GF6018A 20A 1050V Multifunction Electric Ammeter and Voltmeter calibrator for clamp meter calibration

 


The equipment of clamp type multimeter calibrator is compliance with national verification regulation: JJG124-2005 “Ammeter, voltmeter, power meter and resistance meter verification rules” and related national standards. 0.05 single-phase AC/DC standard source, can verify magnitude 0.2 and the following AC/DC voltmeter, ammeter, frequency meter, resistance meter. The source signals by using DSP and 16-bit high-speed analog-to-digital converters of controllable sine wave, the distorted wave signal source.The equipment comes with RS-232 interface, can be connected with PC to a semi-automatic verification system. It has precise interface and multi-functions including verification, storage and query.The equipment adopts color LCD screen, chart character display is clear, with high precision, stable and reliable, convenient operation and flexible characteristics.
 
Calibration of Voltmeter, Ammeter & Wattmeter using Potentiometer The calibration is the process of checking the accuracy of the result by comparing it with the standard value. In other words, calibration checks the correctness of the instrument by comparing it with the reference standard.
Calibration unit has a Voltmeter and an Ammeter, which can be Calibrated by Potentiometer. A Potentiometer is an instrument designed to measure an unknown voltage by comparing it with a known voltage. It has a high degree of accuracy.
The best method of calibration is to measure the true value of current or voltage in the circuit and compare it with the value measured by the given ammeter or voltmeter. The true value of current or voltage in the circuit can be measured up to a fair grade of accuracy by using a potentiometer.
 
Theory


Calibration of multimeter in measurement technology and metrology is the comparison of measurement values delivered by a device under test with those of a calibration standard of known accuracy. Such a standard could be another measurement device of known accuracy, a device generating the quantity to be measured such as a voltage, or a physical artefact, such as a metre ruler.
The outcome of the comparison can result in no significant error being noted on the device under test, a significant error being noted but no adjustment made, or an adjustment made to correct the error to an acceptable level. Strictly speaking, the term calibration means just the act of comparison, and does not include any subsequent adjustment.
The calibration standard of multimeter and clamp meter is normally traceable to a national standard held by a National Metrological Institute.

 

FEATURES

 

  • It can be semi-automatic or manual inspection of analog multi meter, clamp type current meter, all kinds of electric meter, (voltmeter, ammeter, frequency meters, resistance meter) basic error

  • The equipment can be generated with 2-31 harmonic wave, harmonic number, order, amplitude and harmonic phase of fundamental wave can be programmed

  • Power amplifier working frequency range is 40Hz-1KHZ, has a good linear. Current amplifier is constant current source; voltage amplifier is constant voltage source. Because of light weight, this device is more suitable for field use

  • With RS-232 interface, this device is controlled by PC through the PC software (optional), can be automatic or manual inspection, and the results are processing and management

  • Has a non-volatile memory, it can store 500 pieces of the raw data of tested meter, for access and upload

PARAMETERS

 

Electrical parameters  

Power supply

Single phase AC 220V±10% or 110V±10%, 50/60HzAC

AC Voltage output  

range

200mV, 400mV, 2000mV, 4000mV,
20V, 100V, 200V, 500V, 1000V

Range 20V-1000V

Accuracy

±(0.03%RD+0.02%FS)

Stability

≤0.01% FS /60s

Range 200mV-4000mV

Accuracy

±(0.3%RD+0.2%FS)

Stability

≤0.04% FS /60s

Load capacity

20VA

Waveform distortion

≤0.3%

Adjustable Range

0-120% FS (1000V not included)

Fineness

5×10-5

AC Current Output  

Range

2mA, 4mA, 20mA, 40mA, 0.2A, 0.5A, 2A, 5A, 10A, 20A

Range 0.2A-20A

Accuracy

±(0.03%RD+0.02%FS)

Stability

≤0.01% FS /60s

Range 2mA-40mA

Accuracy

±(0.3%RD+0.2%FS)

Stability

≤0.04% FS /60s

Load capacity

20VA

Waveform distortion

≤0.3%

Adjustable Range

0-120% FS

Fineness

5×10-5

Frequency Output  

Range

45-65Hz

Fineness

0.001Hz

Accuracy

0.01Hz

Harmonic Wave  

Times

2-31

Extent

0-20%

Each harmonic phase fineness

0.01·N (N is harmonic order)

DC voltage output  

Range

200mV, 400mV, 2000mV, 4000mV,
20V, 40V, 100V, 200V, 400V, 1000V

Load capacity

20W

Adjustable range

0-110% FS (1000V not included)

Adjustable fineness

5×10-5

Stability

≤0.01% FS /60s (peak-peak value)

Accuracy

±(0.03%RD+0.02%FS)

Ripple wave

≤1% FS

DC current output  

Range

20uA, 40uA, 200uA, 400uA, 2mA, 4mA,
20mA, 40mA, 0.2A, 1A, 2.5A, 5A, 10A, 20A

Load capacity

20W

Adjustable Range

0-110% FS (1000V not included)

Adjustable Fineness

5×10-5

Stability

≤0.01% FS /60s (peak-peak value)

Accuracy

±(0.06%RD+0.04%FS)

Ripple wave

≤1% FS

Resistance box  

Range

0-200Ω, 200-400Ω, 0-2kΩ, 2kΩ-4kΩ, 0-20kΩ, 20kΩ-40kΩ,
0-200kΩ, 200kΩ-400kΩ, 0-2MΩ, 2MΩ-4MΩ

Fineness

0.02%

Range 0-2kΩ, 0-20kΩ, 0-200kΩ  

Range 0-2kΩ

Adjustable Range

0-100%FS

Accuracy

±(0.12%RD+0.08%FS)

Range 0-200Ω,0-2MΩ

Adjustable Range

0-100% FS

Accuracy

±(0.3%RD+0.2%FS)

Range 2kΩ-4kΩ, 20kΩ-40kΩ, 200kΩ-400kΩ  

Range 2kΩ-4kΩ

Adjustable Range

5%-100%FS

Accuracy

±(0.12%RD+0.08%FS)

Range 200-400Ω,
2MΩ-4MΩ

Adjustable Range

5%-100% FS

Accuracy

±(0.3%RD+0.2%FS)

Clamp meter range  

200A

Accuracy

±0.2%

600A

Accuracy

±0.5%

1000A

Accuracy

±1%

Dot frequency square wave output 

Range

1kHz, 5kHz, 10kHz, 50kHz, 100kHz, 200kHz, 1000kHz

Square wave amplitude

4.5V

Setting accuracy

5×10-5

Mechanical parameters  

Dimension (L×W×H) (mm)

440x360x160

Weight (kg)

12

Environmental conditions 

Operating temperature

20°C±10°C

Relative humidity

≤85%RH

 

Photo

 

Hot Products

LZCG530-10 Protection relay device using 5P20 50/1A Zero Sequence Current Transformer Zero Sequence ...
LZCG530-10 Zero Phase Current Transformer, Ring net cabinet 5P10 Ratio 100A / 1A Zero Sequence ...
LZCK2500-10 120mm 50/5a Resin Casting Split core Protection Zero Sequence Current Transformer 5P20 ...
LZCK-55 GFUVE 100A/5A IP67 Outdoor installation Split Core Clamp On Current Transformer Clamp-On ...
LMCK185-10 2000A /5A 0.5s IEC60044-1 IP67 Outdoor split core current transformer manufacturers ...
LZCK-55 Outdoors Split Core Current Transformer Description LZCK series current transformers are ...