Home Companies Guangzhou BioKey Healthy Technology Co.Ltd

Infectious Hypodermal Hematopoietic Necrosis Virus ( IHHNV ) Nucleic Acid Taqman QPCR Kit

Guangzhou BioKey Healthy Technology Co.Ltd

Contact Us

Trade Verify

Address: Add:No.2, Ruitai Rd, Guangzhou Hi-Tech Industrial Development Zone, 510670, CHINA

Contact name:Lisa

Inquir Now

Guangzhou BioKey Healthy Technology Co.Ltd

Verified Suppliers
  • Trust
    Seal
  • Verified
    Supplier
  • Credit
    Check
  • Capability
    Assessment

Infectious Hypodermal Hematopoietic Necrosis Virus ( IHHNV ) Nucleic Acid Taqman QPCR Kit

City & Province guangzhou
Categories Dyestuff Intermediates
InquireNow

Product Details

Infectious hypodermal and hematopoietic necrosis virus(IHHNV) Nucleic Acid Detection Kits

(PCR-Fluorescent Probe Method)

 

 

Intended use:

 

This kit is suitable for the specific detection of IHHNV in water samples, shrimp seedlings, bait, etc., and is used for the etiological diagnosis of suspected infected persons.

 

Principle:

 

This kit is chosen infectious subcutaneous and hematopoietic organ necrosis virus (IHHNV) specific fragment to design primers probe probe, the probe with primer amplification area in the middle of a specific binding DNA template, in the process of PCR extension, the circumscribed Taq polymerase enzymes from the 5 'end of fluorescent groups on the probe will be cut down, make it free in the reaction system, Thus, the 3 '-end quenched group was separated and fluorescence was emitted, which was detected by fluorescence PCR instrument to realize the detection of infectious subcutaneous and hematopoietic necrosis virus (IHHNV) in a completely closed reaction system.

 

Main components:

 

ComponentSpecification
PCR reaction liquid 24 tubes
Positive control1 tube
Negative control1 tube
Note: components in different batches of kits should not be interchangeable.

 

Sample requirements:

 

1. Sample type

Seedlings: parent, shrimp eggs, seedlings, coarse standard products

Pond: bottom mud, excrement, water body

Bait: live bait, chilled bait and non-high-temperature treated bait samples

2. Transport

Sample collection and transportation shall be carried out in accordance with NY/T 541. Samples shall be transported to the laboratory as soon as possible under refrigerated conditions to avoid repeated freeze-thaw.

3. Sample preservation

It can be stored for 24 hours at 2℃~8℃, and can be stored for a long time at -20℃.

 

Using Micgene 242/244/244IVD and ABI QuantStudio5:

 

procedures are set as below:

Steps

Cycle number

TemperatureTime
1195℃3min
24595℃10s
60℃

30sCollect the fluorescent

FAM was selected for detection channel.

 

The results of interpretation

ResultsCt Value
PositiveCt≤45
NegativeNo Ct value or Ct value≥45
Gray area40<Ct≤45

Note: The result is judged to be grey area, which needs to be rechecked. If the recheck result is Ct value < 45 is considered as positive, and no Ct value or Ct value ≥45 is considered as negative.

 

Result analysis:

Automatic analysis using instrumentation

 

Quality control:

Negative control: No obvious amplification curve of FAM detection channel;

Positive control: FAM channel showed obvious amplification curve, Ct value ≤30;

 

The above conditions are met simultaneously in the same experiment; otherwise, the experiment is invalid and re-testing is required.

 

 

Hot Products

BioPure series nucleic acid extractor is a set of highly automatic equipment integrating nucleic ...
Nucleic acid extraction96-well Nucleic Acid Extractor: Features -Easy operation 13-inch color touch ...
BioPure series nucleic acid extractor is a set of highly automatic equipment integrating nucleic ...
Nucleic acid extraction 32-well Nucleic Acid Extractor: Features -Easy operation 13-inch color touch ...
Infectious hypodermal and hematopoietic necrosis virus(IHHNV) Nucleic Acid Detection Kits (PCR...
Along with the frequent outbreaks, a variety of diseases as well as the farmers breeding consciousne...