Home Companies Tunkia Co., Ltd.

TD1880 Fluke Multifunction Calibrator High Precision Multi Product Calibrator

Tunkia Co., Ltd.

Contact Us

[China] country

Trade Verify

Contact name:

Inquir Now

Tunkia Co., Ltd.

Verified Suppliers
  • Trust
    Seal
  • Verified
    Supplier
  • Credit
    Check
  • Capability
    Assessment

TD1880 Fluke Multifunction Calibrator High Precision Multi Product Calibrator

Country/Region china
InquireNow

Product Details

TD1880 High-Precision Multi-Product Calibrator
 

Introduction

  • TD1880, a wide-range, multi-value and high-precision calibrator, can accurately output ultra-precision AC/DC voltage, current, phase adjustment and power, and also has the functions of pulse output and simulating DC resistance, thermocouple (TC), thermal resistance (RTD) and capacitance. Applied to calibrate 6 and below digits multimeters or other electrical measuring instruments.

Reference

  • JJG 124-2005 Verification Regulation of Amperemeters, Voltmeters, Wattmeters and Ohmmeters.

  • JJF 1587-2016 Calibration Specification for Multimeters

  • JJF 1284-2011 Calibration Specification of Calibrators for Electrical Meters

  • JJF 1638-2017 Calibration specification of multifunctional standard source

Features

  • Electrical Parameters Output: Voltage is up to 1100 V, current is up to 30 A at a frequency 10 Hz ~ 500 kHz.

  • Resistance Simulation: Continuously variable range, 1 ~ 1100 M, pulse output range: 1 Hz ~ 2 MHz.

  • The output current has high stability (the typical value is better than 10 ppm/h) and high accuracy (the typical value is 50 ppm/a), and its measurement accuracy is higher than the international similar products.

  • Value Adjustment Method: Fixed-point, rotary encoder, step and potentiometer (option).

  • Output Switch: The output can be turned on/off arbitrarily by one-touch operation.

  • TC and RTD simulation: Support for simulation of up to 8 types of TC and 6 types of RTD.

  • Strong load capacity of AC, can drive various types of analogue meters and digital meters, so the applicable range of measured instruments is very wide.

  • The protection of voltage and current sources ensures high reliability under complex conditions.

  • Excellent Human-Computer Interaction: Large LCD touch screen display, touch and button operation modes, easy for users to measure manually.

  • Transmitter Measurement (Option): Measure the secondary DC signal of the transmitters.

  • Proprietary Software (Option): Support the calibration of the checked meters by semi-automatic or full-automatic, data management and report export.

Specification

 

AC Voltage Output

Range Resolution Frequency (Hz) Accuracy (k = 2, ppm*RD +μV)

Max. Load

Current (mA)

90 Days

(23 ±5) °C

1 Year

(23 ±5) °C

30 mV 100 nV 10 ≤ F≤ 45 600 + 6 800 + 6 50
45 < F < 400 80 + 5 100 + 5
400 < F < 10 k 80 + 10 100 + 10
10 k < F < 20 k 160 + 10 200 + 10
20 k < F < 50 k 800 + 10 1000 + 10
50 k < F < 100 k 2800 + 12 3500 + 12
100 k < F < 500 k 6000 + 50 8000 + 50
100 mV 1 μV 10 ≤ F≤ 45 250 + 8 300 + 8 50
45 < F < 400 80 + 5 100 + 5
400 < F < 10 k 80 + 10 100 + 10
10 k < F < 20 k 80 + 10 200 + 10
20 k < F < 50 k 280 + 10 350 + 10
50 k < F < 100 k 640 + 10 800 + 10
100 k < F < 500 k 1600 + 70 2000 + 70
300 mV 1 μV 10 ≤ F≤ 45 250 + 8 300 + 8 100
45 < F < 10 k 80 + 10 100 + 10
10 k < F < 20 k 80 + 10 200 + 10
20 k < F < 50 k 280 + 10 350 + 10
50 k < F < 100 k 640 + 10 800 + 10
100 k < F < 500 k 1600 + 70 2000 + 70

 

