All reagents were returned to room temperature prior to
testing.Testing shall be performed at room temperature.
I. Sample extraction (see Figure 1)
After 400L (~10 drops) of sample buffer was added vertically to the
sample extraction tube, the sample was inserted into the tube and
rotated against the inner wall about 10 times so that the sample
was dissolved in as much solution as possible.
2. Sample along the inner wall of the extraction tube, leave as
much liquid as possible inside the tube, then remove and discard
it.
3. Cover with the dropper.
II test procedure (see Figure 2)
1. Remove the test card from the sealed bag.
2. Add 2 drops (~80L) of the treated sample extract to the sample
hole of the test card and start the timer.
3. Place the test card at room temperature for 15 minutes and read
the results.The results were not valid after 20 min.
Schematic diagram of the card result judgment:
Invalid (1) Results: the quality control line (line C) has no
reaction line, and the test is invalid, so the test is repeated.
(2) Negative result: red ribbon, quality control line (line C line)
coloring.
(3) Positive result: two red bands, test line (T line) and quality
control line (C line) are all colored.