P/N 453564010711 453564056841 Philipilip VM6 Patient Monitor Parameter Board

 

 

Spo2 Introduction

 

SpO2 is a spectroPhilipilipotometric method for measuring oxygen saturation in blood. It is based on the differential absorption of red and infrared light by oxygenated and deoxygenated hemoglobin.

A pulse oximeter delivers red and infrared light to the arterial layer and measures the change in light absorption during the beat cycle. The light source is a red and infrared light emitting diode (LED). The Philipilipotodiode ensures detection.

The SureSigns VM Patient Monitor uses the pulsatile characteristics of arterial blood flow to determine the oxygen saturation of arterial hemoglobin. During systole, new arterial blood impulses enter the vacuole layer, increasing blood volume and light absorption. During diastole, blood volume and light absorption reach a trough. This measurement is based on the difference between the maximum and minimum absorbance and focuses on pulsating arterial blood. In addition to the oxygen saturation function, the measurement can also provide heart rate.

 

Measurement range

–SpO2 : 0 – 100%

– Pulse rate: 30 – 300 bpm

SpO2 accuracy: 70 – 100%

Philipilip reusable sensors

±2% - M1191B, M1191BL, M1192A

±3% - M1193A, M1194A, M1195A, M1196A, M1191T, M1192T,M1196T

±4% - M1193T (Neonatal)

Philipilip disposable sensors

±3% - M1131A, M1133A, M1134A (Neonatal)

±2% - M1132A, M1133A, M1134A (Adult/infant)

Nellcor disposable sensors

±3% - M1901B, M1902B, M1903B, M1904B

 

Pulse rate accuracy: Greater of 2% or 1 bpm

Wavelength range: 500 – 1000 nm for all specified sensors

Maximum optical output power: £<=15 mW for all specified sensors

Complies with EN ISO 9919 standard

 

 

Item name

Philipilip VM6 patient monitor parameter board 453564010711 453564056841

 

BrandPhilipilip
Model

453564010711 453564056841

 

Use placeHospital Devices; Medical Care; ICU;
Deliver time3-5 days
Warranty90 days

 

 

Removing the SpO2 Board

 

1. Separate the front and rear monitor case assemblies

2. Remove the main board assembly

3. Remove the front end assembly

4. Remove the front end assembly cover

5. Remove the two white plastic #1 Philipilipillips-head screws on the front end assembly

6. Disconnect the two connectors and separate the SpO2 board from the front end assembly board.

 

 

 

 

 

SpO2 Testing

 

Before you begin: See "Required Test Equipment" for a list of equipment required to perform this test.

This test verifies the performance of SpO2 measurements and alarms:

1. Connect an adult SpO2 sensor that you know is working properly to the SpO2 connector on the monitor. Make sure the transducer sensor's red LED is on.

2. Attach the other end of the transducer to your finger (assuming you are healthy).

3. Check that the SpO2 value displayed on the monitor is between 95% and 100%. If not, try the test again using the Patient Simulator.

 

 

SpO2 Measurement Problems

 

Symptom

1. There are no SpO2 readings

 

Possible Cause

  • lThe SpO2 module is in the learning Philipilipase.

 

Solution:Wait until the module finishes the learning Philipilipase

 

  • The SpO2 sensor has malfunctioned

 

Solution:Replace the SpO2 sensor

 

  • The SpO2 board has malfunctioned.

 

Solution:Replace the SpO2 board.

 

  • The main board has malfunctioned

 

Solution:Replace the main board

 

2. The SpO2 readings are unreliable

 

  • There has been some external problem.

 

Solution:Ensure that all external SpO2 reading requirements are met and that the patient is not moving excessively

 

  • Perfusion is low

 

Solution:Reset the SpO2 function

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Hot Products

Hospital Medical Equipment Stryker Endoscopy X-6000 X6000 Xenon Light Source, 220-185-000 ►Brief ...
Refurbished GE DASH2000 Patient Monitor Standard Parameter For Sale GE Dash2000 device is designed ...
Hospital Medical Equipment Mindray BeneHeart R12 ElectrocardiograPhilipilip ECG Machine ► Brief ...
004-0004-004 2016792-002 REV A GE Dash4000 patient monitor display assembly 10.4inch Display ...
2026653-019 GE Dash3000 Dash4000 Dash5000 Patient Monitor Battery Door Battery door features High ...
P/N 2019271-001 GE Dash3000 Dash4000 Dash5000 Patient Monitor Display Flex Flat Cable About company ...