Automatic Ethylene Oxide Residue Detector Gas Chromatograph High Precision

Dongguan HOOHA Electrical Equipment Company Limited
Active Member

Contact Us

[China] country

Address: ROOM 1205, 12/F, TAI SANG BANK BUILDING, 130-132 DES VOEUX ROAD,CENTRAL, HONG KONG

Contact name:Jack young

Inquir Now

Dongguan HOOHA Electrical Equipment Company Limited

Automatic Ethylene Oxide Residue Detector Gas Chromatograph High Precision

Country/Region china
City & Province hong kong hong kong
Categories Other Chemicals
InquireNow

Product Details

Gas Chromatograph

Automatic ethylene oxide residue detector

 

 

Instrument size: 570mm * 465mm * 465mm (length * width * height)

Analysis item: Determination of EO residual content

Executive Standard GBT16886. 7-2015 Biological Evaluation of Medical Devices Part 7: Residue of Ethylene Oxide Sterilization (2)

 

1. Instrument introduction

 

Ethylene oxide (EO) is a broad-spectrum low-temperature sterilant that can kill a variety of microorganisms at room temperature, including spores, tubercle bacillus, bacteria, viruses, and fungi. At the same time, EO is very penetrating and can penetrate micropores to reach the corresponding depth inside the product, thereby greatly improving the sterilization effect. At present, most sterile medical device manufacturers generally adopt ethylene oxide sterilization.

 

2. Reference Standard

 

· GB / T 16886. 7-2015 "Medical Devices Biological Evaluation Part 7: Residue of Ethylene Oxide Sterilization"

· GB / T 14233. 1-2008 "Inspection methods for medical infusion, blood transfusion and injection equipment Part 1: Chemical analysis methods"

· GB19083-2010 "Technical requirements for medical protective masks"

 

3. Instrument features

 

· Using 320x240 dot matrix 5.7-inch large-screen LCD Chinese / English display, real-time display of temperature and operating conditions of each channel, clear and intuitive content, and truly realize man-machine dialogue;

· Power-on self-test, automatic temperature rise, wide-range self-diagnosis function, can accurately determine the fault location and alarm;

· Six independent temperature control (vaporization chamber, capillary vaporization chamber can be independently temperature controlled), eight-step program temperature rise function;

· Over-temperature protection function: any circuit exceeds the set temperature, the instrument automatically power off and alarm;

· The unique vertical heating device makes the sample vaporization more reliable, minimizes the heat radiation generated by the vaporization chamber, and ensures that the temperature deviation in the oven is extremely small;

· After intelligent fuzzy control, the door opening system can automatically track the temperature and adjust it dynamically. The damper angle can realize precise temperature control even near the room temperature;

· Equipped with packed column head sampling, glass-lined sampling, diaphragm cleaning, capillary split / splitless sampling device, and gas sampler can be installed;

· High-precision dual-stabilized gas circuit, four detectors can be installed at the same time;

 

4. Temperature control index

 

· Temperature control range: room temperature 5 ℃ ~ 400 ℃ increment 0. 1 ℃

· Temperature control accuracy: better than±0. 1 ℃

Program temperature increase: 8-step interstage constant temperature time 0 ~ 999min increment 0. 1min temperature increment 0. 1 ℃ heating rate: maximum 40 ℃/ min at 200 ℃ maximum 20 ℃/ min at 200 ℃

 

5. The main technical indicators of the detector

 

Hydrogen flame ionization detector (FID):

· Cylinder collector structure design, quartz nozzle, extremely high response

· Sensitivity / detection limit: Mtslx1012g / s (n-hexadecane / isooctane)

· Baseline noise: Rns5x1014A

· Baseline drift: Rds1x10 14A / 30min

· Linear: z

Flame photometric detector (FPD):

· Detection limit: 5x1010 g sulfur / sec or 2x1010 g sulfur / sec (calculated by H, S)

· Minimum detection amount: 0.05 ppm or 0.02 ppm (calculated as H2S) 0.1 ppm (calculated as SO, calculated)

· Baseline drift: ≤0.2mV / h;

· Relative root mean square error: ≤10%;

· Incubator temperature: higher than room temperature (8 ℃ ~ 120 ℃) ± 2%;

106 Universal Thermal Conductivity Detector (TCD):

· Adopt semi-diffusion structure

· The power supply adopts constant current control

·Sensitivity: ≥8000mVml / mg (n-hexadecane).

