Bovine Vitamin C ELISA Kit with High Spesificity and Sensitivity
Cat.No E2050Bo
Standard Curve Range: 5mg/L - 1800mg/L
Sensitivity: 2.54mg/L
Size: 96 wells
Storage: Store the reagents at 2-8°C. For over 6-month storage refer to
the expiration date keep it at -20°C. Avoid repeated thaw cycles.
If individual reagents are opened it is recommended that the kit be
used within 1 month.
* This product is for research use only, not for use in diagnosis
procedures. It’s highly recommend to read this instruction entirely
before use.
Intended Use
This sandwich kit is for the accurate quantitative detection of
Bovine Vitamin C (also known as VC) in serum, plasma, cell culture
supernates, cell lysates, tissue homogenates.
Assay Principle
This kit is an Enzyme-Linked Immunosorbent Assay (ELISA). The plate
has been pre-coated with Bovine VC antibody. VC present in the
sample is added and binds to antibodies coated on the wells. And
then biotinylated Bovine VC Antibody is added and binds to VC in
the sample. Then Streptavidin-HRP is added and binds to the
Biotinylated VC antibody. After incubation unbound Streptavidin-HRP
is washed away during a washing step. Substrate solution is then
added and color develops in proportion to the amount of Bovine VC.
The reaction is terminated by addition of acidic stop solution and
absorbance is measured at 450 nm.
Reagent Provided
Components | Quantity |
Standard Solution (640nmol/ml) | 0.5ml x1 |
Pre-coated ELISA Plate | 12 * 8 well strips x1 |
Standard Diluent | 3ml x1 |
Streptavidin-HRP | 6ml x1 |
Stop Solution | 6ml x1 |
Substrate Solution A | 6ml x1 |
Substrate Solution B | 6ml x1 |
Wash Buffer Concentrate (30x) | 20ml x1 |
Biotinylated Bovine VC Antibody | 1ml x1 |
User Instruction | 1 |
Plate Sealer | 2 pics |
Zipper bag | 1 pic |
Specimen Collection
Serum Allow serum to clot for 10-20 minutes at room temperature.
Centrifuge at 2000-3000 RPM for 20 minutes.
Plasma Collect plasma using EDTA or heparin as an anticoagulant.
Centrifuge samples for 15 minutes at 2000-3000 RPM at 2 - 8°C
within 30 minutes of collection.
320nmol/ml | Standard No.5 | 120μl Original Standard + 120μl Standard Diluent |
160nmol/ml | Standard No.4 | 120μl Standard No.5 + 120μl Standard Diluent |
80nmol/ml | Standard No.3 | 120μl Standard No.4 + 120μl Standard Diluent |
40nmol/ml | Standard No.2 | 120μl Standard No.3 + 120μl Standard Diluent |
20nmol/ml | Standard No.1 | 120μl Standard No.2 + 120μl Standard Diluent |
Standard Concentration | Standard No.5 | Standard No.4 | Standard No.3 | Standard No.2 | Standard No.1 |
640nmol/ml | 320nmol/ml | 160nmol/ml | 80nmol/ml | 40nmol/ml | 20nmol/ml |
Wash Buffer Dilute 20ml of Wash Buffer Concentrate 30x into deionized or
distilled water to yield 500 ml of 1x Wash Buffer. If crystals have
formed in the concentrate, mix gently until the crystals have
completely dissolved.
Assay Procedure
1. Prepare all reagents, standard solutions and samples as
instructed. Bring all reagents to room temperature before use. The
assay is performed at room temperature.
2. Determine the number of strips required for the assay. Insert
the strips in the frames for use. The unused strips should be
stored at 2-8°C.
3. Add 50μl standard to standard well. Note: Don’t add antibody to standard well because the standard solution
contains biotinylated antibody.
4. Add 40μl sample to sample wells and then add 10μl anti-VC
antibody to sample wells, then add 50μl streptavidin-HRP to sample
wells and standard wells (Not blank control well). Mix well. Cover
the plate with a sealer. Incubate 60 minutes at 37°C.
5. Remove the sealer and wash the plate 5 times with wash buffer.
Soak wells with at least 0.35 ml wash buffer for 30 seconds to 1
minute for each wash. For automated washing, aspirate all wells and
wash 5 times with wash buffer, overfilling wells with wash buffer.
Blot the plate onto paper towels or other absorbent material.
6. Add 50μl substrate solution A to each well and then add 50μl
substrate solution B to each well. Incubate plate covered with a
new sealer for 10 minutes at 37°C in the dark.
7. Add 50μl Stop Solution to each well, the blue color will change
into yellow immediately.
8. Determine the optical density (OD value) of each well
immediately using a microplate reader set to 450 nm within 10
minuets after adding the stop solution.