ultrasound Probe GE 4C-D Cable leakage failure reason ...
GE 3S-RS Replace Lens 3D 4D Probe Repair Ultrasound Sensor Repair time: 2-5 days Warranty: 60 days ...
Medical Ultrasound System GE Voluson E6/Personal Sonogram Machine/Picture Ultrasound 1. Condition: ...