GE Logiq P5 System Reports Error

 

 

Failure phenomenon

LOGIQ P5 has no response at start up.

 

 

 

 

Failure analysis

Switch ON the POWER, turn ON the main switch at the back of the machine, press the "POWER ON" button, the monitor and keyboard lights are off, the cooling fan is not turned, and the machine has no response.POWER ON cannot be activated when the UPS POWER supply and the main switch of the machine are restarted. The output voltage of UPS measured on fieldis 222V,which indicating that the POWER supply is normal.After turning on the main switch of the machine, the waiting light next to the transformer can be seen from the transformer cooling hole below the machine, showing that AC 220V has entered the machine.Press POWER ON at this time, still no response, suspected the machine internal POWER supply (APS) may be faulty.

 

Troubleshooting

Open the side cover of the machine, observe the APS panel indicator, only the LPS indicator flashes, and the other lights do not works.According to the working principle diagram of APS, LPS is the preparation indicator light of APS, which should be in the long and bright state.Flashing indicates that the machine is in a protected state.Test the voltage at each connection from the motherboard of the machine, and the voltage deviates from the normal value, so that the lock fault is caused by APS's failure to start vibration normally.Replace the APS, and the machine is in normal use.

 

Ultrasound parts that we can repair:

 

 

Repair List of Featured Ultrasonic Board & Assy
BrandSystemDescriptionPart Number
GELogiq P5CPU mainboard5168431
GELogiq P6CPU mainboard 
GELogiq E9/vivid E9GFI25161631
GEVoluson E6/Voluson E8RSRKTI301394-2/KTI196357
GEVoluson E6/Voluson E8RFI/ RFI21b BoardKTI300614/KTI302197-6
GEVoluson S6/S8/P8BF645396937-2
GEVoluson S6/S8/P8RFS board5364098-2/5364098-3
GEVoluson S6/S10/P8BF192 board5357234
AlokaAlpha 5mainboardEP493700DD
AlokaAlpha 6RX Beam FormerEP555501AA
AlokaAlpha 6CW boardEP555700
AlokaAlpha 6CPU mainboardEP537000/EP563000
AlokaAlpha 6TX/RXEP555300
AlokaAlpha 6/Alpha 7DOMEP550000AA
AlokaAlpha 10RX Beamformer boardEP531800CE
AlokaAlpha 10wave generator BoardEP495000HH/EP495000HJ/EP495000FF
HitachiF37TX/RSEP557400/EP557300
HitachiF37mainboardEP575700BC/EP560800
AlokaF75USP boardEP556600
AlokaF75Cell BoardEP558800
HitachiHI VISION AviusRDBF board7513907A/7513612
HitachiArietta 60CELLEU-9160C
HitachiArietta 70TXEP572300AA
HitachiArietta 70CELL 
HitachiNoblusCELL 
HitachiNoblusCONT-B 
HitachiAscendusCELL7352830A
SiemensX150/G40BE10131804
SiemensX300BE 2.0/BE 3.0/BE 7.510131990
SiemensX700BE10136465
SiemensX700RC10136467
SiemensX700TI11014292
SiemensAntaresBE7472421
SiemensAntaresCB7288504
SiemensS2000RM30010852163
SiemensS2000 high version/S3000RM30110854749
iU22/iE33EMB excelsior motherboard453561419431/453561382231/453561419431
iU22/iE33UMB Unified Motherboard453561254492/453561360251
HD15ACB acquisition control board453561197145/453561197144
HD15AIM board4.53561E+11
CX50control panel4.53562E+11
CX50mainboard453561368033/453561622163/453561495832
Affiniti 70ACQ acquisition control board 
EPIQ 5/EPIQ 7ACQ Acquisition Module board453561734844/453561704246
BiosoundMylab SeriesCPU mainboard9501361000
ToshibaAplio 300/400/500mainboardPM30-38696
ToshibaAplio 500RXywm2056*a
ToshibaXario 200TXPM30-32732-1
ToshibaXario 200RXPM30-32733-1
Remarks
1.The repair cycle is within 3-5 days.
2.Warranty after fixed:60 days.
3. Kindly send us the estimated main faults (better to have image) for our engineer to prejudge.
4.Package price is available if repair 3 pcs or more at one time.
5.Extra discount for VIP partners.
6.You could have different part No.under same description.

 

Hot Products

GE Logiq P5 System Reports Error Failure phenomenon LOGIQ P5 has no response at start up. ecer(...
GE Voluson S8 System Reports Error The parts with high failure rate&available in stock of this ...
GE Logiq P6 System Reports Error Failure phenomenon All probes of GE LOGIQ P6 machine have ...
GE Logiq S8 System Reports Error Failure phenomenon 2D image and blood flow of the machine is ...
GE Voluson S6 System Reports Error The parts with high failure rate&available in stock of this ...
GE Voluson P8 Ultrasound Sonography Machine Repair GE Voluson P8 System Reports Error The parts with ...