M3001-66417 Philipilip MP Series M3001A Module Non-Invasive Blood Pressure NBP Board

 

 

 

Non-invasive blood pressure measurement (NBP)

 

Description

 

The measurement server has a non-invasive blood pressure measurement for the monitoring device. It is designed for use in adult, pediatric, or neonatal patients in ICU and OR settings.

 

Measurements

 

It measures numbers that produce systolic, diastolic, and average blood pressure values. There are no waves associated with this measurement. Three different methods can be used to obtain measurement values, as follows:

• Manual - For each request, a systolic, diastolic, and average pressure measurement is taken.

• Auto: repeats the measurement of three values ​​at user-specified intervals.

 

• Statistics - Measure three values ​​immediately and repeatedly within five minutes. This method uses a faster measurement procedure, but produces less accurate readings.

 

NBP Features

This figure shows the user controls and connectors used for measurement. The parts are described in the paragraPhilipilip that follows the illustration.

 

 

 

The measurement offers adult, pediatric, and neonatal modes.

 

The following table lists the cuff inflation limits for each mode:

 

ModeFirst Inflation

Subsequent Inflations

 

Stat Mode
Adult1652515
Pediatric1302015
Neonatal1001515

 

 

The following table lists the measurement ranges for each mode

ModeSystolic

Diastolic

 

Mean
Adult30 - 27010 - 24520 - 255
Pediatric30 - 18010 - 150120 - 160
Neonatal30 - 13010 - 10020 - 120

 

Safety

The following table lists the maximum limits that ensure patient safety:

 

Mode

Maximum

Measurement Time

Maximum Time/PressureOver-pressure/Maximum
Adult180 seconds

180 seconds for

pressure > 15

mmHg

300 mmHg for > 2

seconds

 

Pediatric180 seconds

180 seconds for

pressure > 15

mmHg

 

300 mmHg for > 2

seconds

 

Neonatal90 seconds

90 seconds for pres-

sure > 5 mmHg

 

150 mmHg for > 2

seconds

 

 

 

 

NBP measurement characteristics

 

Theory of Operation

When the NBP signals are transmitted from the patient to the monitor, they will go through the stages corresponding to the logic part of the circuit, as shown in the block diagram. Circuit-related faults can usually be isolated in one of the stages

1.Acquisition: the pressure sensor receives the signal from the patient through the cuff and the cuff connects to the circuit through a single tube. The cuff is inflated, deflated and monitored by a safety system controlled by a pump, a deflation system and a microprocessor.

A. Cuff inflation - During initial cuff inflation, the pressure pump inflates the cuff to a set pressure determined by the size of the patient. The pressure pump then inflates the cuff to a cuff pressure higher than the patient's systolic pressure. Depending on the measurement method used, the swelling will occur one or more times. When the cuff pressure is greater than the systolic pressure, the artery becomes occluded; then the pressure sensor only detects the cuff pressure.

 

B.Cuff deflation - Cuff pressure is gradually and automatically released by the deflation system until the artery is only partially occluded. At this point, as the cuff pressure is gradually released, the measurement and processing of blood pressure swings begin and continue.

 

2.Detection: blood pressure swing is superimposed on cuff pressure. They are drawn from the cuff pressure through a digital bandpass filter in the microcontroller.

 

3.Measurement: As the cuff deflates, the amplitude of the oscillation as a function of the cuff pressure increases until the mean arterial pressure is reached. When the cuff pressure is less than the mean arterial pressure, the amplitude of oscillation begins to decrease. The systolic and diastolic blood pressure values ​​are derived from the oscillometric signal by extrapolation. Results can be expected to differ from the standard stethoscope method. NBP accuracy conforms to AAMI SP-10

Hot Products

M3016A Philipilip Mainstream CO2 MMS Module M3016A Measurement Server Expansion Working Principle ...
3D4-8ET Samsung Medison Ultrasound Probe for Accuvix V20. Accuvix V10. SonoAce R7 Live 3D. SonoAce ...
Mindray IMEC7 Patient Monitor With ECG, RESP, NIBP, SpO2, PR, TEMP Parameters Mindray iMEC 7 patient ...
Hospital Medical Equipment Used Patient Monitor Edan SE-2003 SE-2012 Holter System ► Brief ...
Hospital Medical Equipment Stryker Endoscopy X-6000 X6000 Xenon Light Source, 220-185-000 ►Brief ...
Refurbished GE DASH2000 Patient Monitor Standard Parameter For Sale GE Dash2000 device is designed ...