IE33 ultrasonic diagnostic instrument common fault troubleshooting

 

 

 

 


Fault phenomenon


During the use of the machine, the image is out of sync (image delay, once every two or three days at the beginning), and there is an eha206 error occasionally. After that, the phenomenon of image delay appears every day. It will be fine after shutdown and restart, but it will appear again after a long time.
 

 


Troubleshooting


Check the fault log, and all the errors point to sip. Focus on troubleshooting sip. First, plug and unplug the PCI wiring of dolphin on SIP, and then plug and unplug the board card on SIP. The fault is still the same. Then swap SIP with host's dolphin, and the fault is still there. After several operations, it was found that the sound of a CPU fan of SIP was sometimes abnormal. After careful observation, it is found that the fan speed is not normal sometimes. Check the log again and select temperature. It is found that the temperature of SIP / CPU is too high to 82 ℃ (normally, it should be around 45 ℃), which results in the decline of SIP processing function and image delay. Replace CPU fan and sip, fan temperature returns to normal, troubleshooting.

 

 

 

Ultrasound parts that we can repair:

 

Repair List of Featured Ultrasonic Board & Assy
BrandSystemDescriptionPart Number
GELogiq P5CPU mainboard5168431
GELogiq P6CPU mainboard 
GELogiq E9/vivid E9GFI25161631
GEVoluson E6/Voluson E8RSRKTI301394-2/KTI196357
GEVoluson E6/Voluson E8RFI/ RFI21b BoardKTI300614/KTI302197-6
GEVoluson S6/S8/P8BF645396937-2
GEVoluson S6/S8/P8RFS board5364098-2/5364098-3
GEVoluson S6/S10/P8BF192 board5357234
AlokaAlpha 5mainboardEP493700DD
AlokaAlpha 6RX Beam FormerEP555501AA
AlokaAlpha 6CW boardEP555700
AlokaAlpha 6CPU mainboardEP537000/EP563000
AlokaAlpha 6TX/RXEP555300
AlokaAlpha 6/Alpha 7DOMEP550000AA
AlokaAlpha 10RX Beamformer boardEP531800CE
AlokaAlpha 10wave generator BoardEP495000HH/EP495000HJ/EP495000FF
HitachiF37TX/RSEP557400/EP557300
HitachiF37mainboardEP575700BC/EP560800
AlokaF75USP boardEP556600
AlokaF75Cell BoardEP558800
HitachiHI VISION AviusRDBF board7513907A/7513612
HitachiArietta 60CELLEU-9160C
HitachiArietta 70TXEP572300AA
HitachiArietta 70CELL 
HitachiNoblusCELL 
HitachiNoblusCONT-B 
HitachiAscendusCELL7352830A
SiemensX150/G40BE10131804
SiemensX300BE 2.0/BE 3.0/BE 7.510131990
SiemensX700BE10136465
SiemensX700RC10136467
SiemensX700TI11014292
SiemensAntaresBE7472421
SiemensAntaresCB7288504
SiemensS2000RM30010852163
SiemensS2000 high version/S3000RM30110854749
iU22/iE33EMB excelsior motherboard453561419431/453561382231/453561419431
iU22/iE33UMB Unified Motherboard453561254492/453561360251
HD15ACB acquisition control board453561197145/453561197144
HD15AIM board4.53561E+11
CX50control panel4.53562E+11
CX50mainboard453561368033/453561622163/453561495832
Affiniti 70ACQ acquisition control board 
EPIQ 5/EPIQ 7ACQ Acquisition Module board453561734844/453561704246
BiosoundMylab SeriesCPU mainboard9501361000
ToshibaAplio 300/400/500mainboardPM30-38696
ToshibaAplio 500RXywm2056*a
ToshibaXario 200TXPM30-32732-1
ToshibaXario 200RXPM30-32733-1
Remarks
1.The repair cycle is within 3-5 days.
2.Warranty after fixed:60 days.
3. Kindly send us the estimated main faults (better to have image) for our engineer to prejudge.
4.Package price is available if repair 3 pcs or more at one time.
5.Extra discount for VIP partners.
6.You could have different part No.under same description.

 

 

Hot Products

IE33 ultrasonic diagnostic instrument common fault troubleshooting Fault phenomenon During the use ...
GE vivid7 maintenance case Fault case Turn On the power and press On/Standby. The ultrasonic ...
Toshiba SSA-580A Maintenance Fault phenomenon The machine starts normally, but there is no echo when ...
Ultrasound Machine SIEMENS Acuson Antares The fault phenomenon After the Antares machine starts, the ...
Repair and Replace Toshiba SSA-580A numeric Keyboard Brand: Toshiba SSA-580A Model: BSM31-3078 ...
GE VIVID I System Maintainance First, the failure phenomenon: The machine starts normally, and after ...