1 V

10 μV 10 ≤ F≤ 45 250 + 50 300 + 50 100
45 < F < 400 40 + 20 50 + 20
400 < F < 10 k 80 + 20 100 + 20
10 k < F < 20 k 160 + 20 200 + 20
20 k < F < 50 k 280 + 20 350 + 20
50 k < F < 100 k 640 + 20 800 + 20
100 k < F < 500 k 2000 + 600 2400 + 600
3 V 10 μV 10 ≤ F≤ 45 250 + 50 300 + 50 100
45 < F < 400 40 + 50 50 + 50
400 < F < 2 k 64 + 50 80 + 50
2 k < F < 10 k 80 + 50 100 + 50
10 k < F < 20 k 160 + 50 200 + 50
20 k < F < 50 k 280 + 50 350 + 50
50 k < F < 100 k 640 + 50 800 + 50
100 k < F < 500 k 2000 + 600 2400 + 600
10 V 100 μV 10 ≤ F≤ 15 250 + 30000 300 + 30000 100
15 < F < 45 250 + 650 300 + 650
45 < F < 400 40 + 150 50 + 150
400 < F < 2 k 64 + 150 80 + 150
2 k < F < 10 k 80 + 150 100 + 150
10 k < F < 20 k 160 + 150 200 + 150
20 k < F < 50 k 280 + 150 350 + 150
30 V 100 μV 10 ≤ F≤ 15 250 + 80000 300 + 30000 250
15 < F < 45 250 + 650 300 + 650
45 < F < 400 40 + 500 50 + 500
400 < F < 2 k 64 + 500 80 + 500
2 k < F < 10 k 80 + 500 100 + 500
10 k < F < 20 k 160 + 500 200 + 500
20 k < F < 50 k 280 + 500 350 + 500
100 V 1 mV 15 < F < 45 250 + 2000 300 +2000 250
45 < F < 400 40 + 1000 50 +1000
400 < F < 2 k 64 + 1000 80 + 1000
2 k < F < 10 k 80 + 1000 100 + 1000
10 k < F < 20 k 160 + 1000 200 + 1000
20 k < F < 50 k 280 + 1000 350 + 1000
300 V 1 mV 15 < F < 45 250 + 6000 300 + 6000 100
45 < F < 400 50 + 5000 50 + 5000
400 < F < 2 k 80 + 5000 80 + 5000
2 k < F < 10 k 80 + 5000 100 + 5000
10 k < F < 20 k 160 + 5000 200 + 5000
20 k < F < 50 k 280 + 5000 350 + 5000
1000 V

10 mV

 

45 ≤ F≤ 400 80 + 10000 100 + 10000 30
400 < F < 2 k 96 + 10000 120 + 10000
2 k < F < 10 k 120 + 10000 150 + 10000
  • Output Range: 3 mV ~ 1100 V; Display: 7 digits;

  • Distortion: <0.5 %;<1 % @ 10 mV…1 V ranges

  • Protection: Short-circuit protection, overload protection.

 

AC Current Output

Range Resolution Frequency (Hz) Accuracy (k = 2, ppm*RD +μV)

Max. Load

Current (mA)

