Quantitative Sugar Analysis HPLC System High Performance Liquid Chromatography Instrument With Optional Detectors
 
Product Description

High-performance liquid chromatography (HPLC), formerly referred to as high-pressure liquid chromatography, is a technique in analytical chemistry used to separate, identify, and quantify each component in a mixture. It relies on pumps to pass a pressurized liquid solvent containing the sample mixture through a column filled with a solid adsorbent material. Each component in the sample interacts slightly differently with the adsorbent material, causing different flow rates for the different components and leading to the separation of the components as they flow out of the column.

 

Product Parameter

 

Quantitative Sugar Analysis HPLC System High Performance Liquid Chromatography Instrument With Optional Detectors

Physical Specification 

 
Power requirements220V,50HZ
Ambient temperature4-40℃
Ambient humidity20-85% relative humidity(non-condensing)
System 
Qualitative repeatability(Retention time)RSD≤0.3%
Quantitative repeatability(Area)RSD≤0.6%
Pump 
Flow rate range0.0001-10.000mL/min( increments :0.0001 mL/min)
Max operating Pressure42MPa
Pressure pulsation<0.1MPa( Test condition: 10MPa , at 1.0ml/min methanol)
Flow accuracy±0.2%( Test condition: 10MPa , at 1.0ml/min methanol)
Flow precisionRSD≤0.06%( Test condition: 10MPa , at 1.0ml/min methanol)
Gradient accuracy±1%
Seal washWith programmable on-line seal wash module
Column Oven 
Temperature setting range5 °C ~ 85 °C
Temperature stability≤0.2°C/h
Temperature controlling precision±0.1 °C
Autosampler 
Injection Volume Precision

<0.25%RSD( full loop) ;

<0.5%RSD( partial loop) ;

<1%RSD( no waste)

Linearity>0.999
Cross Contamination(Carry Over)<0.01%
Sample Capacity120 samples( 1.5ml vials)
Syringe250 μL( Standard, with 500μL,1000μL, 2500μL optionals)
Sample Loop20μL( Standard, with 50μL, 100μL, 200μL optional)
DAD Detector 
Noise1.5×10-5Au
Drift≤1×10-3Au/h
LOD5.0×10-9g/mL (naphthalene/methanol)
Wavelength range190-800nm
Wavelength accuracy≤±1 nm
Wavelength reproducibility≤±0.1 nm
Linearity≥1.0×104

 

 
 
Product Performance
 
1. Mainstream split design, complete system testing function, to meet most testing needs

2. The new cam curve design adopts electronic pulsation suppression technology and solvent

3. Compression compensation to further reduce pressure fluctuations and ensure higher repeatability of test results.

4. Optional UV, ELSD, FLD and DAD detectors.
 
Product Application
 
Food, medicine, chemical industry, agriculture, etc.

 
 

 

 

 

 

Hot Products

Needle in Flow Path HPLC Ultra-high Performance High Pressure Liquid Chromatography Machine for ...
Extremely Low Cross-contamination UHPLC Ultra-high Performance High Pressure Liquid Chromatography ...
Quantitative Sugar Analysis HPLC System High Performance Liquid Chromatography Instrument With ...
HPLC Chromatografia Aflatoxin Analysis High Pressure Liquid Chromatography Equipment Product ...
HPLC Chromatografia System High Performance Liquid Chromatography in Pesticide Residue Analysis ...
HPLC Chromatografia System High Performance Liquid Chromatography in Pesticide Residue Analysis ...