220V Flame Retardant Test In Fires Heat Release Rate Checking
Thermal Conductivity Lab
Product Application
The name Cone Calorimeter is derived from the conical heater as
used by Dr. Vytenis Babrauskas to inspect test specimen(100×100)
with flux up to 100/ for the bench scale oxygen depletion of his
development.
ISO Solution Fire Testing Equipment Cone Calorimeter is based on
the theory that pure combustion calories are in proportion to a
necessary amount of oxygen for combustion and 13.1 MJ/kg is
generated whenever oxygen 1kg is consumed, and heat emission,
ignition time, oxygen consumption, CO and CO2 generation, and flow
of ignited gases as generated from test materials are measured.
ISO Solution Fire Testing Equipment Cone Calorimeter, as properly
sized for lab use, is designed to enable auto operation of all
units connected with DAQ(Data Acquisition) Program so that user can
conveniently perform test. Also, it reduces installation space and
automatically advances test with the 17inch Touch Screen. This
tester fully meets the international standard.
Technology data:
1.) 19inch standard instrument frame,17inch touchable screen pc
insert
2.) cone heater powering rating:5000w,heat output rate:0~100kw/m2
3.) in the range of 50*50mm in the center surface of expose test
sample ,center irradiation error no more than +-2%;
4.) sample size:100mm*100mm*50mm
5.) weight range:0~2000g;precision:0.1g;
6.) ignition system with safe cut device high pressure spark
generator ,automatic positioning
7.) clockwise magnetic oxygen analyzer ,adopt pressure change of
clockwise magnetic to measure the oxygen density
8.) density range:0-25%,T90<1.5S,zero
drift:<0.5%month,measure data deviation<0.5%month,
9.) non-dispersion infrared co and co2 analyzer:co:0~1%;co2:0~10%;
10.) smoke analyzer use laser system test smoke density,and this
system was made up of photoelectricity two-electrode valve,0.5mv
He-Ne laser,main graphics detector and auxiliary graphics detector
.
11.)exhaust system was made up of fan,hottle,air intake and exhaust
tube of fan and orifice meter.exhaust gas flow 0~50g/s can be
adjusted,precision 0.1g/s.
12.)ring sampler need to be located in the air take tube at
distance of 685mm against hottle ,need to be 12 holes existed in
the sampler in order to homogeneous air flow component.
13.) exhaust gas flow need to be measured by the differencial
pressure of side of hole plate above the fans at the distance of
350mm,the inner diameter of hole plate is 57mm+-1mm;
14.) the temperature of air flow is tested by armored thermocouple
with closed node of diameter 1.6mm,thermocouple need to be located
in 100 mm above the orifice plate;
15.) gas sampling system including ring sampler ,sampling
pump,filter,cold trap,waste water exhaust,water filter and co2
filter;
16.) cold trap:0~5 degree,diaphragm pump,flow rate :261/min,vacuum
degree:700mm Hg
Sample size | 100*100*50mm |
Weight range | 0-200g |
precision | 0.1g |
Density range | 0-25% |
T90 | <1.5S |
Zero drift | <0.5%month |
Measure data deviation | <0.5%month |
Structure of cone calorimeter
1.)The combustion chamber: the conical heater, the 10KV ignition
device, the control circuit, the windshield, etc..
2.)Oxygen analyzer: Servomex analyser, oxygen analyzer can
accurately test the content of oxygen in the air pipe in the time
of the change, and then by the instantaneous oxygen concentration
and oxygen consumption principle to determine the heat release of
the material.
3.)Load table: a device for measuring the quality of the sample,
which can accurately record the mass change of the sample in the
process of combustion
4.)Smoke measurement system: He Ne laser emitter and double
electron beam measuring device, which can be used to determine the
specific extinction area (SEA) of the smoke in the smoke pipe.
5.)Ventilation system: the ventilation system is a sample of the
combustion products from the combustion chamber to the atmosphere
in the device
6.) Other improvement equipment: according to different needs, can
also add other analysis device, such as the analysis of the
combustion products, can increase the infrared spectrum analysis
device; if the temperature distribution in the sample, the
corresponding thermocouple or infrared imaging device.
7.) Auxiliary equipment: auxiliary equipment includes microcomputer
processor, heat flow meter device, removing CO2 and H2O (gas),
etc..