Home Companies ZHANGJIAGANG MEDPHARM MACHINERY LTD.

High Pressure Muti Cycle Selection CSSD Sterilizer For Hospital B-D Test

ZHANGJIAGANG MEDPHARM MACHINERY LTD.

Contact Us

[China] country

Trade Verify

Address: B631, HUAFANG INTERNATIONAL MANSION, No.178 CHENGBEI STR., ZHANGJIAGANG, JIANGSU, PRC, 215600

Contact name:Juliet

Inquir Now

ZHANGJIAGANG MEDPHARM MACHINERY LTD.

Verified Suppliers
  • Trust
    Seal
  • Verified
    Supplier
  • Credit
    Check
  • Capability
    Assessment

High Pressure Muti Cycle Selection CSSD Sterilizer For Hospital B-D Test

Country/Region china
City & Province jiangsu
Categories Other Beauty Equipment
InquireNow

Product Details

           

High Pressure Muti Cycle Selection CSSD Sterilizer For Hospital B-D Test

 

Sterilization Cycles

 

Cycle No.

Cycle Name

Further Details

1

Instruments/Tools

3 times Vacuum, expose temperature 132℃, expose time 8 min, drying time 5 min.

2

Clothing

3 times Vacuum, expose temperature 121℃, expose time 15 min, drying time 25 min.

3

Liquid

15 min exchange, expose temperature 121℃, expose time 15 min, cooling time 25 min.

4

B-D Test

3 times Vacuum, expose temperature 134℃, expose time 4 min, drying time 3 min.

5

Vacuum Leak Test

Test the chamber connecting parts’ tightness effect. It’s important to ensure the sterilization effect.

6

User Programmable

 All preset parameters can be resettable; Multiple customized programs can be set and stored by user.

 

 

 

 

 

 

            

 

                                                                                                                                                                    

Description

 

Door Safety

 

  • The autoclave doors are designed with a number of independent mechanical and digital safety features.
  • A safety device prevents the operator from opening the door when chamber is pressurized
  • Steam will not enter the chamber when the door is open.
  • A cycle cannot start if the door is open or not properly locked.
  • The door cannot unlock until chamber pressure reaches room pressure.
  • Sliding Door Safety - sliding door progress will automatically stop if an obstruction is detected.
  • Double Door Safety – interlocks prevent both doors from being opened simultaneously.

 

General Safety Features

 

  • Double Independent Monitoring: The combined electronic and mechanical monitoring ensures that the operator has two independent means to monitor pressure.
  • Safety Valves: If the pressure exceeds the allowed limit the safety valves will discharge.
  • Built-in Steam Generator Safety: A water level monitoring system maintains a constant water level and ensures safe operation of the heaters.
  • Emergency shut-off: Easily accessible emergency switches for immediate cycle shut-off.

 

Competitive Advantage

 

  1. During Sterilization a Microprocessor / PLC controls temperature, Pressure, Steam Injection,
  2. Sterilization time, Drying and all other critical time based parameters.
  3. Important sterilization parameters like Phase Status, Current Values & set values, continuously displayed at all times on Controller Screen.
  4. Continuous display of Temperature and Pressure On-line allows user to be aware of Cycle status at all times.
  5. Error Code and alarm is generated and cycle stops automatically in case any error occurs.
  6. Computer Screen displays Error, possible cause and remedy.
  7. The advanced SCADA software provided stores detailed records of Process- Data for every cycle.
  8. Reports are available in Data, Summary and Graph formats for every batch of goods sterilized.

 

 

 

 

Brochure load down from here

Medpharm medical large steam autoclave brochure.pdf

Hot Products

High Vacuum Degree Sterilizers Autoclave With Low Noise Water-ring Vacuum Pump For Silence ...
Rectangular Chamber With Smooth Round Corners And Jacket Medical Instrument Sterilizer Machine Model ...
Liquid Cycle With Pressure Ballasting Steam Sterilization Using Autoclave Feature 1. Hinged door ...
High Pressure High Vacuum Autoclave Sterilizers Ensuring Perfect Sterilization For Hospital And ...
Overpressure Protection Autoclave and Sterilizers With Safety Door System Main parameter PLC based ...
Overpressure protection autoclave and sterilizers with safety door system Main parameter: PLC based ...