Home Companies State (Beijing) Medical Technology Co., Ltd.

Pig / Dog / Horse Veterinary Ultrasound Scanner Full Digital M Display Format

State (Beijing) Medical Technology Co., Ltd.
Active Member

Contact Us

[China] country

Address: Room D312,Floor 3,Building 3,Jia 2 Jinfu Road,Economic Development-area, Daxing District,Beijing,China.

Contact name:Cathy Wu

Inquir Now

State (Beijing) Medical Technology Co., Ltd.

Pig / Dog / Horse Veterinary Ultrasound Scanner Full Digital M Display Format

Country/Region china
City & Province beijing beijing
InquireNow

Product Details

Pig/Dog/Horse/Cattle Used Ultrasound & Full Digital Vet Palm Ultrasound Scanner

 

 

Check Before Starting

 

Before starting, pls check the following item

NOItemInspection
1Power supply in line with standard single-phase three-wire power(AC220V±22V or AC110V±11V )Test
2Probes and connectors must not be damaged, scratched and shedding; joints no loose connections, no cracks in the shellVisual inspection
3power line. Ensure no anomalies(such as scratch and crack)Visual inspection
4Cleaning and disinfection of the probeVisual inspection
5Systems and peripheral equipment can not be deformed, destroyed or with stains and other phenomenaVisual inspection
6Debris is not placed on the control panelVisual inspection
7tempreture,humidity and pressure should meet with requirementsTempreture/humidity inspection
8clean venues and environmentVisual inspection
9There must be no obstructions in the vicinity of the deviceVisual inspection
10Condensation can not occurVisual inspection
11Control panel, display and casing, to ensure that no anomalies (such as cracks)Visual inspection
12Check the Local auxiliary power outlet and all I / O ports to ensure that no anomalies (such as damage or foreign material clogging)Visual inspection


 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Power Examination

 

Should always check the instrument power supply, when the supply voltage exceeds the specified range of adaptation instrument (AC220V ± 22V, 50Hz ± 1Hz), there shall be booted

Always check the power cord and the instrument probe cable, if found damaged, breakage, should be replaced immediately.

After the boot panel cooling fan should check whether the normal operation should be stopped if found shutdown process.

 

Open Power Supply

 

1) Open power supply, allow the device to complete the boot menu and the image will appear.

2) Check whether the probe surface when using abnormal heat

 

warning:Use abnormal heat may burn the patient probe
 

 

Attention :When you start the system or switch probe, may hear "crackling" sound, which is a normal phenomenon.
 

3) Complete the following steps:

Check image,ensure there are no any non-normal situation

Check the control panel, make sure buttons and knobs work properly.

You should also check the following items are normal:

NOIteminspection
1There must be no abnormal sound, smell or overheatingVisual inspection, hear
2No error message is displayedVisual inspection
3Abnormal noise may not be obvious, without the presence of black or continuously displayed on the B-mode image area.Visual inspection
4Acoustic lens surface probe must not overheat (hand-checked)Touch feeling
5Function keys on the keyboard is normaloperation
6No black line screen images without interferenceVisual inspection

 

warning:If you find any abnormality, indicating that the machine is defective. Should be shut down immediately and contact the marketing department, service department or agency.

 

4.4 Diagnosis

Put gel on veterinary’s parts, and make the probe touch examination organs. In this case the screen section image will be displayed on the screen. If changing the direction and inclination of the probe can observe different images section , select the desired image section , as a basis for diagnosis. In the meantime, if the image's brightness , contrast and gain inappropriate , you can adjust the brightness of the monitor repeatedly , contrast, or adjust the signal gain until you get a clear picture.

 

Reboot Instrument

 

When an exception occurs appear following situations , you can switch off / on the way to restart the system :

1) The screen displays an error message, further more, the message is always displayed ;

2 ) display abnormal ;

3 ) the system is not operating properly .

 

Power off

 

1) be sure to turn the unit off when don’t use machine. Before turning off the power, you should follow these steps:

2) probe should be put in the corresponding probe cup sleeve.

3) according to the operation manual request to close all of the power device.

Warning: In order to ensure the safe and effective operation, it should routine maintenance and inspection. Please perform the following examination before each start. Upon checking into the abnormal situation, immediately shut down the system and contact our company marketing department, service department or dealer. If you use the non-normal system will hurt patient and equipment.

 

 

 

Hot Products

Veterinary Ultrasound Machine/vet Handheld Ultrasound Scanner/vet Ultrasound Warning: 1. Don’t move ...
Portable Vet Ultrasound & Veterinary Portable Ultrasound Scanner 1. Requirement on Environment 1. ...
3.5 MHz Probe Palm Veterinary Ultrasound Scanner M Mode Display High Resolution 1. Technical ...
Clear Image Swine Veterinary Ultrasound Scanner Machine 3 Stage Frame Function OOBA After the device ...
Pig/Dog/Horse/Cattle Used Ultrasound & Full Digital Vet Palm Ultrasound Scanner Check Before ...
Medical Digital Palm Ultrasound Scanner With Mini Design Operation Precautions 1 ) If the fuse or ...