CONTEC MS400 ECG Simulator Multi-Parameter Patient Signal Generator Patient Simulator

Country/Region china
City & Province shenzhen guangdong
Categories Other Consumer Electronics
InquireNow

Product Details

 

CONTEC MS400 ECG Simulator Multi-Parameter Patient Signal Generator Patient Simulator

 

Introduction:

 

MS400 Multi-parameter Simulator is a portable device, which can provide comprehensive physiological waveform to content general test and train requirement. It can check the monitor rapidly near the bed, also can test the quality warranty of ECG equipment, monitor and remote therapy equipment, At the same time, it can be used as the training of distinguishing from arrhythmia waveforms to the nurses.

 

Specification:

 

3.5-inch color TFT touch-screen, 320*240 pixels, friendly operation interface.

Display the output waveforms synchronously, which is convenient for user to compare with the test equipment.

It integrates the functions of SD socket interface, user can output the custom waveform by this device.

Built-in templates list to provide the waveform that is expressed by each

preset sequence number, which is convenient for searching.

Two operation interfaces: Chinese and English.

Operate with keystroke and touch-screen expediently and rapidly.

Built-in 3.7V/4000mAh(Rechargeable alkaline battery).

Optical external power: AC 220V, 50Hz or AC 120V, 60Hz

 

ECG/Arrhythmia

ECG

Conventional: All leads have the same voltage with the II-lead as the reference

Amplitude: 0.5,1.0,1.5 and 2.0mV

The accuracy of Amplitude: ±2%

ECG Rate:

15,20,25,30,40,60,80,100,120,140,160,180,200,220,240,260,280,300,320,340 and 350 BPM

ECG artifacts: 50 or 60 Hz, muscle, baseline drift and breathing

The elevation of the ST segment: -0.8 mV to +0.8 mV, the step is 0.1 mV

The default: 80 BPM, 1mV

Arrhythmia

Atrial: Atrial Fib(coarse), Atrial fib(fine), Atrial Flutter, Sinus Arrhythmia,

Missed Beat, Atrial

Tachycardia, Nodal Rhythm, Supraventricular Tachycardia

Ventricular: Vent Fib (coarse), Vent Fib (fine), Premature Ventricular, Contraction (PVC) 6/min

PVCs 12/min, PVCs 24/min, Frequent Multifocal, Bigeminy, Trigeminy, Paired PVCs, Run 5

PVCs, Run 11 PVCs, Vent Tachycardia, Asystole

Premature Beat: Atrial Premature Atrial Contraction (PAC), Nodal PNC, Left Ventricular Focus

PVC1, Early Left Ventricular Focus PVC1, Right Ventricular Focus PVC1, Early Right

Ventricular Focus PVC1, R on T Left Ventricular Focus, R on T Right Ventricular Focus, Multifocal PVC

Conduction Defects: First Degree Heart Block, Second Degree Heart Block, Third Degree Heart Block, Right Bundle Branch Block, Left Bundle Branch Block

ECG Performance Testing

ECG amplitude: 0.5,1.0,1.5 and 2.0mV

The accuracy of Amplitude: ±2%

Pulse: 30 and 60 BPM, the width of pulse wave is 60 ms

Square: 0.125 and 2.0 Hz

Triangle: 2.0 and 2.5 Hz

Sine: 0.5, 5, 10, 40, 50, 60 and 100 Hz

ECG Rate: 15, 20, 25, 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240,

260, 280, 300, 320, 340 and 350 BPM

The accuracy of ECG Rate: ±1%

Respiration

Respiration Rate: 0 (Apnea), 15, 20, 30, 40, 60, 80, 100, 120 and 150 R/MIIN

Respiration Impedance: 0.2, 0.5, 1.0 and 3.0 Ω

Baseline Impedance: 500,1000,1500 and 2000

Apnea: 12 s, 22 s, 32 s

ECG simulator with adjustable heart rate/12-lead output

33 kinds of Arrhythmia waveforms

Pacemaker

Two IBP channels

Variety of respiratory impedance,apnea pauses (0 BrPM) of 12, 22 and 32 seconds.

Standard ECG calibration

Export square, triangle and sine with adjustable frequency

IBP:

The default: 00/00mmHg

Dynamic IBP:

Channel 1: Artery 120/80mmHg

Radial 120/80mmHg

Left Ventricle 120/00mmHg

Right Ventricle 25/00mmHg

Channel 2: Artery 120/80mmHg

Radial 120/80mmHg

Left Ventricle 120/00mmHg

Right Atrium 15/10mmHg

Right Ventricle 25/00mmHg

Pulmonary Artery 25/10mmHg

Left Atrium 14/4mmHg

Static-Pressure Levels

Channel 1: 0, 80, 160, 240, 320 and 400mmHg

Channel 2: 0, 50, 100, 150, 200 and 250mmHg

Hot Products

Hotel Or Travel Carry sharp Twin Blades Disposable Shaving Razor with shaving cream Blades type: 2/ ...
LSXA1000 Pink Color Razor for women With Dual 3 Blades Head Model number: LSXA1000(shai 3+3) Quick ...
6 Blades Shaving Razor Pace 6 with blue lubricating strip Razor Cartridges Model number: SXA1000 ...
Male Shaving Razor 1 Handle with 2 Cartridges PACE 4 Razor Model number: FRA1000 Standard Accessory: ...
Phellolips / Clarisonic Replacement Toothbrush Head 2013 New Product oral b toothbrush heads for ...
Oral B Replacement Toothbrush Head , Elite Brush Heads 2013 New Product toothbrush heads for oral B ...