G08 General Purpose Gas Flow Analyzer Medical Equipment Test Tool

Country/Region china
City & Province beijing beijing
InquireNow

Product Details

Introduction
The Flow Analyzer Go8, referred to hereafter as the Analyzer, is a general-purpose gas flow analyzer. The Analyzer measures
unidirectional gas flow, airway pressure and high pressure. The Analyzer can also measure tidal volume, inspiratory peak pressure, positive
end-expiratory pressure, etc.
This manual provides a description of the Analyzer and its applications. lt should be used as a guide when using the instrument to analyze ventilators and test lungs.
Key Features
Unidirectional flow rate, airway pressure, low pressure, high pressure and oxygen concentration measurement
Flow, pressure real-time waveform display
Touch buttons control
RS232 serial port and can be used for flow calibration

Intuitive interface, easy to operate
Easy to carry

Unpacking and inspection
Use the following checklist when unpacking the Analyzer. There are no special unpacking instructions, but be careful not to damage the instrument when unpacking it.
Check the shipping carton is in good condition. lf the shipping carton is damaged, carefully continue unpacking the instrument and note any dents and scratches on the Analyzer. Save the damaged shipping carton and packing material for the carrier inspection.
lf there is no shipping damage, continue removing the Analyzer from the shipping case. Save the box and packing materials.
Perform a visual inspection to ensure that the Analyzer is intact and if there is any physical damage, such as bent or broken parts, dents, or scratches. If the Analyzer has been damaged, call the theSiriusmed Service Center immediately.
After unpacking the Analyzer, check to see if all of the standard accessories listed under the packing list have been included with the analyzer. if anything is missing, contact the Siriusmed ServiceCenter.

 

Operating Environmental Conditions
Temperature Range10 - 40 °c
Ambient Humidity0- 80 %non-condensing
Barometric Pressure520-800 mmHg
Storage Environmental Conditions 
Temperature Range-25 to 50°℃
Humidity0 -95 % non-condensing
Power Requirements
lnput Voltage Range12V DC
Power Consumption36 VA
Flow Port
Flow Range0 -200 LPM
Accuracy3% of reading or 0.1 LPM
Resolution0.01 LPM flow <5 LPM;0.1 LPM flow >5 LPM

 

Notes

Flow accuracy is specified for dry air or oxygen.

Do not use with humidified gases.

Flow accuracy is obtained by allowing the Analyzer to fully warm up.

 

Airway Pressure
Pressure Range0 - 100 cmH2O
Accuracy3 % of reading or 0.5 cmH2O
Resolution0.1 cmH2O
High Pressure
Operating Pressure Range0 - 690 kPa
Accuracy2 % of reading or 5 kPa
Resolution0.1 kPa
Barometric Pressure 
Operating Pressure Range120 - 860 mmHg
Accuracy2 % of reading or 5 mmHg
Resolution1 mmHg
Gas Temperature 
Temperature Range0 - 70 °C
Accuracy2 %FS
Resolution0.1 °C
Oxygen Concentration(Optional) 
Range0 - 100 %
Accuracy2 %FS
Resolution0.1 °C

 

 

 

 

Hot Products

Vet Anesthesia Gas Machine, screen with touch, AGSS, PEEP valve, CO2 absorber circuit *Tidal Volume ...
Unique designed variable orifice flow sensor 8.4" TFT LCD with touch screen control, easy to operate ...
Anesthetic Workstation X40, 8.4” LCD Human-computer interaction system with touch screen anesthetic ...
Vet Anesthesia Gas Machine, screen with touch, AGSS, PEEP valve, CO2 absorber circuit *Tidal Volume ...
Anesthesia Workstation Features lntegrated structure & Attractive DesignAccurate Manual Ventilation ...
Technical Specifications Mode of Ventilation VCV, PCV, SIMV-VCV, SIMV-PCV, PS with Backup(VCV,PCV), ...