Hospital Medical equipment Anesthesia / anasthesia / Anestesia machine
Characteristics:
1. Pneumatically driven electronically controlled system
2. Built-in ventilator, 5.4" LCD display screen
3. With 1 patient circuit for adult (Option: patient circuit for children).
4. 4-tube flowmeters, O2 & N2O ;0.1 L/Min ~ 10L/Min;
5. Vaporizer, Enflurane/Isoflurane/Sevoflurane;
(Option: Halothane)
6. Ptr: -10~10 hPa;
7. Respiration Mode: IPPV, SIPPV(VCV), IMV, SIMV, PEEP, Manu, Sigh
8. Tidal Volume: 0~1500mL;
9. Respiration Frequency: 1~99bpm;
10. I/E Ratio: 4:1~1:10
Details Images
TECHNIQUES SPECIFICATIONS:
Physical specifications
Note
Screen
5.4'' Display Screen
Standard
Suitable
Adult & Child
Standard
Mode
pneumatically powered and electrically controlled system
Standard
Working Mode
Closed; Semi-Closed; Semi-Open
Standard
Circuit
Breathing circuit integrated standards
Standard
Tube
4Tubes Flowmeters: O2:0.1~10L/Min, N2O:0.1~10L/Min
optional
Trolley
Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated
brake provisions
Drawer unit
ONE drawer totally extractable
Technical Specifications:
Gas requirement:
Medical oxygen and nitrous oxide with a pressure ranging from
O2: 0.32~0.6MPa ; NO2: 0.32 MPa to 0.6 MPa.
safety valve
<12.5 kPa
Respiratory Rate
1~99bpm
Flow meter
O2:0.1~10L/Min, N2O:0.1~10L/Min
oxygen concentration in the mixed gas N2O/O2
> 21%
Oxygen Flush
25~75 L/min
modes of ventilation
A/C , IPPV, SIPPV, IMV, SIMV, MANUAL
I/E Ratio
4:1 ~ 1:10, Has the inverse ratio ventilation
Tidal Volume
0~1500ML
ptr inspiratory triggering pressure
-0.1kPa~1.0kPa
Inspiratory Plateau
0~1s
Maximum safety pressure
≤ 12.5 kPa
Pressure limit range
0 ~ 6.0 kPa
Airway pressure alarm: Audible and visual and with yellow and red color indicating
Lower: 0.2kPa~5.0kPa; Upper: 0.3~6.0 kPa
±0.2 kPa
Tidal volume alarm: Audible and visual and with yellow and red color indicating
upper alarm: 50 to 2000ml, lower alarm: 0~1800ml
Power Supply Alarm
Ac/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s
The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.
Company Profile

Hot Products

#detail_decorate_root .magic-0{width:750px}#detail_decorate_root .magic-1{overflow:hidden;width...
#detail_decorate_root .magic-0{width:750px}#detail_decorate_root .magic-1{overflow:hidden;width...
#detail_decorate_root .magic-0{border-bottom-style:solid;border-bottom-color:#53647a;font-family...
#detail_decorate_root .magic-0{border-bottom-style:solid;border-bottom-color:#53647a;font-family...
#detail_decorate_root .magic-0{border-bottom-style:solid;border-bottom-color:#53647a;font-family...
#detail_decorate_root .magic-0{border-bottom-style:solid;border-bottom-color:#53647a;font-family...