Home Companies Xi'An LIB Environmental Simulation Industry

IEC 62660-2 130℃ Temperature Cycling Test Chamber Constant Battery

Xi'An LIB Environmental Simulation Industry

Contact Us

[China] country

Trade Verify

Address: No.1 Zhangba First Street, Xian High-Tech Area, Shaanxi, P.R. China 710065

Contact name:Serena

Inquir Now

Xi'An LIB Environmental Simulation Industry

Verified Suppliers
  • Trust
    Seal
  • Verified
    Supplier
  • Credit
    Check
  • Capability
    Assessment

IEC 62660-2 130℃ Temperature Cycling Test Chamber Constant Battery

Country/Region china
City & Province Shaanxi
Categories Battery Packs
InquireNow

Product Details

Description 

IEC 62660-2 130℃ Battery Temperature Cycling Test Chamber are suited for testing battery reliability under extreme temperature changes.

Battery need to be able to withstand drastic changes in temperature.

The thermal shock and temperature cycling tests called out in the IEC, UL, and SAE standards

 

Technical Specification

NameIEC 62660-2 130℃ Battery Temperature Cycling Test Chamber
ModelTK10-220
Internal Dimension (mm)500*600*750
Overall Dimension (mm)960*1150*1860
Temperature Range (℃)-70℃ ~+150 ℃
Cooling Rate10℃/min (15℃/min)
Heating rate10℃/min (15℃/min)
Temperature Fluctuations± 0.5 ℃
Temperature Deviation± 2.0 ℃
Interior Volume220 liters
ControllerProgrammable color LCD touch screen controller, Multi-language interface, Ethernet , USB
Cooling SystemMechanical compression refrigeration system
Exterior MaterialSteel Plate with protective coating
Interior MaterialSUS304 stainless steel
Power Supply380V
Maximum Noise65 dBA

 

Thermal Test

Battery Temperature Cycling Test Chamber,This test is performed to characterize cell responses to high-temperature environment.

a) Adjust the SOC of the cell to 100% for BEV application, and to 80% for HEV application.

 

b) The cell, stabilized at room temperature, shall be placed in a gravity or circulating air convection oven. The oven temperature shall be raised at a rate of 5K/min to a temperature of 130℃±2K. The cell shall remain at this temperature for 30 min before the test is discontinued.

 

Note:

If necessary, to prevent deformation, the cell may be maintained during the test in a manner that does not violate the test purpose. The manner to prevent deformation should be representative of cells inside battery systems and battery packs.

 

IEC 62660-2 130℃ Battery Temperature Cycling Test Chamber

 

Test Standard

1. IEC 62660-2:The oven temperature shall be raised at a rate of 5K/min to a temperature of 130℃±2K.

2. IEC 62133 - Temp cycling: 75 to 20 to -20°C (30 min. transitions)

3. UL 2054 - Temp cycling: 70 to -40 °C (30 min. transitions)

4. UL 1642 - Temp cycling: 70 to 20 to -40 °C (30 min. transitions)

5. SAE J2464 - Thermal shock test: 70 to -40°C (15 min. transitions)

 

The LIB Battery Temperature Cycling Test Chamber meet IEC 62660-2 standard for test battery

 

Contact us for spare parts!

Hot Products

Description IEC 62660-2 130℃ Battery Temperature Cycling Test Chamber are suited for testing battery ...
Description IEC 62660-2 5K/Min Battery Temp Circulating Air Convection Oven,LIB INDUSTRY supply a ...
Description IEC -40℃ Environmental High Temp Battery Thermal Test Chamber,There are many UL and IEC ...
Description IEC 62660 85℃ Battery High Temperature Environment Test Chamber,The battery industry is ...
Description T-100 Programmable IEC Battery -40℃ Cell Thermal Test Chamber,Whether your company ...
Description IEC 62660 55 ℃ Thermal Cell Stabilized Temperature Test Chamber,environmental chambers ...