90 Days

(23 ±5) °C

1 Year

(23 ±5) °C

300 μA 1 nA 10 ≤ F≤ 20 400 + 0.2 500 + 0.2 30
20 < F < 45 300 + 0.1 350 + 0.1
45 < F < 400 240 + 0.1 300 + 0.1
400 < F < 2 k 320 + 0.2 400 + 0.2
2 k < F < 10 k 400 + 0.5 500 + 0.5
10 k < F < 20 k 4000 + 0.6 5000 + 0.6
20 k < F < 50 k 6400 + 0.7 8000 + 0.7
1 mA 10 nA 10 ≤ F≤ 20 400 + 0.5 500 + 0.5 30
20 < F < 45 200 + 0.1 250 + 0.1
45 < F < 400 160 + 0.1 200 + 0.1
400 < F < 2 k 160 + 0.2 200 + 0.2
2 k < F < 10 k 240 + 0.5 300 + 0.5
10 k < F < 20 k 2400 + 0.6 3000 + 0.6
20 k < F < 50 k 5600 + 0.7 7000 + 0.7
3 mA 10 nA 10 ≤ F≤ 20 400 + 1.5 500 + 1.5 30
20 < F < 45 200 + 0.1 250 + 0.1
45 < F < 400 160 + 0.1 200 + 0.1
400 < F < 2 k 160 + 0.2 200 + 0.2
2 k < F < 10 k 240 + 0.5 300 + 0.5
10 k < F < 20 k 1600 + 0.6 2000 + 0.6
20 k < F < 50 k 4800 + 0.7 6000 + 0.7
10 mA 100 nA 10 ≤ F≤ 20 400 + 5 500 + 5 30
20 < F < 45 200 + 0.5 250 + 0.5
45 < F < 400 80 + 0.5 100 + 0.5
400 < F < 2 k 160 + 1 200 + 1
2 k < F < 10 k 400 + 2 500 + 2
10 k < F < 20 k 1600 + 3 2000 + 3
20 k < F < 50 k 4800 + 4 6000 + 4
30 mA 100 nA 10 ≤ F≤ 20 400 + 15 500 + 15 30
20 < F < 45 200 + 2 250 + 2
45 < F < 400 80 + 2 100 + 2
400 < F < 2 k 160 + 3 200 + 3
2 k < F < 10 k 400 + 5 500 + 5
10 k < F < 20 k 1600 + 6 2000 + 6
20 k < F < 50 k 4800 + 7 6000 + 7
100 mA 1 μA 10 ≤ F≤ 20 400 + 50 500 + 50 30
20 < F < 45 200 + 5 250 + 5
45 < F < 400 40 + 5 50 + 5
400 < F < 2 k 80 + 10 100 + 10
2 k < F < 10 k 160 + 20 200 + 20
10 k < F < 20 k 800 + 30 1000 + 30
20 k < F < 30 k 2400 + 40 3000 + 40
30 k < F < 50 k 5000 + 40 6000 + 40
1 A 10 μA 10 ≤ F≤ 20 400 + 500 500 + 500 15
20 < F < 45 200 + 20 250 + 20
45 < F < 400 40 + 50 50 + 50
400 < F < 2 k 50 + 100 50 + 100
2 k < F < 10 k 160 + 500 200 + 500
10 k < F < 20 k 800 + 600 1000 + 600
20 k < F < 30 k 2400 + 700 3000 + 700
3 A 10 μA 10 ≤ F≤ 20 400 + 1500 500 + 1500 10
20 < F < 45 200 + 100 250 + 100
45 < F < 400 40 + 100 50 + 100
400 < F < 2 k 80 + 100 100 + 100
2 k < F < 10 k 160 + 500 200 + 500
10 A 100 μA 45 < F < 400 40 + 200 50 + 200 2.5
400 < F < 2 k 80 + 200 100 + 200
2 k < F < 10 k 160 + 500 200 + 500
30 A 100 μA 45 ≤ F≤ 400 80 + 600 100 + 600 2
400 < F < 2 k 80 + 600 100 + 600
2 k < F < 5 k 160 + 1500 200 + 1500
  • Output Range: 30 μA ~ 30 A; Display: 7 digits;

  • Distortion: <0.5 %;<1 % @ 300 μA…10 mA ranges

  • Protection: Short-circuit protection, overload protection.

 

DC Voltage Output

Range Accuracy (K = 2) (ppm*RD + μV)

Max. Load Current

(mA)

Ripple Coefficient

(%)

24 Hours 90 Days 1 Year
30 mV 15 + 2 24 + 2 30 + 2 50 < 0.5
100 mV 12 + 2 20 + 2 25 + 2 50 < 0.5
300 mV 10 + 2 16 + 2 20 + 2 100 < 0.5
1 V 10 + 2 16 + 2 20 + 2 100 < 0.5
3 V 8 + 5 12 + 5 15 + 5 100 < 0.5
10 V 8 + 10 12 + 10 15 + 10 100 < 0.5
30 V 8 + 20 12 + 20 15 + 20 250 < 0.5
100 V 10 + 100 16 + 100 20 + 100 250 < 0.5
300 V 10 + 300 16 + 300 20 + 300 100 < 0.5
1000 V 10 + 1500 16 + 1500 20 + 1500 30 < 0.5
  • Output Range: ±(3 mV ~ 1100 V); Resolution: 0.0005 %*RG; Display: 7 digits

  • Range Switching: Automatic or Manual.

