Home Companies Sinuo Testing Equipment Co. , Limited

ASTMD 1525 Vicat Softening Temperature Of Determination Thermal Testing Equipment

Sinuo Testing Equipment Co. , Limited

Contact Us

[China] country

Trade Verify

Address: Room 101, 1st Floor, No. 6, Third Street, Pingshan Industrial Zone 511495, Shibi Street, Panyu District, Guangzhou, China

Contact name:Zoe Zou

Inquir Now

Sinuo Testing Equipment Co. , Limited

Verified Suppliers
  • Trust
    Seal
  • Verified
    Supplier
  • Credit
    Check
  • Capability
    Assessment

ASTMD 1525 Vicat Softening Temperature Of Determination Thermal Testing Equipment

Country/Region china
City & Province guangzhou guangdong
Categories Other Metals & Metal Products
InquireNow

Product Details

 

ASTMD 1525 Vicat Softening Temperature Of Determination Thermal Testing Equipment

 

Product Information For Vicat Softening Temperature Of Determination Thermal Testing Equipment:

 

The vicat softening temperature determination tester conforms to ISO 306-2013 clause 5.2 figure 1, ISO 2507, ISO 75, ASTMD 1525, ASTMD 648, etc.

It is mainly used for the determination of thermal deformation temperature and Vicat softening point temperature of non-metallic materials such as plastics, hard rubber, nylon, electrical insulating materials, long fiber reinforced composite materials, high-strength thermosetting extrusion materials, etc. Which is widely used in colleges and universities, scientific research units and product quality supervision and inspection units.

The instrument adopts PLC (programmable controller), touch screen display and operation, and the modular design temperature control unit, which make the performance of the instrument more stable and the operation easier and more convenient.

During the operation, operator only inputs the specimen size, the standard deflection and required weight of the test will be automatically calculated out, avoiding human calculation errors.

The test process can be completed by following the displayed prompts, with high degree of automation, low failure rate, and simple maintenance. It is an advanced instrument used for the determination of thermal deformation and vicat softening point temperature of non-metallic materials.

 

Technical Parameters For Vicat Softening Temperature Of Determination Thermal Testing Equipment:

 

1 Temperature range: room temperature -300℃.

2 Heating rate: 120±10℃/h (12℃±1℃/6min)

                         50 ± 5℃/h(5℃±0.5℃/6min)

3 Temperature error: ±1℃

   Deformation error: ±0.01mm

4 Minimum resolution of temperature: 0.1℃

Minimum resolution of deformation: 0.001mm

5 Load range: 0.864N-49.03N

    Load error: ±0.5%

6 Support span: 64-100mm continuously adjustable

7 Heating power: 3KW

8 Power supply voltage: Transformer primary AC 127V/60Hz 3.5KVA

                                       Transformer secondary AC220V/50Hz

9 Heating medium: methyl silicone oil (the flash point temperature of the medium should be higher than the maximum test temperature).

10 Cooling method: natural cooling above 150℃, water cooling or natural cooling below 150℃.

11 This machine has a horizontal host, and its dimensions: 560×620×640

12 Fuel tank capacity: 12.5 liters

 
Safety Notices For Vicat Softening Temperature Of Determination Thermal Testing Equipment:
 

1 Must be heated in the presence of heating medium, otherwise the heating tube will burst due to overheating by dry burning.

2 The test specimen should be taken out if it falls into the fuel tank. The stirring motor must be turned off when taking specimens out.

3. Before using cooling water (or air) to cool down, the water (or air) pipe must be fixed to prevent burns. After cooling down, be sure to drain the water in the cooling pipe or blow it out with high-pressure air.

4. Regularly check the liquid level position of the heating medium to ensure that the oil level reaches the specified position.

5 The instrument limits the heating up to 300°C. It should be noted that the flash point temperature of the oil used should be greater than 300°C and the viscosity of the methyl silicone oil is less than 500 centistokes.

6 When loading and unloading weights, move away the magnetic core of the displacement sensor to avoid damage to the magnetic core.

 

 

 

Hot Products

ISO 5151 60K BTU Computer Control Dual Code Air-Enthalpy Calorimeter System introduction: The air ...
IEC61855 Figure 9 Dryer Air Volume Test System For Household And Similar Purposes Standards: QB...
Dryer Air Volume Test Equipment For Measure Air Volume Or Airflow Performance Of Dryer IEC 61855 ...
IEC 61591: 2014 Range Hood Air Performance (Air Volume) Testing System For Pressure Efficiency Test ...
IEC 60879 Tower Fan Table Fans Energy Efficiency Performance Test Chamber Standards Application: IEC ...
SPWM High Frequency Pulse AC Constant Current Source Width Pulse Width Modulation Product Informatio...