Full Digital Vascular Color Doppler Ultrasound Scanner Convex Arra Canning Mode
Performance requirements in color flow imaging mode:
a) Color ultrasound in the color flow imaging mode of the
performance of the probe should be consistent with the provisions
of Table 2.2.
b) The color blood flow image and the grayscale image of the
pipeline where it is located should basically coincide.
c) The direction of blood flow should be able to correctly
identify, no aliasing.
The performance of each probe in color flow imaging mode
Doppler mode | Convex array(model:TC60A) | Cavity(model:TC10A) | Line array(model:TL40A) |
Detection depth in color flow mode | ≥100mm | ≥40mm | ≥50mm |
Detection Depth in Spectral Doppler Mode | ≥100mm | ≥40mm | ≥50mm |
Blood flow velocity reading error | ≤±15% |