Sinuo Testing Equipment Co. , Limited |
Verified Suppliers
|
|
IEC 62368-1 Clause 4.1.4 Programmable Constant Temperature And Humidity Test Chamber
Product information:
This programmable temperature chamber is required by IEC 62368-1:2018 clause 4.1.4, Y4.4, Y6.2, etc.
This test chamber is used for cold, heat, humidity and dryness testing of raw materials and devices such as electronics, electricians, electrical appliances, plastics, etc. The purpose is to confirm the products reliability of storage and transportation under high and low temperature alternating heat and humidity.
It conforms to standards:
IEC60068-2-1:2007 “Environmental Testing-Part 2-1: Tests-Test A: Cold, IDT”
IEC60068-2-2:2007 “Environmental Testing-Part 2-1: Tests- Test B-Dry heat, IDT”
IEC60068-2-78:2012 “Environmental Testing-Part 2-78: Tests-Test Cab: Damp heat, steady state, IDT”
IEC60068-2-30:2005 “Environmental Testing-Part 2-30: Tests-Test Db: Damp heat, cyclic (12h+12h cycle), IDT”
Features:
1. The test system has advanced and reasonable structural design,
standardized manufacturing process, beautiful appearance and
generous appearance.
2. The main functional components of the test chamber are world
famous branded ones. The test chamber has best control of noise and
energy saving.
3. The matching of components and assembly is good. The main
functional components are imported from the internationally branded
companies, which improves the safety and reliability of the test
chamber, and can ensure the long-term, high-frequency usage
purposes.
4. The equipment has good operability, maintainability, good
temperature stability and durability, good safety performance. It does not pollute the
environment and endanger human health.
Technical parameters:
Applied standard | IEC 62368-1 |
Applied clause | 4.1.4 |
Sample limitation | This test chamber prohibits: Testing or storage of flammable, explosive, and volatile substance samples Testing or storage of samples of corrosive substances Testing or storage of biological samples Testing or storage of strong electromagnetic emission source samples Testing or storage of radioactive material samples Test or storage of highly toxic substance samples Testing or storage of samples that may produce flammable, explosive, volatile, highly toxic, corrosive and radioactive materials during testing or storage |
Volume, size and weight | |
Nominal content volume | (2880L) |
Inner chamber size | W 2400×D 800×H 1500 (mm) |
External dimensions | Approx. W3100×D1320×H2100 (mm) |
Weight | Approx. (1500) KG |
Performance | |
Temperature range | -40 ℃~+85℃ (C type) Air-cooled type |
Humidity range | (30 ~ 95)% RH (Refer to temperature and humidity controllable range chart, no active wet and heat load) Temperature and humidity controllable range chart |
Temperature fluctuation | ≤0.5℃ (The temperature fluctuation is half of the difference between the measured maximum temperature and the minimum temperature at the center point) |
Temperature uniformity | ≤2.0℃ (Temperature uniformity is the arithmetic mean of the difference between the highest and lowest temperature measured in each test) |
Humidity deviation | ≤ 2%R.h (>75%), 3%R.h. (<75%) |
Heating time | RT~ 100℃ (2℃-3℃)/min |
Cooling time | RT~ -40℃(0.7℃-1℃)/min |
Working noise
| 80dB(A) (1m in front of the gate, 1.2m above the ground, in free space) |
Control System | |
Controller | 7-inch color touch screen temperature controller, English display |
Operation mode | Fixed value/program mode: users can perform programming test and constant test according to test requirements and needs |
Setting method | Set on the touch screen |
Input | Thermal resistance signal |
Output | Multi-point temperature and humidity signal output |
Control method | Control and adjust the temperature and humidity in the test
chamber, and automatically determine the working status of each
executive elements (such as compressor, fan, heater, water circuit,
etc.) PID action and automatic calculation function Ratio band (PB): 0.1%~999.9% Integration time (IT): SR-92/SR-93 1~3600 seconds Differential time (DT): SR-92/SR-93 1~1200 seconds |
Setting range | Temperature: -100~+300℃, humidity: 0~100%RH |
Control precision | Temperature: 0.1°C, humidity: 0.1%RH, time 0.1 minutes |