PN SR 92B370 GE Datex-Ohmeda S5 Patient Monitor Power Supply For Hospital

 

 

About Power Supply

The menu shows the voltage and temperature measured by the display unit power management board (PMB voltage) and the CPU main board (DU voltage). The measurement starts approximately 100 ms after the start. The values ​​in the Average column show the average for the last second, the Min column shows the minimum average, and the Max column shows the maximum average of the voltage and temperature measured during the current power-on period.

The PMB measures the following voltages on the frame drive

 

Raw CC

+15V 14.65….15.35

–15V -14.65 ...- 15.35V

+ 15VD 14.65 ... 16.00V

+ 5V 4.85 ... 5.25V

.

The temperature values ​​(°C) correspond to the internal temperature of the frame unit measured by the PMB.

The numbers on this page are just instructions, not absolute values.

The following voltages are measured by the central processing board of the display unit:

+ 12V 11.50 ... 12.50V

+ 3.3V 3.00 ... 3.60V

VBUS 10.00… 17.00V

 

The temperature (°C) corresponds to the internal temperature of the display unit, which is measured by the central processing board.

The MPS input current is the current supplied to the module's power supply. The current must be <4A.

The AC/DC capacity used is the capacity information that the PMB receives from the AC/DC unit.

 

AC/DC voltage is the AC/DC power supply output voltage measured by the PMB.

 

Specification

 

Part NoSR 92B370
Rated voltage range and frequencies100 to 240 VAC 50/60 Hz (limits 90-264 VAC)
Stability±10 % of nominal voltage
Power consumption140 VA (max)
Fuses2x2A, T/SLOW, 250 V
GroundingHospital grade
InterruptabilityData memory and alarm settings are saved up to 15 minutes after monitor standby

 

 

The storage section receives power (+5V, +12V) from the CPB. PCMCIA card programming voltage is +12V. The card programming voltage is controlled by the interface controller.

 

 

Display Unit DC/DC power supply

 

 

The universal AC/DC power supply supplies the compact S/5 monitor with 15.5 V isolated power supply.

The power supply is attached to the chassis with four screws and a D25 power connector.

Cooling air for the AC/DC unit comes from the chassis unit fan through the cooling vents at the rear of the chassis unit. The power supply generates a charge signal that is read by the PMB card in the frame unit.

The AC/DC power supply is protected against overvoltages. When the output voltage exceeds 17V, the protection will be activated and the AC/DC power supply will not resume normal operation until the power cord is momentarily disconnected.

 

This DC/DC converter board converts primary voltages between +10 and +18 VDC to +3.3V, +5V and +12V secondary voltages. All output voltages have a common ground.

The DC/DC power supply is powered by the Vbus voltage produced by the frame unit AC/DC converter (power supply). In the event of a mains power failure, the DC/DC power supply is powered by batteries.

The turn-on and turn-off sequence of the DC/DC power supply is controlled by the logic on the DC/DC board and the CPU board software. If there is overvoltage on the output line, the converter will automatically shut down. The display unit DC/DC converter also provides a display unit fan.

Hot Products

GE Dash3000 Dash4000 Dash5000 patient monitor general electric 801536-002 421263 circuit board ...
Used Nonin Model 7500 Pulse Oximeter, Hospital Medical Monitoring Devices ► Brief Introduction The ...
Hospital Medical Monitoring Devices Mindray BC-2800 Auto Hematology Analyzer ► Brief Introduction ...
Hospital Medical Equipment Philipilip Glodway G30 Patient Monitor ►Brief Introduction Goldway G30 ...
P/N M3535-62301 philip MRX Defibrillator Spare Parts NBP Module NBP Module The NBP module performs ...
051-000633-00 Mindray MEC1200 MEC2000 Patient Monitor Mainboard Motherboard The main control of the ...