Guangzhou Anlai Airtech Co.,Ltd |
|
Clean work bench/Vertical Laminar Air Flow Hood
• Economic laminar flow hoods provide with a Polypropylene
worksurface can be positioned on nearly any bench.
• Available with Powder coated steel mobile Stand
• Consists of a integrated Fan Filter module. Meets IEST recommended practices. low sound, low watts Measuring only 50 dBA. Running at only 240 watts @ 90 fpm 490 CFM @ 250 Watts
• Choice of HEPA: 99.99% efficient @ 0.3µm particles or ULPA: 99.999% efficient @ 0.12µm particles
• Minihelic gauge (0 – 3” Water Column) monitors pressure differential when it exceeds the limit, it indicates the need to replace the HEPA filter(s).
• Each model includes a tilt-up sash that increases air speed by restricting the air flow
• Including Built-in Fluorescent Light
• Optional Hospital Grade Power Strip (2-Outlets)
Horizontal Flow Clean Benches
If you do not use large or bulky equipment on the work surface and
want to ensure the lowest level of turbulence, horizontal flow
might be best for you.
Risks of contamination from operators may be reduced in horizontal
laminar flow hoods because hands are located downstream of the
airflow. That means anything you might introduce when working
within the hood is immediately pushed out of the enclosure.
Application:
It can be widely used in medicine, bio-pharmaceutical, food,
optics, microelectronics, sterile laboratory, lant tissue culture
etc.
Horizontal CLEAN BENCH Specifications:
Model | AL-LC1200 | AL-LC1500 | AL-LC1800 |
External Size (WxLxH)mm | 1200x780x1580 | 1500x780x1580 | 1800x780x1580 |
Internal Size (WxLxH)mm | 1100x600x650 | 1400x600x650 | 1700x600x650 |
Air Flow (m³/h) | 800 ~ 1200 | 1000 ~ 1500 | 1500 ~ 2000 |
Wind Velocity | 0.25 ~ 0.45 m/s | ||
Purification rank | 100 rank (ISO14644-1 standard) |
Vertical Laminar Flow Hoods- Economical Clean Bench
Economical laminar flow cabinet with clean, simple, low-maintenance design is a work bench with class 100 / ISO5 environment can be used in assembly, testing and inspection procedures that require a continuous flow of ultra clean air.
These bench-top polypropylene vertical laminar flow hoods feature 99.99% efficient HEPA or 99.999% efficient ULPA fan / filter modules that remove particulate airborne contamination.
Vertical CLEAN BENCH Specifications:
Model | AL-VC1200 | AL-VC1500 | AL-VC1800 |
External Size (WxLxH)mm | 1200x700x1800 | 1500x700x1800 | 1800x700x1800 |
Internal Size (WxLxH)mm | 1100x650x600 | 1400x650x600 | 1700x650x600 |
Air Flow (m³/h) | 800 ~ 1200 | 1000 ~ 1500 | 1500 ~ 2000 |
Wind Velocity | 0.25 ~ 0.45 m/s | ||
Purification rank | 100 rank (ISO14644-1 standard) |
What does a clean bench do?
The clean bench is recommended for work with non-hazardous
materials where clean, particle-free air is necessary to avoid
contamination. The clean bench ensures that the work surface is
constantly flooded with HEPA-filtered air in a laminar flow. Unlike
a biological safety cabinet, the clean bench protects the work on
the work surface, but not personnel or the surrounding environment
from aerosols created on the work surface. The HEPA air filter can
trap 99.999% of particles with a diameter greater than and less
than 0.3 micron.
Why Laminar flow?
Clean benches are equipped with either vertical or horizontal
laminar flow. Both of these configurations provide a unidirectional
velocity along parallel flow lines. Simply, all of the air inside
the confines of the bench flows in a single direction without
backwash, typically an an area of turbulent airflow created by
obstructions, equipment, or movement within the clean bench
enclosure. Proper enclosure design helps prevent backwash and
ensures that the vertical or horizontal laminar air flow continues
to protect work on the surface of the clean bench.