6 Panel Multiple Drug Abuse Test Kit Detecting Drug Metabolites In Human Urine

Intended Use

The Multi-drug Rapid Test Panel (Urine) is rapid chromatographic immunoassays for the qualitative and simultaneous detection up to sixteen of the following drugs in human urine. The designed cutoff concentrations for these drugs are as follows:

 

TestCut-off (ng/mL)
AMP(Amphetamine)500
MOR(Morphine)200
MET(Methamphetamine )500
MTD(Methadone)300
BZO(Oxazepam)200
FYL(Fentanyl)100
OXY (Oxycodone)100
PGB (Pregabalin)500
Adulteration ( Oxidants,Specific Gravity,pH)/

 

 

Principle

 

The Multi-drug Rapid Test Panel (Urine) is one-step immunoassay in which chemically labeled drugs (drug-protein conjugates) compete for limited antibody binding sites with drugs which may be present in urine. The test panel contains membrane strips which are pre-coated with drug-protein conjugates on the test band(s). Each strip, the drug antibody-colloidal gold conjugate pad is placed at one end of the membrane. In the absence of drug in the urine, the solution of the colored antibody-colloidal gold conjugate move along with the sample solution upward chromatographically by capillary action across the membrane to the immobilized drug-protein conjugate zone on the test band region. The colored antibody-gold conjugate then attach to the drug-protein conjugates to form visible lines as the antibody complex with the drug conjugate. Therefore, the formation of the visible precipitant in the test zone occurs when the test urine is negative for the drug.When the drug is present in the urine, the drug/metabolite antigen competes with drug-protein conjugate on the test band region for the limited antibody. When a sufficient concentration of the drug is present, it will fill the limited antibody binding sites. This will prevent attachment of the colored antibody (drug-protein conjugate)-colloidal gold conjugate to the drug-protein conjugate zone on the test band region. Therefore, absence of the color band on the test region indicates a positive result.A control band with a different antigen/antibody reaction is added to the immunochromatographic membrane strip at the control region (C) to indicate that the test has performed properly. This control line should always appear regardless of the presence of drug or metabolite. If the control line does not appear the test panel should be discarded.

 

STORAGE

 

Store as packaged in the sealed pouch either at room temperature or refrigerated (2-30°C). The test is stable through the expiration date printed on the sealed pouch. The test must remain in the sealed pouch until use. DO NOT FREEZE. Do not use beyond the expiration date.

 

LIMITATION

 

1. The 6-MAM Rapid Test Cassette provides only a qualitative, preliminary analytical result. A secondary analytical method must be used to obtain a confirmed result. Gas chromatography/mass spectrophotometry (GC/MS) is the preferred confirmatory method2.3.

2. It is possible that technical or procedural errors, as well as other interfering substances in the urine specimen may cause erroneous results.

3. Adulterants, such as bleach and/or alum, in urine specimens may produce erroneous results regardless of the analytical method used. If adulteration is suspected, the test should be repeated with another urine specimen.

4. A positive result indicates presence of the drug or its metabolites but does not indicate level of intoxication, administration route or concentration in urine.

5. A negative result may not necessarily indicate drug-free urine. Negative results can be obtained when drug is present but below the cut-off level of the test.

 

performance characteristics

 

A side-by-side comparison was conducted using the Rapid Test Cassette (Urine) and GC/MS. The following results were tabulated:

 

Specimen Correlation

 

Method  GC/MSTotal Results
ACL Rapid Test ResultsPositive Negative 
 Positive23 225
Cassette (Urine)  
 Negative2 6870
   
Total Results  25 7095
% Agreement 92.0% 97.1%95.8%

 

 

Table 2: Specimen Correlation

  GC/MS
AMP One Step Test Strip +-
+1329
-4155
 

 

Positive agreement with GC/MS: 132/(132+4) = 97%

Negative agreement with GC/MS: 155/(155+9) = 95%

Total agreement with GC/MS: (132+155)/(132+4+9+155) = 96%

 

 

 

Cross Reactivity

 

A study was conducted to determine the cross-reactivity of the test with compounds in either drug-free urine or drug positive urine containing, Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Marijuana, Methadone, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Tramadol , Methaqualone, Ketamine ,Phencyclidine, Propoxyphene or Tricyclic Antidepressants. The following compounds show no cross-reactivity when tested with the Multi-Drug Rapid Test Panel with Adulteration at a concentration of 100 µg/mL. Visual interpretations occurred at 5 minutes after specimen application. Results are presented in table below.

Table: Cross Reactivity

AcetophenetidinCortisoneZomepiracd-Pseudoephedrine
N-AcetylprocainamideCreatinineKetoprofenQuinidine
Acetylsalicylic acidDeoxycorticosteroneLabetalolQuinine
AminopyrineDextromethorphanLoperamideSalicylic acid
AmoxicillinDiclofenacMeprobamateSerotonin
AmpicillinDiflunisalMethoxyphenamineSulfamethazine
l-Ascorbic acidDigoxinMethylphenidateSulindac
ApomorphineDiphenhydramineNalidixic acidTetracycline
AspartameEthyl-p-aminobenzoateNaproxenTetrahydrocortisone,
Atropineb-EstradiolNiacinamide3-acetate
Benzilic acidEstrone-3-sulfateNifedipineTetrahydrocortisone
Benzoic acidErythromycinNorethindroneTetrahydrozoline
BilirubinFenoprofenNoscapineThiamine
d,l-BrompheniramineFurosemided,l-OctopamineThioridazine
CaffeineGentisic acidOxalic acidd,l-Tyrosine
CannabidiolHemoglobinOxolinic acidTolbutamide
Chloral hydrateHydralazineOxymetazolineTriamterene
ChloramphenicolHydrochlorothiazidePapaverineTrifluoperazine
ChlorothiazideHydrocortisonePenicillin-GTrimethoprim
d,l-Chlorpheniramineo-Hydroxyhippuric acidPerphenazined,l-Tryptophan
Chlorpromazine3-HydroxytyraminePhenelzineUric acid
Cholesterold,l-IsoproterenolPrednisoneVerapamil
ClonidineIsoxsuprined,l-Propanolol 
 

Conclusion: The compounds listed in above table show no cross-reactivity when tested at a concentration of 100 mg/ml in 5 minutes.

 

 

 

 

Hot Products

High Accuracy Fecal Occult Blood Test kit, 4mm cassette, simple collection tube, proffesional use ...
High Accuracy Tumor Mark FOB Rapid Test kit, for detecting Fecal Occult Blood in Feces, gold ...
CAT Rapid Diagnostic Test/ Cathinone Rapid Diagnostic Test High Sensitivity Cut - Off 150ng/Ml ...
CAT Rapid Diagnostic Test/ Cathinone Rapid Diagnostic Test Cassete CAT Rapid Diagnostic Test/ ...
6 Panel Multiple Drug Abuse Test Kit Detecting Drug Metabolites In Human Urine Intended Use The ...
Drug Abuse test kit 5 panel drug test, 4mm strip , human urine test, high accuracy Intended Use 5 ...