Bring all reagents to room temperature before testing. The test
should be conducted at room temperature.
I. Sample extraction (see Figure 1)
1. Add 400 μL (about 10 drops) of buffer solution vertically to the
sample extraction tube, insert the sample into the solution in the
sample extraction tube, rotate it about 10 times through the inner
wall, and dissolve it in the solution as much as possible.
2. Extrude the tip of the sample along the inner wall of the
extraction tube, leaving as much liquid as possible in the tube and
taking out the sample.
3. Apply drip.
II. Test Procedure (See Figure 2)
1. Remove the test card from the sticker bag.
2. Add 2 drops (about 80 μL) of the processed sample extract to the
sample hole of the test card and activate the timer.
3. Place the test card at room temperature for 15 minutes before
reading the results. It could not be read after 20 minutes.
Diagram of test card result judgment:
(1) Invalid result: The test is invalid because there is no
reaction line on the quality control line (C line). This experiment
is redone
② Negative result: red ribbon, quality control line color (C line)
③ Positive result: Two red bands, the inspection line (t line) and
the quality control line (C line), are displayed in color.