Used Philip Goldway UT4000F Multi-Parameter Patient Bedside Monitor

Philip Goldway UT4000F Patient Monitor

 

About Philip Goldway UT4000F Patient Monitor

 

The Philip Goldway UT4000F patient monitor is a medical device used to monitor patients' vital signs and parameters in a healthcare setting.

 

Some of the key features of the UT4000F are:

 

  • Display: It has a high-resolution color TFT display that displays multiple waveforms and numerical data at the same time. The screen size varies depending on the model.

 

  • Parameters: It can measure a number of parameters including ECG, SpO2 (Oxygen Saturation), NIBP (Non-Invasive Blood Pressure), Respiratory Rate, Temperature and optionally Invasive Blood Pressure and cardiac output.

 

  • Alarms: Offers customizable upper and lower alarm limits for each parameter. The monitor alerts healthcare providers when any vital signs exceed pre-set alarm limits.

 

  • Battery: It has a built-in rechargeable battery that ensures uninterrupted monitoring even during power outages or patient transport.

 

  • Data storage and connectivity: It has a built-in memory that can store patient data for up to 72 hours. Enables data transfer via USB or Ethernet to an electronic health record (EMR) system.

 

  • Intuitive user interface: The monitor has an intuitive user interface with touch screen functions for easy navigation and setup.

 

  • Portability: Relatively light and compact, suitable for bedside monitoring or ambulatory use.

 

Troubleshooting for Philip Goldway UT4000F Patient Monitor

 

► when the instrument is turned on. There is no on-screen display and no indicator light.

 

Monitoring and analysis: If the instrument is connected to AC power without being turned on, the power light should be orange (or red); after switching on. The light should be green. Therefore:

(1) Check whether the mains socket and the two fuses on the back of the instrument are normal, that is, check whether there is AC input;

(2) Turn on the instrument and determine whether the switching power supply has a 15V DC output if there is a 220V AC input;

(3) If the instrument is turned on without AC power, the display will not appear and the indicator will not light. Indicates that the battery is low or damaged.

Solution:

(1) Connect all connecting parts reliably and make sure there is AC input;

(2) If your switching power supply does not have 15V output, you need to replace the switching power supply:

(3) Connect AC power to charge the instrument. If the battery is still low after 24 hours of charging, it means the battery is damaged and needs to be replaced.

 

► When the meter is turned on, the screen is blank. The light is always orange (or red).

 

Control and analysis: The light is orange. Indicates that the switching power supply has a 15V DC output:

(1) Turn on the instrument and use a multimeter to check whether the power switch is damaged;

(2) Check that the connection cable between the switch and the motherboard is not loose or damaged.

Solution:

(1) Replace the power switch;

(2) Connect the cable reliably; if it does not turn on, replace the motherboard.

 

 

 

► When the meter is turned on, there is no display on the screen, but the light is green.

 

Check and Analysis: If the button can start inflation at this time or you hear a heartbeat while measuring blood oxygen, it means the meter is turned on, but the TFT screen does not light up. Open the tool and check:

(1) There is an input voltage between 2.5V and 5.4V on the inverter power board or the connector is loose:

(2) Use a multimeter to check whether the inverter has high voltage output.

Solution:

(1) Connect the connection cable reliably and ensure the 2.5-5.4V output:

(2) If there is no high voltage output, replace the inverter.

 

 

 

► When the meter is turned on, the screen is blank. But there is a raster (commonly known as white screen) or the display colors are messed up.

 

Control and analysis: the TFT screen lights up. But the display part of the PC104 board is defective. The display signal cannot be output correctly

Solution: replace the PC104 board.

 

 

 

Specification of Philip Goldway UT4000F Patient Monitor

 

Product Name

Philip Goldway UT4000F Patient Monitor

 

BrandPhilip Goldway
Model noUT4000F
ConditionUsed
Power supplyAC 22V±22V 50Hz
ParameterSPO2 ,ECG ,TEMP,NIBP
MOQOne piece
Warranty90 days

 

 

 

 

 

Thank you so much for your attention .If you're interested in Used Philip Goldway UT4000F Multi-Parameter Patient Bedside Monitor ,feel free to contact us .Thank  you .

Hot Products

Used Philip Goldway UT4000F Multi-Parameter Patient Bedside Monitor Philip Goldway UT4000F Patient ...
Mindray T8 Patient Monitor ,Can Monitor Various Physical Indicators Of Patients Mindray T8 Patient ...
Goldway UT4000F PRO Patient Monitor Philip Goldway UT4000F PRO Patient Monitor By GOLDWAY INDUSTRIAL ...
Goldway UT4000F PRO Multi-parameter patient monitor Goldway UT4000F PRO Multi-parameter patient ...
Mindray Pulse Oximeter PM-60 Brief Introduction of Mindray Pulse Oximeter PM-60 The PM-60 is a ...
Sino-K Medical SPH100 Handheld Spo2 Finger Tip Pulse Oximeter product features: The working ...