Home Companies Orient New Life Medical Co.,Ltd.

Methaqualone Rapid Test 3mm Stirp , High Sensitivity Cut-Off 300ng/Ml

Orient New Life Medical Co.,Ltd.
Active Member

Contact Us

Address: 12F,Guting Building,Hefeng Creative Square,No.495 Jiangdong North Road,Ningbo,Zhejiang, China

Contact name:Jerry Meng

Inquir Now

Orient New Life Medical Co.,Ltd.

Methaqualone Rapid Test 3mm Stirp , High Sensitivity Cut-Off 300ng/Ml

City & Province ningbo zhejiang
Categories Water Treatment
InquireNow

Product Details

Drug Abuse Test Kit Methaqualone Rapid Test 3mm Stirp, high sensitivity cut-off 300ng/ml

 

 

Product Name

 

MQL ( Methaqualone) Rapid Diagnostic Test

Specimen: Urine

 

Background

 

Methaqualone (Quaalude, Sopor) is a quinazoline derivative that was first synthesized in 1951 and found clinically effective as a sedative and hypnotic in 1956. 2 It soon gained popularity as a drug of abuse and in 1984 was removed from the US market due to extensive misuse. It is occasionally encountered in illicit form, and is also available in Europeon countries in combination with diphenhydramine (Mandrax). Methaqualone is extensively metabolized in vivo principally by hydroxylation at every possible position on the molecule. At least 12 metabolites have been identified in the urine.

 

The MQL Rapid Test Panel (Urine) is a rapid urine screening test that can be performed without the use of an instrument. The test utilizes a monoclonal antibody to selectively detect elevated levels of Methaqualone in urine. The MQL Rapid Test Panel (Urine) yields a positive result when Methaqualone in urine exceed 300 ng/mL.

 

 

 

Test Principle

 

The FYL Rapid Test Cassette (Urine) is an immunoassay based on the principle of competitive binding. Drugs which may be present in the urine specimen compete against the drug conjugate for binding sites on the antibody.

During testing, a urine specimen migrates upward by capillary action. Fentanyl, if present in the urine specimen below 20 ng/mL, will not saturate the binding sites of antibody-coated particles in the test device. The antibody-coated particles will then be captured by immobilized FYL conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the FYL level exceeds 20 ng/mL because it will saturate all the binding sites of anti-FYL antibodies.

 

A drug-positive urine specimen will not generate a colored line in the test line region, while a drug-negative urine specimen or a specimen containing a drug concentration less than the cut-off will generate a line in the test line region. To serve as a procedural control, a colored line will always appear at the control line region indicating that proper volume of specimen has been added and membrane wicking has occurred.

 

SPECIMEN COLLECTION AND STORAGE

 

The urine specimen must be collected in a clean and dry container. Urine collected at any time of the day may be used. Urine specimens exhibiting visible particles should be centrifuged, filtered, or allowed to settle to obtain clear specimen for testing. Specimen.

 

Urine specimens may be stored at 2-8°C for up to 48 hours prior to testing. For long-term storage, specimens may be frozen and stored below -20°C. Frozen specimens should be thawed and mixed before testing.

 

 

STORAGE

 

Store as packaged in the sealed pouch either at room temperature or refrigerated (2-30°C). The test is stable through the expiration date printed on the sealed pouch. The test must remain in the sealed pouch until use. DO NOT FREEZE. Do not use beyond the expiration date.

 

LIMITATION

 

1. The 6-MAM Rapid Test Cassette provides only a qualitative, preliminary analytical result. A secondary analytical method must be used to obtain a confirmed result. Gas chromatography/mass spectrophotometry (GC/MS) is the preferred confirmatory method2.3.

2. It is possible that technical or procedural errors, as well as other interfering substances in the urine specimen may cause erroneous results.

3. Adulterants, such as bleach and/or alum, in urine specimens may produce erroneous results regardless of the analytical method used. If adulteration is suspected, the test should be repeated with another urine specimen.

4. A positive result indicates presence of the drug or its metabolites but does not indicate level of intoxication, administration route or concentration in urine.

5. A negative result may not necessarily indicate drug-free urine. Negative results can be obtained when drug is present but below the cut-off level of the test.

 

performance characteristics

 

A side-by-side comparison was conducted using the Rapid Test Cassette (Urine) and GC/MS. The following results were tabulated:

 

 

 

 

 

 

MethaqualonePercent ofnVisual Result
Concentration (ng/mL)Cutoff NegativePositive
00%30300
150-50%30300
225-25%30273
300Cutoff301515
375+25%30426
450+50%30030
9003X30030

 


 

ORIENT NEW LIFE MEDICAL CO., LTD.
Contact:Jerry Meng
Email:Jerry @ newlifebiotest .com
Tel.+86 18657312116
SKYPEenetjerry

Hot Products

Drug Abuse Test Kit Lysergic Acid Rapid Test 3mm strip ,high sensitivity cut-off 20ng/ml Product ...
Drug Abuse Test Kit MPD Rapid Test 4mm cassette ,high sensitivity cut-off 1000ng/ml Product Name MPD ...
Drug Abuse Test Kit Methaqualone Rapid Test 4mm cassette ,high sensitivity cut-off 300ng/ml Product ...
Drug Abuse Test Kit Methaqualone Rapid Test 3mm Stirp, high sensitivity cut-off 300ng/ml Product ...
Drug Abuse Test Kit Tropicamide Rapid Test 4mm cassette ,high sensitivity cut-off 350ng/ml Product ...
Drug Abuse Test Kit Tropicamide Rapid Test strip, high sensitivity cut-off 350ng/ml Product Name ...