CJ-RoII Series Dialysis Water Systems Flow Can Be Chose From 1000l/Hour ~ 3000/Hour Support 20~100 beds for dialysis
ROII Series Reverse Osmosis Water Filtration For Hemodialysis Peritoneal Dialysis Center RO System
A reliable hemodialysis water center equipment is not only bio
safe, but also portable and comfortable to use.
Hemodialysis centers require a specialized water treatment system
to ensure that the water used for dialysis is free from
contaminants and meets the appropriate quality standards. This is
because even small amounts of impurities in the water can pose a
significant risk to patients undergoing dialysis.
The water treatment system used in hemodialysis centers typically
includes multiple stages of filtration and purification to remove
impurities such as bacteria, viruses, endotoxins, and chemical
contaminants from the water. The system may also incorporate
reverse osmosis(RO) to remove dissolved salts and other minerals
from the water, and ultraviolet (UV)disinfection to kill any
remaining microorganisms.
The resulting water must meet specific quality standards set by
organizations such as the Association for the Advancement of
Medical Instrumentation (AAMli) and the American Society for
Artificial Internal Organs(ASAIO). These standards specify the
acceptable levels of various contaminants in the water used for
dialysis, including total dissolved solids, endotoxins, bacteria,
and other impurities.
Data detection:
The dialysis water and dialysate should be tested in detail
regularly. The microbial requirements in the dialysis water and
dialysate are mainly that the total number of bacteria is
<100CFU/ml;
The intervention level should be based on knowledge of system
microbial dynamics and should be one-half of the maximum allowable
level. At the same time, endotoxins in dialysis water are
<0.25EU/ml, and endotoxins in dialysate are <0.5EU/ml.
Corresponding intervention levels must be established, and the
maximum allowable levels are also one-half of each;
When it is found that bacteria and endotoxin indicators exceed the
standard, be sure to use disinfectant immediately to carry out
systematic disinfection with professional procedures. The removal
rate of endotoxins and bacteria must be ≥99%;
The removal rate of dissolved salts must be ≥97%, while the water
recovery rate must be >70%, and the system’s emptying rate must
be ≥99%, with no dead space.
Application | Techniques |
---|---|
Hemodialysis (HD) | This is the most common type of dialysis and involves removing waste and excess fluid from the blood by circulating it through a dialysis machine, which filters the blood and returns it to the body. Hemodialysis water is used to prepare the dialysate solution, which is the fluid that circulates through the dialysis machine and removes waste products from the blood. |
Hemodiafiltration (HDF) | This is a type of dialysis that combines hemodialysis with ultrafiltration, which is the process of removing excess fluid from the blood. In HDF, hemodialysis water is used to prepare both the dialysate solution and the replacement fluid, which is the fluid that is infused into the blood to replace the removed fluid. |
Features | Description |
---|---|
Low demand for manual intervention | Fully automatic operation, and Real-time display of water quality status, bringing reliable operational efficiency to the hospital. |
Reliable water quality | Fully reliable and compliant with AAMI and ASAIO dialysis water quality requirements, we have a medical device production license. |
Low maintenance cost | Using customized reverse osmosis components with ultra-long lifespan and customized filtering consumables, long service life, and low usage cost. |
Excellent production materials | The key parts are sealed with nitrogen gas in the 304 pipelines, ensuring that there are no dead spaces in the pipeline and preventing bacterial growth during use. |
Excellent mobility | The bottom is equipped with a highly load-bearing wheel system, which ensures that the equipment can be easily moved to different positions. |
Excellent electrical safety performance | Use internationally famous brand electrical control components and components to ensure that the equipment can operate continuously for a long time. |
CJ-ROII series performance and configuration indicators | |||||||
MAKE-UP FLOW RATE (@25°C) | WATER QUALITY | Tap water requirement | INORGANICS REJECTION RATE | Total bacteria | Endotoxin | Disinfection method | Power and voltage-related information |
1000l/hour~3000l/hour | AAMI / ISO dialysis water standard | ≤400μs/cm 0.2-0.5Mpa | ≥98.7% | <50CFU/ml | <0.25EU/ml | Automatic chemical disinfection | AC380/6.0~12.5 |