LABNOVATION LD-100 POCT HbA1c Analyser Boronate Affinity Portable For HbA1c

Country/Region china
City & Province shenzhen
Categories Solar Chargers
InquireNow

Product Details

HbA1c Analyzer LD-100 POCT HbA1c Analyzer Boronate Affinity Portable Analyzer For HbA1c

 

Intended Use

LLD-100 POCT Analyzer is intended for the percent determination of hemoglobin A1c (HbA1c) in human whole blood which is a semi-automatic instrument used to analyze glycated hemoglobin. LD-100 is small and light, it can be used to detect of HbA1c in medical units at all levels. Fually Automatic system can help the test half work with double results.

 

Feature

  • Protable, easy to carry
  • Fast test speed, WIthin 3 minutes
  • Cv<3%, Excellent Correlation With HPLC
  • RFID For Batch Management
  • Color touch screen
  • External printer (Blue-tooth)
  • USB Port, LIS Support

Specification

ModelLD-100 Analyzer
MethodBornate Affinity
Measurement Time≤3 min
Sample TypeWhole blood
Shelf-Life of test kit9 Months
Weight 800gs

 

Reliable

IFCC calibratorNormalAbnormalNormalAbnormal
Mean HbA1c%5.49.65.39.2
CV (%)2.5%1.9%2.4%1.8%

 

Operation

1. Place cartridge in analyzer.

2. Push bead into reagents.

3. Insert sample into cartridge.

 

Performance Parameters

  • Accuracy: Is tested with the reference material as samples, and the relative deviation of the determination results of the analyzer should be within the range of ±8%.
  • Precision:Samples with a concentration of 4.0%-6.5% are detected, sample coefficient of variation (CV) of the test results is not more than 3%.
  • Linearity:In the range of 4%-16% of HbA1c, the linear correlation coefficientrof the test results should be no less than 0.9900.
  • Stability:Within 8 hours after the instrument is turned on and stable, the relative deviation of the test results is not more than ±3%.
  • Functions of the analyzer:The function of report output in English;  Abnormal alarm function.

 

Certificate

 

 

 

 

Hot Products

HbA1c POCT Analyzer Mini Analyzer For HbA1c Testing IFCC&NGSP Certificated Best Performance Intended ...
POCT Immunoassay Analyzer HbA1c Analysis For Diatebes Diagnosis Product Application This instrument ...
Automatic POCT HbA1c Analyzer For Glycosylated Hemoglobin Diaglosis Intended use This instrument is ...
Automated POCT HbA1c Analyzer For Glycated Hemoglobin Test Intended Use Automated POCT HbA1c ...
HbA1c Analyzer LD-100 POCT HbA1c Analyzer Boronate Affinity Portable Analyzer For HbA1c Intended Use ...
HbA1c Analyzer Fully Automated POCT Analyzer Detect Glycosylated Hemoglobin In Human Blood The HbA1c ...