 

DC Current Output

Range

Accuracy (K = 2)

(ppm*RD + ppm*RG)

Max. Load Voltage

(V)

Ripple Coefficient

(%)

24 Hours

 

90 Days

 

1 Year
30 μA 50 + 0.01 80 + 0.01 100 + 0.01 11 < 0.5
100 μA 50 + 0.02 80 + 0.02 100 + 0.02 11 < 0.5
300 μA 50 + 0.02 80 + 0.02 100 + 0.02 11 < 0.5
1 mA 25 + 0.05 40 + 0.05 50 + 0.05 11 < 0.5
3 mA 25 + 0.05 40 + 0.05 50 + 0.05 11 < 0.5
10 mA 25 + 0.2 40 + 0.2 50 + 0.2 11 < 0.5
30 mA 25 + 0.5 40 + 0.5 50 + 0.5 11 < 0.5
100 mA 25 + 1 40 + 1 50 + 1 11 < 0.5
1 A 25 + 20 40 + 20 50 + 20 4 < 0.5
3 A 25 + 50 40 + 50 50 + 50 4 < 0.5
10 A 25 + 200 40 + 200 50 + 200 4 < 0.5
30 A 30 + 600 48 + 600 60 + 600 3.5 < 0.5
  • Output Range: ±(3 μA ~ 30 A); Resolution: 0.0005 %*RG; Display: 7 digits

  • Range Switching: Automatic or Manual.

 

AC/DC Power Output

DC Power Range Combination of DC voltage range and DC current range.
Accuracy DC voltage accuracy + DC current accuracy.

AC Power

(45 Hz ~ 65 Hz)

 

Range Combination of 30 V...300 V range and 10 mA…30 A range.
Accuracy Range of 10 mA…30 mA: 0.05 %*FS
Range of 100 mA…30 A: 0.02 %*FS
Power Factor -1.0000…0.0000…+1.0000
  • Active Power (|cosφ| ≥0.5), Reactive Power (|sinφ| ≥0.5)

 

Capacitance

Range Accuracy (k = 2) (%*RD+ nF or μF) Resolution Allowable Frequency
1.1 nF ~ 3.2999 nF 0.5 + 0.04 nF 0.1 pF 10 Hz ~ 3 kHz
3.3 nF ~ 10.9999 nF 0.25 + 0.04 nF 0.1 pF 10 Hz ~ 1 kHz
11 nF ~ 32.9999 nF 0.25 + 0.4 nF 0.1 pF 10 Hz ~ 1 kHz
33 nF ~ 109.999 nF 0.25 + 0.4 nF 1 pF 10 Hz ~ 1 kHz
110 nF ~ 329.999 nF 0.25 + 0.3 nF 1 pF 10 Hz ~ 1 kHz
330 nF ~ 1.09999 μF 0.25 + 1 nF 10 pF 10 ~ 600 Hz
1.1 μF ~ 3.29999 μF 0.25 + 3 nF 10 pF 10 ~ 300 Hz
3.3 μF ~ 10.9999 μF 0.25 + 10 nF 100 pF 10 ~ 150 Hz
11 μF ~ 32.9999 μF 0.40 + 30 nF 100 pF 10 ~ 120 Hz
33 μF ~ 109.999 μF 0.45 + 100 nF 1 nF 10 ~ 80 Hz
110 μF ~ 329.999 μF 0.45 + 300 nF 1 nF 0 ~ 50 Hz
330 μF ~ 1.09999 mF 0.45 + 1 μF 10 nF 0 ~ 20 Hz
1.1 mF ~ 3.29999 mF 0.45 + 3 μF 10 nF 0 ~ 6 Hz
3.3 mF ~ 10.9999 mF 0.45 + 10 μF 100 nF 0 ~ 2 Hz
11 mF ~ 30 mF 0.75 + 30 μF 100 nF 0 ~ 0.6 Hz

 

Resistance

Range Allowable Current Accuracy (k = 2), ppm*RD +Ω
24 Hours

90 Days

 

