Please email to Sandy@yuyangtest.com if you need details.

 

ISO 5659-2 NBS Smoke Density Chamber Tester

 

Introduction

The Smoke Density Test Chamber has proved to be one of the most versatile fire testing instruments since its inception in the 1980s. This tester is the most widely accepted apparatus for the measurement of smoke from burning materials and measures specific optical density under flaming and non-flaming conditions. It is also used for the extraction of toxic gas.

  • This smoke density tester is satisfied with standards (ASTM E 662, BS6401, ISO 5659, NES 711, NEPA 258, etc).
  • Using ISO 5659 conical radiant heat furnace, We can change the heating condition from 10KW/㎡to 5KW/㎡by user.

  • In the test of NES 711, there are non-flaming test and flaming test in test conditions different from ASTM E 662.

 

 

Features

  • The inner dimensions of the smoke density chamber is 914mm±3mm long, 914mm±3mm high and 610mm±3mm deep.

  • The inner surface is consist of Teflon coating, which is resistant to chemical attack and corrosion and easily cleaned.

  • Smoke Density Test chamber is equiped with full width opening door, allowing easy access for sample loading and chamber cleaning.

  • Provided with a hinged front-mounted door with an observation window and a removed opaque door cover to the window to prevent light entering into the smoke density chamber tester.

  • Chamber walls are pre-heated for easier start-up and convenient equipment operation.

  • Safety blow-out panel, easily replaceable, allows for safe operation of test method.

  • Two optical windows, each with a diameter of 75mm, mounted in the top and bottom of cabinet.

  • The light measuring system is consist of Incandescent lamp and photomultiplier tube.

  • Provided with range-extension filter in the light path, adjusting the accuracy of instrument under the any scale.

  • Conical Heater - is wound in the form of a truncated cone, rated 2600 W at 230 V with a heat output of 50 kW/m2.

  • Auto Split Shutter automatically opened to transmit radiation to the sample, Spark Igniter to ignite the sample.

  • Heat Flux Meter - for setting the irradiance level at the surface of the specimens, the range from 0-100KW/m2.

  • Furnace Heater, rated 550W at 110 V with a heat output of 25 kW/m2.

  • The movement of sample is controlled by air cylinder automatically.

  • Air Cooled calorimeter for setting the irradiance level at the surface of the specimens.

  • Provided with no pilot flame and pilot flame modes. Digital display flow rate of air and propane.

  • Gas measurement ports are provided, for optional measurements of toxic gases.

  • 19 analysis rack, 15 touch screen panel type PC for the whole control and automatic testing.

  • DAQ(Data Acquisition) Program controllable with the touch  panel PC.

  • Testing results: Light transmission,Optical density,Mass optical density(MOD),Mass loss rate,the Clear-beam correction factor etc.

 

 

 

Installation requirements of ISO 5659-2 NBS Smoke Density Chamber Tester

Electrical110V AC 60Hz / 230V AC 50Hz, 30A
Ambient TemperatureOperating 10°C to 35°C
Dimensions1560 mm (W) x 2220 mm (H) x 1060 mm(D)
Gas Suppliesmixture of air & propane

 

 

Hot Products

ISO 6722-1 Lab Single Core Wire Fire Testing Equipment with 60V/600V Introductions: Single wire / ...
Please email to Sandy@yuyangtest.com if you need details. Solar Cell Spread of Flame & Burning Brand ...
ISO 8301, EN 12667 Thermal Conductivity Tester of Heat Flow Meter 0.1 ~ 8.0 m2K/W Description: It is ...
100~450℃ Melt Flow Index Tester MFR MVR Thermoplastics ISO 1133 ASTM D1238 Description: The melt ...
BS EN136-1998 Medical Surgical Mask Flame Retardant Tester (800 ± 50) ℃ Description: The Flammabilit...
100~450℃ Melt Flow Index Tester MFR MVR Thermoplastics ISO 1133 ASTM D1238 Similarities ISO 1133 and ...