in pharmaceutical and food industries.
Principle
HTZ-008B is based on Mullen theory, the piston injects silicon oil
into the chamber covered by
plastic membrane which contacts with the test sample continuously
until the sample bulges to
break, the maximum pressure is just the result of bursting
strength.
Features
Technical indexes
1.Test range:50-2,000kPa ( 7–300psi,0.5–20.4kgf/cm²)
2.Resolution: 0.1kPa
3.Error of reading: ±1%
4.Fluctuation of Reading: ≤1%
5.Oil injection rate:(95±5) ml/min
6.Sample clamping pressure: 100~1400kPa, Adjustable
7.System leakage: <10%P max @ 1min
8.Upper clamp plate hole size: (30.5±0.05)mm
9.Lower plate hole size: (33.1±0.05)mm
10. Sample thickness: Up to 21mm.
11. Plastic membrane resistance: (30±5)kPa @ Bulge (9±0.2)mm
12.Gas supply: Air, 0.6MPa-0.8MPa (self-provide)
13. Gas consumption: 0.4m3/h
14. Dimension: 380mm ×240 mm ×530 mm
15. Power supply: AC220V 50Hz
16.Net weight: 33kg
Standard : ISO 2758, APPITA AS, 1301.403, ASTM D774, CPPA D.8, JIS
8112, SCAN P 24, TAPPI T403.
Configuration
Standard
configuration: Mainframe, silicon oil(1 bottle), plastic membrane,
power line, calibration plug.
Optional:
Micro-printer paper, DSM system.