1 Year
10 Ω 1 mA ~ 150 mA 20 + 0.01 32 + 0.01 40 + 0.01
20 Ω 1 mA ~ 150 mA 15 + 0.015 24 + 0.015 30 + 0.015
100 Ω 1 mA ~ 80 mA 12 + 0.015 20 + 0.015 25 + 0.015
300 Ω 1 mA ~ 40 mA 12 + 0.02 20 + 0.02 25 + 0.02
1 kΩ 1 mA ~ 20 mA 12 + 0.02 20 + 0.02 25 + 0.02
3 kΩ 0.1 mA ~ 6 mA 12 + 0.2 20 + 0.2 25 + 0.2
10 kΩ 0.1 mA ~ 2 mA 12 + 0.2 20 + 0.2 25 + 0.2
30 kΩ 10 μA ~ 600 μA 15 + 1 24 + 1 30 + 1
100 kΩ 10 μA ~200 μA 15 + 1 24 + 1 30 + 1
300 kΩ 1 μA ~ 60 μA 17 + 10 28 + 10 35 + 10
1 MΩ 1 μA ~ 20 μA 17 + 10 28 + 10 35 + 10
3 MΩ 250 nA ~ 6 μA 30 + 150 48 + 150 60 + 150
10 MΩ 250 nA ~ 2 μA 65 + 250 104 + 250 130 + 250
30 MΩ 25 nA ~ 600 nA 125 + 2.5 k 200 + 2.5 k 250 + 2.5 k
100 MΩ 25 nA ~ 200 nA 250 + 3 k 400 + 3 k 500 + 3 k
300 MΩ 2.5 nA ~ 60 nA 1,5 k + 0.1 M 2.4 k + 0.1 M 3 k + 0.1 M
1000 MΩ 1 nA ~ 20 nA 7.5 k + 0.5 M 12 k + 0.5 M 15 k + 0.5 M
  • Output Range of Resistance: 0 Ω ~ 1100 MΩ; Resolution: 0.0004 %*RG; Display: 7 digits

 

Thermal Resistance (RTD)

Type Temperature Range (°C) Accuracy (k = 2) (°C)
90 Days 1 Year
Pt385, 100 Ω -200 ~ 0 0.05 0.05
0 ~ 300 0.08 0.08
300 ~ 850 0.12 0.12
Pt385,200 Ω -200 ~ 250 0.04 0.04
250 ~ 630 0.10 0.15
Pt385,500 Ω -200 ~ -30 0.36 0.40
-30 ~ 630 0.10 0.11
Pt385,1000 Ω -200 ~ 0 0.027 0.03
0 ~ 300 0.054 0.060
300 ~ 600 0.063 0.070
CU50 -50 ~ 150 0.09 0.09
CU100 -50 ~ 150 0.045 0.045

 

Thermocouple (TC)

Type Temperature Range (°C) Accuracy (k = 2) (°C)
90 Days 1 Year
B 410 ~ 600 0.30 0.35
600 ~ 900 0.30 0.31
900 ~ 1800 0.26 0.30
E -200 ~ 0 0.10 0.11
0 ~ 600 0.07 0.08
600 ~ 1000 0.09 0.10
J -200 ~ -100 0.12 0.13
-100 ~ 750 0.09 0.09
750 ~ 1200 0.10 0.10
K -200 ~ -100 0.15 0.20
-100 ~ 1000 0.10 0.10
1000 ~ 1370 0.12 0.12
N -200 ~ -100 0.20 0.25
-100 ~ 400 0.12 0.12
400 ~ 1300 0.10 0.12
R -50 ~ 50 0.30 0.40
50 ~ 300 0.28 0.32
300 ~ 1000 0.22 0.23
1000 ~ 1750 0.25 0.25
S -50 ~ 50 0.30 0.40
50 ~ 300 0.31 0.34
300 ~ 1000 0.24 0.24
1000 ~ 1750 0.22 0.23
T -200 ~ -100 0.17 0.25
-100 ~ 0 0.08 0.11
0 ~ 400 0.08 0.08

 

Pulse

  • Frequency Range: 1 Hz ~ 2 MHz, TTL Level

  • Resolution: 0.001 %*RG

  • Accuracy: ±(0.002 %*RD + 20 μHz)

  • Jitter: <20 ns

 

DC Small Signal Measurement (Option)

  • Range: 1 V, 10 V, 2 mA, 20 mA

  • Measuring Range: ± (0 ~12 V);± (0 ~ 24 mA)

  • Accuracy: ±(0.006 %*RD + 0.004 %*RG)

  • Notes: Applied to measure the secondary DC signal of the transmitters.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Hot Products

TD1600 Precision AC Calibrator Introduction TD1600, a single-phase AC calibrator with wide range, ...
TD1880 High-Precision Multi-Product Calibrator Introduction TD1880, a wide-range, multi-value and ...
TD1880 High-Precision Multi-Product Calibrator Introduction TD1880, a wide-range, multi-value and ...
TD1880 High-Precision Multi-Product Calibrator Introduction TD1880, a wide-range, multi-value and ...
TD1600 Precision AC Calibrator Introduction TD1600, a single-phase AC calibrator with wide range, ...
TD1050A Calibration Device for Clamp Meter Introduction TD1050A, a calibration device for clamp ...