Home Companies China World Technology Medical Equipment Service Group

V60 Ventilator Gas Delivery Air Oxygen Flow Sensor Assembly PN 1033547

China World Technology Medical Equipment Service Group

Contact Us

[China] country

Trade Verify

Address: TianHe District,GuangZhou,GuangDong

Contact name:Kiara

Inquir Now

China World Technology Medical Equipment Service Group

Verified Suppliers
  • Trust
    Seal
  • Verified
    Supplier
  • Credit
    Check
  • Capability
    Assessment

V60 Ventilator Gas Delivery Air Oxygen Flow Sensor Assembly PN 1033547

Country/Region china
City & Province guangzhou guangdong
InquireNow

Product Details

Philipilip Respironics V60 Ventilator Gas delivery Air and Oxygen Flow Sensor Assembly Parts Used PN 1033547

 

 

 

► Brief Introduction

The ventilator delivers gas to the patient through a main flow (inspiratory) bacteria filter, a single-limb patient breathing circuit, a humidification device (optional), and a patient interface such as a mask or ET tube. A pressure tap proximal to the patient is used to monitor patient pressure. The internal exhalation port continually clears gas from the ventilator airway to ensure delivery of an accurate oxygen mixture.

Oxygen Solenoid Valve

The oxygen solenoid valve and valve driver circuitry control the flow of oxygen according to the set O2 and flow. The oxygen solenoid valve is closed when there is a loss of power or system reset.

Oxygen Flow Sensor

The oxygen flow sensor measures a subset (bypass flow) of the total flow. These measurements are interpolated according to constants that are calculated during GDS calibration. Together, the oxygen solenoid valve and flow sensor provide closed-loop control for delivered oxygen flow.

  • Calibration of selected gas (air or oxygen).
  • Provide a signal source for oxygen and air flow signals. Convert the manifold gas flow into an analog signal, which is sent to DA PCBA for filtration and conversion.
  • The airflow sensor measures flows from -240 to 240 SLPM. The oxygen flow sensor measures flows from 0 to 240 SLPM.
  • Includes EEPROM for calibration data, panel identification information, and operating hours for each flow sensor.

 

Specifications

IPAP4 to 40 cm H₂O
CPAP4 to 25 cm H₂O
EPAP4 to 25 cm H₂O
Max P (PPV maximum pressure limit)5 to 40 cm H₂O
Max P (AVAPS maximum IPAP)6 to 40 cm H₂O
Min P (AVAPS minimum IPAP)5 to 30 cm H₂O
Inspiratory time0.30 to 3.00 s
Oxygen percentage21 to 100 %

 

 

Pneumatics

The pneumatic subsystem delivers and monitors pressurized gas to the patient in response to commands from the CPU subsystem. The pneumatic subsystem includes the:

  • Manifold
  • Blower
  • Oxygen solenoid valve
  • Air and oxygen flow sensors
  • Pressure transducers
  • Solenoid valves
  • Motor controller (MC) PCBA
  • Data acquisition (DA) PCBA

 

► Installation

Follow these steps to remove the air and oxygen flow sensor assembly. Reverse to install.

  1. Remove GDS.
  2. Disconnect the flow sensor cable from the DA PCBA and flow sensor assembly.
  3. Remove the screws that attach the DA PCBA to the flow sensor assembly.
  4. Remove screws that attach the flow sensor assembly to the manifold block.

When installing a new flow sensor assembly, install the new o-rings.

 

 

 

Please kindly contact us, if you are interested in Philipilip Respironics V60 Ventilator Gas delivery Air and Oxygen Flow Sensor Assembly Parts Used PN 1033547, Thank you!

 

Hot Products

Hospital Medical Equipment Used Patient Monitor Edan SE-2003 SE-2012 Holter System ► Brief ...
Hospital Medical Equipment Stryker Endoscopy X-6000 X6000 Xenon Light Source, 220-185-000 ►Brief ...
Refurbished GE DASH2000 Patient Monitor Standard Parameter For Sale GE Dash2000 device is designed ...
Hospital Medical Equipment Mindray BeneHeart R12 ElectrocardiograPhilipilip ECG Machine ► Brief ...
004-0004-004 2016792-002 REV A GE Dash4000 patient monitor display assembly 10.4inch Display ...
2026653-019 GE Dash3000 Dash4000 Dash5000 Patient Monitor Battery Door Battery door features High ...