Jd-8002 Temperature And Humidity Box Heating Time ≤5-8 °c/min High
Temperature Oven
Manufacturing Overview:
The high temperature safety oven is designed and manufactured
according to the national standard (JB/T 5520-91 drying box
technical conditions), and is widely used in petroleum, chemical,
composite materials, electronics, medical pharmacy, scientific
research laboratories and other related production and research
fields. Used for drying, baking, heat treatment, etc. of articles.
This device has the following basic features:
Adopt centrifugal multi-wing fan circulation system, unique air
duct structure, stable temperature control and uniform temperature
field. The manufacturing process specification is beautiful in
appearance, convenient in operation and maintenance, and the buyer
only needs to provide power and meet the required working
environment.
The matching of components and assembly is good. The main
functional components are all original parts with international
advanced level.
The temperature adopts intelligent digital display temperature
control system. The control system adopts digital display and has
PID control function. It can realize automatic operation of test
program, and can automatically combine subsystems such as heating
and fan to ensure high quality control of the whole temperature
system. .
Designed with a safe venting hole, located at the rear of the
box, can automatically discharge pressure when the gas expands in
the working chamber, to avoid damage to the door when the gas
inside the studio expands and further damage the staff.
The control system is equipped with a stand-alone digital
over-temperature power-off protection system, which can be used
with confidence when unattended.
Configure high temperature resistant graphite sealing strips,
humanized environmental awareness, and protect the health of users.
First, the main technical indicators:
1.1 | Temperature range: | R.T.+10°C~500°C; |
1.2 | Temperature fluctuation: | ≤ ± 1 ° C |
1.3 | Temperature uniformity | ≤ ± 2.5% ° C |
1.4 | Heating time | ≤5-8 °C/min |
1.5 | Power | 5KW |
1.6 | Power supply voltage | 220V 50HZ |
Second, the room structure characteristics:
2.1 Structure: The horizontal integrated structure is used, and the
control operating system is located on the upper side of the
studio.
2.2 Box outer wall material: It is made of 1.5mm cold-rolled steel
plate, and the outer surface is electrostatically sprayed;
2.3 Box body wall material: made of 304 stainless steel plate;
2.4 Gate and insulation sealing material: single door, high density
asbestos,
2.5 partition: 2 layers, adjustable up and down;
Third, the air conditioning system:
3.1 Temperature control mode: Balanced temperature regulation
(BTC), temperature difference with temperature control mode (TB).
The method not only controls the output of the heater through the
PID automatic calculation output according to the set temperature
point, but also compensates for the heat loss, and finally achieves
a dynamic balance control method.
3.2 Air circulation device: There are 2 high-temperature resistant
stainless steel long-axis centrifugal multi-wing fans in the air
conditioning room. The unique air duct forced closed circulation
mode is used to push the circulating airflow through the heater to
the working chamber, which greatly improves the air flow. And heat
exchange capacity with the heater, thereby greatly improving the
temperature uniformity.
3.3 Air heating mode: stainless steel closed electric heating tube,
non-contact control (solid state relay).
Fourth, the control system
4.1 Control platform Shanghai Haihong intelligent temperature
controller;
4.2 Display mode LED digital tube display;
4.3 Setting method Film type button input;
4.4 Operation mode Fixed value operation mode, full digital PID
self-tuning adjustment;
4.5 Setting range Temperature: 0~500°C;
4.6 Resolution Temperature: 1 ° C;
4.7 Other functions Heating and over-temperature LED dynamic
prompts;
4.8 Control Features Modular PID control with high precision and
stable control;
Thin-film buttons for setting and digital calibration are
convenient and simple;
LED digital display, intuitive and easy to read.
Five, safety protection device
5.1 Safe and reliable grounding device;
5.2 studio digital over-temperature protection; (optional)
5.3 Heater short circuit, motor overheating heat, leakage
protection;
5.4 Alarm device: sound and light alarm. (optional)
Sixth, equipment use conditions
6.1 Working voltage
Frequency and error: 50±0.5Hz;
Voltage and error: AC 220±10%V, three-phase five-wire system;
Main power switch: aerodynamic switch with leakage protection;
6.2 Spatial location:
For ventilation, operation and maintenance, please place the
equipment according to the space requirements listed below;
At least one space should be left around the equipment, ie ≥800mm;
The space at which the door of the device is opened can be
arbitrarily switched.
6.3 Environmental conditions around the equipment
Temperature range: 5 ~ 40 ° C;
Humidity range: ≤85%R.H.;
Atmospheric pressure: 86 ~ 106KPa;
Ground requirements: smooth, well ventilated;
Ambient air quality requirements: no high concentration of dust and
flammable, explosive gas or dust, no strong electromagnetic
radiation source nearby.
Seven, product implementation standards:
7.1 JB/T 5520-91 drying oven technical conditions.