1. How to use defibrillator during treatment
1. Prepare all the supplies next to the bed and turn on the power.
2. Expose the patient's chest and establish ECG monitoring if
necessary.
3. Determine the patient’s arrhythmia type.
4. Apply conductive glue evenly to the electrode plate.
5. Choose the appropriate energy: synchronous is generally 70-100J,
non-synchronous is generally 200-360J (for adults, single-phase is
200J for the first time, 200~300J for the second time, 360J for the
third time; biphasic is 150J-150J-200J).
6. Charging: Place the electrode plate at the appropriate location
(the bottom of the second intercostal space on the right edge of
the sternum: the fifth intercostal space and the apex of the heart
on the left front axillary line). The distance between the two
electrode plates is greater than 10cm; loudly ask other personnel
to leave the patient and hospital bed. .
7. Press the discharge buttons under the two electrode plates with
both hands at the same time. (Electrode plate adhesion strength
10-12kg)
8. Observe the patient’s electrocardiogram changes.
9. If ventricular fibrillation/ventricular flutter (pulseless
ventricular tachycardia) continues to occur, recharge immediately
and repeat the steps.
10. After the operation is completed, return the energy switch to
the zero position.
11. Clean the skin and place the patient.
12. Monitor heart rate and rhythm, and take medication as directed
by your doctor.
13. Record.
14. Final treatment.
1. Check the performance of the defibrillator in time and
charge it in time.
2. The conductive glue should be applied evenly to prevent skin
burns.
3. Energy selection for children: 2J/kg for the first time, 2~4J/kg
for the second time, and 4J/kg for the third time.
4. During discharge defibrillation, pay attention to the insulation
between the patient and other people and objects.
5. For ventricular tachycardia that can clearly distinguish between
QRS and T waves, synchronized electrical cardioversion should be
performed; if the distinction cannot be made,
asynchronous electrical defibrillation should be used.
Be applicable: | Mindray BeneHeart D6 Defibrillator |
Model: | 0651-20-76722 |
Material: | Polymer materials, metal materials, ceramic materials and nanomaterials |
MOQ: | one piece |
Imported product registration standards: | YZB/GER 1059-2011 "Defibrillator" |
Application scope: | ICU, private clinic, hospital |
Working humidity: | 10%~90% |
Package: | Safe BOX packaging |
We default to express delivery by DHL. We will contact you as soon as possible when sending express delivery. If you have any questions after receiving the express delivery, please feel free to contact us at any time.