GE Voluson S6 System Reports Error

 

The parts with high failure rate&available in stock of this machine:

 

1.DBM64

2.DBM128

3.BF64

4.BF128 Transformer

5.CPS Power Supply

6.RFS Mainboard

 

 


 

 

Common faults of GE ultrasound machine

 

SystemFaultsSolutions
LOGIQ 500No image when scanning far fieldCheck power supply, reinsert the circuit board. Suggest to do metal earthing.
LOGIQ 500PROArtifacts appear in color flow imaging (betweentimes)Replace ACFP3 board
LOGIQ 200ENo image on displayer, whereas normal image in workstationFix displayer
LOGIQ 200ECurved lines interpolated into sector-shaped image, from near field to far fieldReplace DSC board, and make sure the system earthing
LOGIQ 200No sound for the probe, requires multiple startupDetect low voltage and external power supply (MST)
LOGIQ BOOKLithium battery- short lifeReplace lithium battery, and upgrade the system to 2349040-23 KB824146
LOGIQ BOOKCrash
Reinstall the system
LOGIQ 400MDcardiac probe shows clutter filte, interference in far fieldRepair cardiac probe S317, then test the voltage
LOGIQ 400MDNoisy in probe, no waves in scan areaConnecting board failure, try to connectwith TRDR and HBFR
LOGIQ 400CLAutomatic interruptionHigh voltage failure, need repair
LOGIQ 180Screen wobbleDisplayer failure, need stabilized voltage supply
LOGIQ 500MDNo image after boot (betweentimes)Detect five low voltage value
LOGIQ 200PRONo imageDetect the channel board
LOGIQ 9No scan area, no ultrasound imageMSTE and high voltage power supply failure, need repair

 

Hot Products

GE Logiq P5 System Reports Error Failure phenomenon LOGIQ P5 has no response at start up. ecer(...
GE Voluson S8 System Reports Error The parts with high failure rate&available in stock of this ...
GE Logiq P6 System Reports Error Failure phenomenon All probes of GE LOGIQ P6 machine have ...
GE Logiq S8 System Reports Error Failure phenomenon 2D image and blood flow of the machine is ...
GE Voluson S6 System Reports Error The parts with high failure rate&available in stock of this ...
GE Voluson P8 Ultrasound Sonography Machine Repair GE Voluson P8 System Reports Error The parts with ...