KTU310 Online Turbidity Sensor 90 ° Angle Scattering Principle Fiber Structure, Strong Anti-Interference Ability

1.Principle

The KTU310 integrated on-line turbidimeter is designed and made by the principle of scattering light turbidimeter. When a beam of light enters a water sample, the light is scattered by the turbidity material in the water sample. The turbidity in the water sample can be calculated by measuring the intensity of the scattered light in the vertical direction of the incident light and comparing it with the internal calibration value, the final value is output after linearization.

2.Feature

90 ° angle scattering principle, built-in temperature sensor

Support RS-485, Modbus/RTU protocol

Fiber structure, strong anti-interference ability of external light

Infrared LED light source, high stability

IP68, shield, water depth, 20 meters

Convenient, fast, stable and easy to maintain

3.Technical Specifications

Model numberKTU310
Measuring principleScattering method
Range0 ~ 100 NTU
Resolution0.1 NTU, 0.1 °C
Precision

± 3% or ± 2 NTU

± 0.5 °C

Calibration modeTwo-point calibration
Temperature compensationAutomatic temperature compensation (PT1000)
Output modeRS-485(Modbus/RTU)
Working conditions0ー50 °C, < 0.2 mpa
Storage temperature- 5 ~ 65 °C
Installation modeImmersion mounting, 3/4 NPT thread
Power consumption< 0.05 W
Power supply12 ~ 24 VDC ± 10%
Protection levelIP68
 

 

4.Dimensional Drawing

5.Installation and Electrical Connection

5.1 Installs

Installation distance: 5 cm above the side wall and 10 cm above the bottom.

5.2 Electrical Connection

The cable is a 4-core twisted pair shielded wire defined in line sequence:

Red wire -- power cord (12V-24VDC)

Black line (GND)

Blue Line -- 485a

White line -- 485b

Before electrification, the wiring sequence should be carefully checked to avoid unnecessary losses caused by wiring errors.

Wiring instructions: Consider the cable long-term immersion in water (including seawater) or exposed to air, all wiring are required to do waterproof treatment, customer cable should have a certain anti-corrosion ability.

 

6. Maintenance

6.1 Maintenance Procedures and Methods

6.1.1 Maintenance Schedule

Measuring the cleanliness of windows is important for maintaining accurate readings.

Maintenance taskRecommended maintenance frequency
Calibration of sensors (if required by competent authority)According to the maintenance schedule required by the competent authority
 

6.1.2 Maintenance Methods

Sensor outer surface: clean the sensor outer surface with tap water, if there is still debris residue, wipe with a damp soft cloth, for some stubborn dirt, you can add some household cleaning liquid in tap water to clean.

Check the wire of the sensor: the wire should not be taut when working normally, otherwise it is easy to break the wire inside the wire and make the sensor can not work normally.

Check if the measuring window of the sensor is dirty and the cleaning brush is normal.

6.1.3 Points to Note

The probe contains sensitive optical and electronic components. Make sure the probe doesn’t get hit by any heavy machinery. There are no parts inside the probe that require user maintenance.

6.2 Calibration of Sensors

Zero-point calibration: take a proper amount of zero-turbidity liquid with a large beaker and place the sensor vertically in the solution. The front of the sensor should be at least 10 cm from the bottom of the Beaker. zero-point calibration should be carried out after 3 ~ 5 minutes of numerical stabilization. Instruction Reference Appendix.

Slope calibration: the sensor probe is placed in standard solution, the sensor front end is at least 10 cm from the bottom of the Beaker, and slope calibration is carried out after 3 ~ 5 minutes of numerical stabilization. Instruction Reference Appendix.

6.3 FAQ

QuestionsProbable causeSolution
The interface can not be connected or the measurement results are not displayedMeasurement value is too high, too low or value continues to be unstableReconnect the controller to the cable
Cable failurePlease contact us
Measurement value is too high, too low or value continues to be unstableThe sensor window is attached to somethingClean the surface of the sensor window
 

 

 

 

7. Quality and Service

7.1 Quality Assurance

Quality inspection departments have standardized inspection procedures, with advanced and perfect testing equipment and means, and in strict accordance with the rules of inspection, 72-hour aging of products to do experiments, stability experiments, do not let a unqualified product factory.

The consignee will return the lot of the product with 2% nonconformity rate, and all the cost will be borne by the supplier. The standard of consideration refers to the product description provided by the supplier.

Guarantee the quantity and speed of delivery.

7.2 Spare Parts

This product includes:

1 sensor

1 copy of instruction manual

Certificate 1

7.3 After-sales Service Commitment

Our company provides after-sale service for one year from the date of sale, but does not include damage caused by improper use. If repair or adjustment is needed, please return it at your own expense, the company will repair the damage of the instrument free of charge.

 

 

Hot Products

1.Application Environment Drinking water/surface water/various water supplies/industrial water ...
1.Principle KFDO310 integrated on-line fluorescence dissolved oxygen sensor is designed and made ...
KPH310 Online PH Sensor Signal Output: RS-485 Double High Impedance Differential Amplifier Ip68 1...
KTU310 Online Turbidity Sensor 90 ° Angle Scattering Principle Fiber Structure, Strong Anti...
KOP300 KOP300 Online ORP Sensor Drinking Water Waste Water Surface Water Glass Electrode Method IP68 ...
KEC300 Online Conductivity Sensor Water Quality Sensor Instrument Introduction The conductivity ...