Home Companies China World Technology Medical Equipment Service Group

P/N 453564055981 VM6 Patient Monitor Parts Mainboard Motherboard

China World Technology Medical Equipment Service Group

Contact Us

[China] country

Trade Verify

Address: TianHe District,GuangZhou,GuangDong

Contact name:Kiara

Inquir Now

China World Technology Medical Equipment Service Group

Verified Suppliers
  • Trust
    Seal
  • Verified
    Supplier
  • Credit
    Check
  • Capability
    Assessment

P/N 453564055981 VM6 Patient Monitor Parts Mainboard Motherboard

Country/Region china
City & Province guangzhou guangdong
Categories Motherboards
InquireNow

Product Details

P/N 453564055981 Philipilip VM6 Patient Monitor Mainboard Motherboard New Version

 

Main PCB

 

The main PCB contains these components:

• Intel X_Scale µProcessor

– 32 Mb flash memory

– 64 Mb SDRAM

• LAN controller – 10/100T

• Power management control

– Voltage regulator

– Smart battery charger that monitors status of battery

• USB – 1.1 compliant

• LCD drivers

• Front end interface circuitry

• Fan control

• LVDS buffer for LCD interface

• Power distribution and monitoring

• VGA DAC

• Audio

• NBP pump control circuitry

• Connectors

– etCO2 (wire to etCO2)

– DC input (wire to AC/DC module)

– Speaker – LCD – LCD backlight

– Control board – B2B: patient daughter board

– B2B: VGA daughter board

– B2B: I/O daughter board

– B2B: recorder board

 

Item name

Philipilip VM6 patient monitor mainboard new version 453564055981

 

BrandPhilipilip
Model453564055981
Use placeHospital Devices; Medical Care; ICU;
Deliver time3-5 days
Warranty90 days

 

 

 

Removing the Main PCB

 

Cautions

After replacing the main PCB, you must reset the monitor serial number, reconfigure the system, and recalibrate the monitor

 

1.Separate the front and rear monitor case assemblies

 

2.Remove the main board assembly.

 

3.Remove the front end assembly

 

NOTE - Before reassembling the monitor, please note the calibration values. These values are needed to reset the serial number and calibrate the monitor

 

4.Remove the five Torx T10 screws from the main board.

5.Disconnect all of the connectors and NBP tubes from the main board

 

6.Remove the main board from the assembly

 

7.If the version of the motherboard you are installing does not require the LCD adapter board, remove the adapter board from the LCD frame and connect the LCD directly to the motherboard.

 

 

8.After replacing the motherboard and reassembling the monitor, you must do

Follow the procedure below before use:

• "Reset Serial Number"

• "Set System Configuration"

• "Calibrate IBP (SureSigns VM6 and VM8 SE)"

• "Calibrate SureSigns VM6 and VM8 SE Temperature"

 

 

 

 

Reinstalling the Main Board Assembly

 

1.Slide the board assembly into the rear case (guiding the power cord retaining clip through the rear cover) and ensuring enough clearance for the speaker and fan cables.

 

2.Push the assembly until it snaps into place in the case.

3.Replace the four Torx T10 screws and washers on the bottom of the rear case to ensure that the main board assembly is aligned correctly in the case.

 

Main Board Versions

Depending on the version of the motherboard in the monitor, the LCD may require a breakout board. Use the image below to determine the board version. If the board has a connector in the area shown below, an adapter board must be installed. If there are two connectors in this area, you don't need a breakout board and you can connect the LCD cable directly to the motherboard.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Hot Products

Hospital Medical Equipment Stryker Endoscopy X-6000 X6000 Xenon Light Source, 220-185-000 ►Brief ...
Refurbished GE DASH2000 Patient Monitor Standard Parameter For Sale GE Dash2000 device is designed ...
Hospital Medical Equipment Mindray BeneHeart R12 ElectrocardiograPhilipilip ECG Machine ► Brief ...
004-0004-004 2016792-002 REV A GE Dash4000 patient monitor display assembly 10.4inch Display ...
2026653-019 GE Dash3000 Dash4000 Dash5000 Patient Monitor Battery Door Battery door features High ...
P/N 2019271-001 GE Dash3000 Dash4000 Dash5000 Patient Monitor Display Flex Flat Cable About company ...