Home Companies Guangzhou Kingrise Enterprises Co., Ltd.

Lab Use Three Phase Meter Test Bench , Meter Test System High Precision CT / PT

Guangzhou Kingrise Enterprises Co., Ltd.

Contact Us

[China] country

Trade Verify

Address: Rm 210,No 69 Guangpu Rd(W), Huangpu Dist. Guangzhou, 510663 CHINA

Contact name:

Inquir Now

Guangzhou Kingrise Enterprises Co., Ltd.

Verified Suppliers
  • Trust
    Seal
  • Verified
    Supplier
  • Credit
    Check
  • Capability
    Assessment

Lab Use Three Phase Meter Test Bench , Meter Test System High Precision CT / PT

Country/Region china
City & Province guangzhou guangdong
Categories Switching Power Supply
InquireNow

Product Details

 

 

KE8300B Three phase Meter Test Equipment

 

Accuracy : 0.02 , Used for the test of standard meter and mounted meter with a accuracy class below 0.1, Number of meter position : 3-12

 

Features:

  1. High accuracy: 0.02%
  2. Integrated / separated aluminum alloy structure, light and reliable, flexible choices
  3. Meter positions can be chosen from 3-12
  4. Auto protection of short circuit of voltage, open circuit of current, and overload.
  5. Test of neutral current (optional)
  6. Full automatic, semi automatic and manual operation available
  7. Supporting bar code input, which increase the efficiency of input
  8. Simultaneous test of the meters with same specification, different constants and different classes.
  9. Forward and reverse active energy/ Forward and reverse reactive energy error test
  10. Creeping and starting test
  11. Standard deviation, 24 hour variation test
  12. Timing error test
  13. Influencing quantity test of voltage, frequency, harmonic
  14. Calibration and verification of constant
  15. Register indicating value, register combination error test
  16. Powerful computer management system, easy and user-friendly
  17. Perfect database management makes the storage, searching, and viewing of the data possible, and the data can be converted and exported in other forms

Functions:

  1. Base error test
  2. Forward / reverse active energy error test
  3. Forward / reverse reactive energy error test
  4. Creeping test (no load test)
  5. Starting test
  6. Verification of constant
  7. Balanced and imbalanced load test
  8. Dial test (energy accumulation test)
  9. Error repeatability test
  10. Standard deviation test
  11. 24-hour variation test
  12. Influencing shelf-heating test
  13. Reversed phase sequence test
  14. Influencing quantity test of voltage, frequency, harmonic
  15. Voltage dips and short interruptions test
  16. Register indicating value, register combination error test
  17. Timing error test (optional)

 

Technical Specifications:

 KE8300B-3KE8300B-6KE8300B-12
Number of meter position3612
Accuracy±0.02%
Voltage output
Output range (P-N)3×(24V - 300V)
Output power of voltage3×200VA3×200VA3×400VA
ResolutionBetter than ± 0.001%
StabilityBetter than ± 0.002% / 150s
Load regulation rateBetter than ± 0.01%
Distortion factorBetter than ± 0.05% (linear resistance load)
Harmonics output2nd -21st free programmable
Current output
Output range1mA - 120A
Output power of current3×150VA3×300VA3×600VA
ResolutionBetter than ± 0.001%
StabilityBetter than ± 0.002% / 150s
Load regulation rateBetter than ± 0.01%
Distortion factorBetter than ± 0.05% (linear resistance load)
Harmonics output2nd -21st free programmable
Phase output
Output range0 -- 360°
Resolution± 0.01°
Frequency
Output range45Hz – 65Hz
Resolution± 0.001Hz
Operation condition
Power supply220V±10%,50/60Hz±2Hz
Temperature-10°C -- +40°C
Humidity35% -- 85%RH

 

 

 

 

 

 

 

 

Hot Products

KE8302M-1000 1000A Three Phase Meter Test Equipment KE8302M-1000 is used for testing and calibration ...
KE8300A Three phase Meter Test Equipment Accuracy : 0.01 , Used for the test of standard meter and ...
KE8101 single phase meter test equipment is used for the test of various single phase electric and ...
KE8300M three phase meter test equipment Accuracy : 0.02 , works as the standard of meter factory, ...
KE8300B Three phase Meter Test Equipment Accuracy : 0.02 , Used for the test of standard meter and ...
KE8301 Three Phase Meter Test Equipment Accuracy : 0.05 , Used for the test of three phase CT/PT ...