· Baseline noise: ≤20uV.

· Baseline drift: ≤100uV / 30min.

Nitrogen and Phosphorus Detector (NPD):

· Sensitivity: Mfs5x10 ”g / s azobenzene

· Noise: ≤0.02mV

· Drift: ≤0. 15mV / h

· Imported rubidium beads, applicable residue analysis

Miniature thermal conductivity detector:

· Sensitivity: ≥12000mVml / mg (n-hexadecane).

· Baseline noise: ≤15pV.

· Baseline drift: ≤100uV / 30min.

Zirconia detector (ZD):

· Baseline noise: ≤0.05mV

· Baseline drift: ≤0.2mV / 30min

· The detection limit for hydrogen and oxygen is 0.1ul / L

 

 

 

6. Host

 

ConfigurationParameterQuantity
Detector

 

(1) Hydrogen flame detector (FID)

(2) Structural design of cylindrical collector electrode, quartz nozzle, extremely high sensitivity / detection limit: Mts1x10-12g / s (n-hexadecane / isooctane)

(3) Baseline noise: Rns5x10-14A

(4) Line drift: Rds1x10-14A / 30min

(5) Linearity: ≥10

1 set
SamplerCapillary sampling system1 set
Chromatography column

Special column for ethylene oxide

 

1 set
WorkstationDual Channel All Chinese1 set
Spare partsSyringe, pressure reducing valve connector, sampling pad, sealing pad, gas line, etc.1 set

 

 

 

6. Chromatography work peripheral configuration

 

ConfigurationparameterQuantity

 

 

 

generator

HK-3 hydrogen generator

Hydrogen purity: 99.999%

Hydrogen flow: 0 ~ 300ml / min

Flow display: LED digital display

Working pressure: 0 ~ 0. 5MPa (set at the factory to 0.3MPa)

Power supply: ~ 220V ± 10% 50Hz

Power consumption: 130W

1 set

AK-2 air generator

Output flow: 0 ~ 2000ml / min

Output pressure: 0 ~ 0. 4MPa,

System pressure: 0.55 ~ 0. 75MPa

Voltage stability accuracy: 0. 003 MPa

Working noise: <40Dbd (A)

Power supply: ~ 220V ± 10% 50Hz

Power consumption: 150W

1 set
40L nitrogen cylinderHigh-purity nitrogen cylinder + pressure reducing valve (self-equipped)1 set
Autosampler (optional according to need)

(1) 7-inch LCD touch screen, Chinese menu input, fool-free key operation

(2) Head space bottle station: 20

(3) Three-way independent temperature control accuracy: better than ± 0.1 ℃

(4) Microcomputer control, PID self-tuning temperature control mode temperature control is more accurate (gift 200 head space bottles)

1 set
StandardEO standard solution1 set
Computers and printersBrand Computer (Lenovo)1 set

Hot Products

Mask ultraviolet sterilization furnace 1. Overview Can be used to sterilize air and objects. Widely ...
Mask tensile testing machine 1. Overview The mask tensile testing machine is suitable for the ...
Mask bacterial filtration efficiency (BFE) detector 1. Overview It is used to test the bacteria ...
Gas Chromatograph Automatic ethylene oxide residue detector Instrument size: 570mm * 465mm * 465mm ...
Forehead gun constant temperature calibration sink with black body 1. Overview The LSK-C20 constant ...
Protective clothing synthetic blood penetration tester The main purpose It is suitable for testing ...