The Component of Product
The BLX-6 high frequency oral X-ray machine is composed of the main
machine and the ball head.
The host machine is mainly composed of power supply, data central
processor, and oscillation, amplification and feedback parts. The
ball head is composed of booster, voltage doubling, frequency
doubling, ball tube, vacuum insulation sealing and other parts. The
accessory shell parts are molded and sealed for protection.
Product Usage
BLX-6 high frequency oral X-ray machine is mainly used for the
comparison of the examination before treatment and effect during
the treatment in the dental clinic, and also the confirmation of
the treatment effect after the treatment.For tooth extraction, to
understand the internal structure of the gums, the depth of the
root, the degree of inflammation of the pulp, even the internal
situation of broken teeth, etc., is an indispensable equipment in
the daily work of the clinic.Because the dose is so small and there
is no need for protection, some hospitals use this machine for
finger and foot photography for bedside examination.
Product Parameters
Power Supply | Voltage AC220V±10% 50Hz Grid Power Resistance 0.52 |
Ball Tube Current | 0.2MA |
Nominal Electric Power | 60kV, 0.2MA,0.5s,12W |
Frequency | 30 KHz |
Exposure Time | 0.5-9.9 seconds, intermittent loading ,continuous operation |
Radiation Leakage | <20ugy/H |
Input Power | 160VA |
Focal Length | ≥20cm |
The Bulb Tube Voltage | 60KV |
Relative Humidity | <80% |
Ambient Temperature | 5-40 degrees |
Gross Weight | 5 kgs |
The Weight of The Machine | 2kgs |
Total Filtration | 2.0mmAL (1.5mmAL inherent filtration, 0.5mmAL additional
filtration) |