Home Companies XI'AN BEICHENG ELECTRONICS CO.,LTD

Portable Multifunction Process Calibrator Rtd Calibration Equipment

XI'AN BEICHENG ELECTRONICS CO.,LTD
Active Member

Contact Us

[China] country

Address: JINGYUAN 7 ROAD, JINGHE INDUSTRIAL PARK, NORTH DISTRICT,XI'AN,CHINA  710200

Contact name:sales

Inquir Now

XI'AN BEICHENG ELECTRONICS CO.,LTD

Portable Multifunction Process Calibrator Rtd Calibration Equipment

Country/Region china
Categories Solar Chargers
InquireNow

Product Details

Portable Multifunction Process Calibrator Rtd Calibration Equipment

 

 

VICTOR 24 Handheld Multifunction Process Calibrator 0.02% Accuracy Two Independent Channels For Input And Output 5 Digits Display

 

Introduction

This multifunction process calibrator (the calibrator in the following) is a handheld, battery-operated instrument that measures and sources electrical and physical parameters. (See Table 1)

Table 1 Source and Measurement Function

 

Measurement

Function

Sourcing

Function

DCV

DCI

OHM

FREQ

TC

RTD

CONT.

LOOP OFF

LOOP ON

DCV

DCI

RAMP ON

×

×

×

×

×

×

×

×

RAMP OFF

OHM

FREQ

×

TC

×

×

RTD

×

×

Except the functions listed in Table 1, the calibrator has the following features as well:

  • You can operate the measurement and source function simultaneously. The LCD screen is divided into two separate districts, whose upper part displays measurement information and lower part displays source information.

  • TC measurement/source terminals and built-in lead connector of same temperature (RJ compensation with auto-reference joint point)

  • Manual step source and auto -step and sweeping –step source

  • Room temperature monitoring under any operation

  • Measurement/source temperature monitoring function

  • Measurement/source mA% display

  • Measurement wave-filter function

  • Measurement manual-holding function

  • Pressure source auto-holding function

  • Contact Us

To purchase parts, obtain operation help or address of the vendor or service center nearest to you, please call us or visit our web (see the bottom page of the Manual).

  • Standard Accessories

Make sure that the package contains all the accessories listed below. And if you find they are damaged or any of them is missing, please contact the vendor from which you purchased the product as soon as possible. Refer to the replacing part list in 15.3 in the Manual if you want to order the replacing parts.

  • Two pair of Industrial testing Lead (H000002)

  • A pair of Testing Probe (H000000)

  • A pair of Alligator clip (H010000)

  • A quick reference guide

  • A User's Manual

  • One Fuse 50mA/250V

  • One Fuse 100mA/250V

Techincal Data 

 

 

 

Project

Model
Performance
parameters

VICTOR 24

VICTOR 25

Source

DC millivolt

Range

-10 ~ 110mV

-10 ~ 110mV

Accuracy=
±(% of the set value+% of range)

0.02%+0.01%

0.02%+0.01%

Resolution

1uV

1uV

DC voltage

Range

-0.1 ~ 11V

-0.1 ~ 11V

Accuracy=
±(% of the set value+% of range)

0.02%+0.01%

0.02%+0.01%

Resolution

10uV/0.1mV

10uV/0.1mV

DC current

Range

0 ~ 22mA

0 ~ 22mA

Accuracy=
±(% of the set value+% of range)

0.02%+0.02%

0.02%+0.02%

Resolution

1uA

1uA

Sink current

Range

0 ~ 22mA

0 ~ 22mA

Accuracy=
±(% of the set value+% of range)

0.02%+0.02%

0.02%+0.02%

Resolution

1uA

1uA

External power
supply

5 ~ 28V

5 ~ 28V

Resistance

Range

0 ~40KΩ

0 ~40KΩ

Accuracy=
±(% of the set value+% of range)

0.02%+0.02%

0.02%+0.02%

Resolution

0.01Ω/0.1Ω/1Ω

0.01Ω/0.1Ω/1Ω

Thermocouple

Range

R/S/K/E/J/T/B/N

R/S/K/E/J/T/B/N

Accuracy

0.4℃

0.4℃

Resolution

0.1℃/1℃

0.1℃/1℃

RTD

Range

Pt100/Pt200/Pt500/
Pt1000/Cu10/Cu50

Pt100/Pt200/Pt500/
Pt1000/Cu10/Cu50

Accuracy

0.2℃

0.2℃

Resolution

0.1℃

0.1℃

Frequency

Range

1Hz ~110KHz

1Hz ~110KHz

Accuracy

±2words

±2words

Resolution

0.01Hz/1Hz/0.1KHz/2KHz

0.01Hz/1Hz/0.1KHz/2KHz

Pulse

Frequency

N/A

1Hz ~10KHz

Range

1 ~100000cyc

Accuracy

±2words

Resolution

1cyc

Switch

Range

N/A

1Hz ~110KHz

Accuracy

±2words

Resolution

0.01Hz/1Hz/0.1KHz/2KHz

Pressure

Range

N/A

Depends on the pressure module

Accuracy=
±(% of the set value+% of range)

