► Brief Introduction
The ICC 200 is a high-frequency surgical unit for cutting and coagulation.
Unintentional activation of an HF generator can lead to burns on the patient if the active electrode hereby touches the patient directly or indirectly through electrically conductive objects or wetcloths.Unintentional activation of an HF generator can, for example, be caused by:
► Setting
Basic setting ICC 200 without ENDO CUT option | ||
AUTO CUT | Effect 3 | 120 watts max |
AUTO COAG | FORCED | 60 watts max |
AUTO COAG (with ARGON option) | Spray | 60 watts max |
Footswitch setting | yellow = AUTO CUT | blue = AUTO COAG |
Basic setting ICC 200 with ENDO CUT option | ||
AUTO CUT | Effect 3 | 120 watts max |
AUTO COAG | FORCED | 60 watts max |
AUTO COAG (with ARGON option) | Spray | 60 watts max |
Footswitch setting | yellow = AUTO CUT | blue = AUTO COAG |
ENDO CUT | ON | |
Pulse length | t ON = 50 ms | |
Pause length | t OFF = 750 ms |
► Soft Coagulation
► Forced Coagulation