Customized Human ECP Sandwich ELISA Kit High Specificity Eosinophil Cationic Protein ELISA Assay Kit

 

Cat.No E1391Hu

 

Intended Use

This sandwich kit is for the accurate quantitative detection of Human Eosinophil Cationic Protein (also known as ECP) in serum, plasma, cell culture supernates, cell lysates, tissue homogenates.

 

Sensitivity: 0.25ng/ml

Standard Curve Range: 0.5ng/ml - 100ng/ml

*This product is for research use only, not for use in diagnosis procedures. It’s highly recommend to read this instruction entirely before use.

 

Precision

Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision.

Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision.

CV(%) = SD/mean x 100

Intra-Assay: CV<8%

Inter-Assay: CV<10%

 

Reagent Provided

ComponentsQuantity
Standard Solution (120ng/ml)0.5ml x1
Pre-coated ELISA Plate12 * 8 well strips x1
Standard Diluent3ml x1
Streptavidin-HRP6ml x1
Stop Solution6ml x1
Substrate Solution A6ml x1
Substrate Solution B6ml x1
Wash Buffer Concentrate (25x)20ml x1
Biotinylated Human ECP Antibody1ml x1
User Instruction1
Plate Sealer2 pics
Zipper bag1 pic

 

Material Required But Not Supplied

  • 37°C±0.5°C incubator
  • Absorbent paper
  • Precision pipettes and disposable pipette tips
  • Clean tubes
  • Deionized or distilled water
  • Microplate reader with 450 ± 10nm wavelength filter

 

Reagent Preparation

All reagents should be brought to room temperature before use.

Standard Reconstitute the 120μl of the standard (120ng/ml) with 120μl of standard diluent to generate a 60ng/ml standard stock solution. Allow the standard to sit for 15 mins with gentle agitation prior to making dilutions. Prepare duplicate standard points by serially diluting the standard stock solution (60ng/ml) 1:2 with standard diluent to produce 30ng/ml, 15ng/ml, 7.5ng/ml and 3.75ng/ml solutions. Standard diluent serves as the zero standard(0 ng/ml). Any remaining solution should be frozen at -20°C and used within one month. Dilution of standard solutions suggested are as follows:

 

60ng/mlStandard No.5120μl Original Standard + 120μl Standard Diluent
30ng/mlStandard No.4120μl Standard No.5 + 120μl Standard Diluent
15ng/mlStandard No.3120μl Standard No.4 + 120μl Standard Diluent
7.5ng/mlStandard No.2120μl Standard No.3 + 120μl Standard Diluent
3.75ng/mlStandard No.1120μl Standard No.2 + 120μl Standard Diluent

 

Standard ConcentrationStandard No.5Standard No.4Standard No.3Standard No.2Standard No.1
120ng/ml60ng/ml30ng/ml15ng/ml7.5ng/ml3.75ng/ml

 

Wash Buffer Dilute 20ml of Wash Buffer Concentrate 25x into deionized or distilled water to yield 500 ml of 1x Wash Buffer. If crystals have formed in the concentrate, mix gently until the crystals have completely dissolved.

 

Assay Procedure

1. Prepare all reagents, standard solutions and samples as instructed. Bring all reagents to room temperature before use. The assay is performed at room temperature.

2. Determine the number of strips required for the assay. Insert the strips in the frames for use. The unused strips should be stored at 2-8°C.

3. Add 50μl standard to standard well. Note: Don’t add antibody to standard well because the standard solution contains biotinylated antibody.

4. Add 40μl sample to sample wells and then add 10μl anti-ECP antibody to sample wells, then add 50μl streptavidin-HRP to sample wells and standard wells (Not blank control well). Mix well. Cover the plate with a sealer. Incubate 60 minutes at 37°C.

5. Remove the sealer and wash the plate 5 times with wash buffer. Soak wells with at least 0.35 ml wash buffer for 30 seconds to 1 minute for each wash. For automated washing, aspirate all wells and wash 5 times with wash buffer, overfilling wells with wash buffer. Blot the plate onto paper towels or other absorbent material.

6. Add 50μl substrate solution A to each well and then add 50μl substrate solution B to each well. Incubate plate covered with a new sealer for 10 minutes at 37°C in the dark.

7. Add 50μl Stop Solution to each well, the blue color will change into yellow immediately.

8. Determine the optical density (OD value) of each well immediately using a microplate reader set to 450 nm within 10 minuets after adding the stop solution.

 

Calculation of Result

Construct a standard curve by plotting the average OD for each standard on the vertical (Y) axis against the concentration on the horizontal (X) axis and draw a best fit curve through the points on the graph. These calculations can be best performed with computer-based curve-fitting software and the best fit line can be determined by regression analysis.

Hot Products

MCP-1 Canine Sandwich ELISA Kit High Precision Monocyte chemotactic protein 1 ELISA Assay Kit Cat.No ...
96 Well Plate Human Copeptin ELISA Test Kit With High Precision and Specificity Customized Reagent ...
Customized Copeptin Human ELISA Test Kit 96 Well Plate Highly Sensitive With Oem Service Cat.No ...
Laboratory Research Human Copeptin ELISA Assay Kit 96 Wells Size With Strong Sensitivity Cat.No ...
Customized Human ECP Sandwich ELISA Kit High Specificity Eosinophil Cationic Protein ELISA Assay Kit ...
Strong Sensitivity and Specificity 96 Wells ECP Human Sandwich ELISA Kit With 2 Hours Assay Time Cat...