Depends on the pressure module

Resolution

5 digit display

Measure

DC millivolt

Range

-5 ~ 550mV

-5 ~ 550mV

Accuracy=
±( % of Reading+ % of Range)

0.02%+0.01%

0.02%+0.01%

Resolution

1uV/10uV

1uV/10uV

DC voltage

Range

-0.5 ~ 55V

-0.5 ~ 55V

Accuracy=
±( % of Reading+ % of Range)

0.02%+0.01%

0.02%+0.01%

Resolution

0.1mV/1mV

0.1mV/1mV

DC current

Range

-5 ~ 55mA

-5 ~ 55mA

Accuracy=
±( % of Reading+ % of Range)

0.02%+0.01%

0.02%+0.01%

Resolution

1uA

1uA

Resistance

Range

0 ~5.5KΩ

0 ~5.5KΩ

Accuracy=
±( % of Reading+ % of Range)

0.05%+0.02%

0.05%+0.02%

Resolution

0.01Ω/0.1Ω

0.01Ω/0.1Ω

Thermocouple

Range

R/S/K/E/J/T/B/N

R/S/K/E/J/T/B/N

Accuracy

0.7℃

0.7℃

Resolution

0.1℃/1℃

0.1℃/1℃

RTD

Range

Pt100/Pt200/Pt500/
Pt1000/Cu10/Cu50

Pt100/Pt200/Pt500/
Pt1000/Cu10/Cu50

Accuracy

0.3℃

0.3℃

Resolution

0.1℃

0.1℃

Frequency

Range

3 Hz~50KHz

3 Hz~50KHz

Accuracy

±2words

±2words

Resolution

0.01Hz/0.1Hz/1Hz

0.01Hz/0.1Hz/1Hz

Switch

Range

N/A

CLOSE/OPEN

Threshold

200 ~300Ω

Pressure

Range

N/A

Depends on the pressure module

Accuracy=
±( % of Reading+ % of Range)

Depends on the pressure module

Resolution

5 digit display

Continuity beeper

Range

0 ~500Ω

0 ~500Ω

Accuracy

≤50Ωsound

≤50Ωsound

Resolution

0.01Ω

0.01Ω

Loop power supply

Range

24V

24V

Accuracy

±10%

±10%

The maximum current

22mA

22mA

Other functions

Backlight

Auto Power-off

Charging function

Room temperature display

Battery detection

The ramp signal

Communication function

RS232 serial communication, can realize remote control instrument by the device driver software.

RS232 serial communication, can realize remote control instrument by the device driver software.

Characteristic

Safety

EN61010-1:2001

EN61010-1:2001

Withstand voltage

The input and output AC350V/1 minutes

The input and output AC350V/1 minutes

Insulation

The input and output DC500V /50MΩ.

The input and output DC500V /50MΩ.

EMC

EN61326-1:2006

EN61326-1:2006

Authentication

CE

CE

Display

Double display on the LCD screen

Double display on the LCD screen

Power supply

4×1.5V AAA Alkaline battery

4×1.5V AAA Alkaline battery

Size

205 × 95 × 42 mm

205 × 95 × 42 mm

Weight

About 500g

About 500g

Standard Quantity Per Carton

10pcs

10pcs

Standard Carton Measurement

395mm*385mm*260mm

395mm*385mm*260mm

Standard Carton Cross Weight

12.4Kg

12.4Kg

 

Hot Products

VICTOR 85D True RMS 600V/10A/60mF Multimeter Intelligent Multi-Purpose Meters Automatically Identify ...
VICTOR 6012A Pen-Type Multimeter With 3 5/6 Digit Display True RMS Multimeter Non-Contact 600V AC/DC ...
ecer({videoId:"d9ex-DOwn_Y",videoId_local:"20210903_6131e156e835d.mp4",idContainer:"#qtGYCmr"}) ...
VICTOR VC9205 1999 3 1/2 Large LCD Display Manual Range Digital Multimeter With Frequency And ...
VICTOR 88C Manual Range Digital Multimeter 1999 Counts large LCD Display With True RMS 1000V/20A AC ...
VICTOR 79 Digital Multimeter 2 In 1 4-20mA Signal Output Multimeter Process Signal Source 40MΩ RTD ...