Home Companies China World Technology Medical Equipment Service Group

X-6000 X6000 Stryker Endoscopy Xenon Light Source 220-185-000

China World Technology Medical Equipment Service Group

Contact Us

[China] country

Trade Verify

Address: TianHe District,GuangZhou,GuangDong

Contact name:Kiara

Inquir Now

China World Technology Medical Equipment Service Group

Verified Suppliers
  • Trust
    Seal
  • Verified
    Supplier
  • Credit
    Check
  • Capability
    Assessment

X-6000 X6000 Stryker Endoscopy Xenon Light Source 220-185-000

Country/Region china
City & Province guangzhou guangdong
Categories Laser Equipment Parts
InquireNow

Product Details

 

Hospital Medical Equipment Stryker Endoscopy X-6000 X6000 Xenon Light Source, 220-185-000

 

►Brief Introduction

The Stryker Endoscopy X-6000 Xenon Light Source is a high performance device designed for use with Stryker Endoscopy medical vidoe cameras to provide outstanding illumination of the surgical site across all endoscopic applications. With the proper light cable and adapters, it can be connected to any filexible or righd endoscope.

The X-6000 is equipped with a special safety feature that when using an ESST cable, will prevent accidental burning caused by an unattended light cable. Electronic Scope Sensing Technology senses when the scope and the light cable are separated and places the light source in standby.

A Stryker Endoscopy X-6000 Xenon Light Source consists of the light source console unit, power cord, lamp, and video cable. Replacement or spare lamps can be purchased separately. The X-6000 Xenon Light Source is compatible with all Stryker Light Cables.

 

 

Functions

  1. Power – Turns unit prower “On” or “Off”.
  2. Hour Meter – Displays cumulative number of hours lamp have operated since last lamp change.
  3. Brightness Display LED – Indication (0-100%) of light intensity level.
  4. Brightness Control Slide Bar – Used to increase or decrease light intensity.
  5. Standby – Momentary switch to reduce lamp intensity to a low level for endoscope removal or direct (non-video) viewing through endoscope. STANDBY is the default setting when the light source is turned on. The unit is in Standby when the Standby word and symbol are lit.
  6. Shutter – Toggle switch used to turn the automatic Shutter on and off. The Shutter is on when the Shutter word and symbol are lit.
  7. Cable – Port for light source end of fiber optic cable. When no cable is presend, the lamp automatically turns to save bulb hours.
  8. In/Out – Opens fiber optic cable holder.

 

No Light Output Troubleshooting

 

1Check the power cord to ensure it is plugged into the AC inlet on the back of the unit and that the power cord is also plugged into a power source.
2Check that the front panel POWER switch is switched "ON".
3Check the fuse compartment on the back of the unit to ensure that all fuses are operating.
4Check the lamp housing to make sure a lamp is properly seated in the lamp housing.
5Check the lamp to ensure it is in operating condition. Replace if necessary.
6Check to see that a fiber optic light cable is seated correctly in the CABLE port.
7Ensure bulb access door is completely shut.

 

 

 

Thank you for your time and interest in Stryker Endoscopy X-6000 X6000 Xenon Light Source, 220-185-000. All through the way we seek for providing perfect service to the satisfaction of customers. If you are interested in our products, welcome to place your orders.

Hot Products

Hospital Medical Equipment Stryker Endoscopy X-6000 X6000 Xenon Light Source, 220-185-000 ►Brief ...
Refurbished GE DASH2000 Patient Monitor Standard Parameter For Sale GE Dash2000 device is designed ...
Hospital Medical Equipment Mindray BeneHeart R12 ElectrocardiograPhilipilip ECG Machine ► Brief ...
004-0004-004 2016792-002 REV A GE Dash4000 patient monitor display assembly 10.4inch Display ...
2026653-019 GE Dash3000 Dash4000 Dash5000 Patient Monitor Battery Door Battery door features High ...
P/N 2019271-001 GE Dash3000 Dash4000 Dash5000 Patient Monitor Display Flex Flat Cable About company ...