IEC 60335 Dual Station Label Marking Abrasion Test Apparatus

 

Standards For Dual Station Label Marking Abrasion Test Apparatus:

 

This label marking abrasion test apparatus conforms to IEC 60335-1 clause 7.14, IEC 60065:2014 clause 5.1, IEC 60950:2013 clause 1.7.11, IEC 60669-1 clause 8.9.

 

Technical parameters For Dual Station Label Marking Abrasion Test Apparatus:

 

Test stationsTwo stations
Test stroke10-60mm, adjustable(Manually adjustable)
Speed5-60 times /min, adjustable
Counter0-9999 adjustable
WeightTwo pieces of 10g, 20g,50g, 100g, 200g, 500g
VoltageAC 220V, 50Hz
Machine size:500*300*550mm
 

Maintenance Notes For Dual Station Label Marking Abrasion Test Apparatus:

 

1 The instrument must be placed on a stable surface before use.

2 It is not allowed to move the machine during use.

3 Select the appropriate power supply voltage, should not be too high, to avoid burning the instrument.

4 If the instrument works abnormal, please contact with us for timely processing.

5 The instrument must be used in a good ventilation environment.

6 Please add lubricating oil to the mechanical transmission part in time due to the large operating load of each component of the button mechanism.

7 After each test is completed, operator should clean the machine.

8 The control box part must be wiped with a dry cloth.

 

Troubleshooting For Dual Station Label Marking Abrasion Test Apparatus:

 

1 If the POWER is turned on, the instrument is not powered on or the power indicator is not light: if the power indicator is damaged, replace the indicator; if there is no power output at the power output terminal, please check the power cord or power socket-outlet.

2 If the power is turned on, the power indicator lights, but the machine does not work when operator presses the SWITCH button. Please check whether there is a disconnection. If there is a disconnection, please connect it in time. If there is no disconnection of the relay, indicating that the counter is damaged, the counter should be replaced in time.

3 When the number of times set on the counter is finished, but the motor does not stop working, indicating that the counter is damaged.

4 If the power indicator light is on but the counter has no display, it means that the counter is damaged and the counter needs to be replaced.

 

Configuration

No.:NameUnitQuantity
1Main deviceLabel Marking Abrasion Test ApparatusSet1
2AccessoryPetroleum spirit/ alcohol abrasion test fixturesPCS2
Rubber head abrasion test fixturesPCS2
Pencil abrasion test fixturesPCS2
Cotton clothPack1
Rubber headPCS2
PencilPCS2
3DocumentUser manualCopy1
Certificate of qualityCopy1
Warranty cardCopy1

 

 

 

 

Hot Products

Freezing Load Test Package for IEC 62552 Refrigerator Energy Efficiency Test Overview: The ...
ISO 5151 60K BTU Computer Control Dual Code Air-Enthalpy Calorimeter System introduction: The air ...
IEC 62552 Refrigerator Freezer Energy Efficiency Lab Overview: This refrigerator performance lab is ...
ISO15502 Household Refrigerating Appliances Performance Lab 6 Stations Overview: This refrigerator ...
5RT Air Enthalpy Laboratory Air Conditioner Energy Efficiency Performance Lab System overview This ...
IEC61855 Figure 9 Dryer Air Volume Test System For Household And Similar Purposes Standards: QB...