► Removing the Display Assembly
- Remove the top cover assembly
- Place the top cover assembly top down on a cloth or other soft
surface.
![](/Philipilipoto/patients-monitor/editor/20211129131558_96583.jpg)
- First release the cable lock by moving it in the direction of the
arrows, then disconnect the white recorder ribbon cable from the
recorder adapter board.
![](/Philipilipoto/patients-monitor/editor/20211129131559_94790.jpg)
- Next, disconnect the silver display ribbon cable from the recorder
adapter board after releasing the cable lock.
![](/Philipilipoto/patients-monitor/editor/20211129131559_68476.jpg)
- Turn over the top cover assembly.
- Remove the two plastic hinge pins on either side of the display
housing. Release them by turning the slotted head anti-clockwise
with a small flat-bladed screwdriver.
![](/Philipilipoto/patients-monitor/editor/20211129131559_48641.jpg)
- Squeeze the arms of the ratchet clip to remove it from the slots on
the rear of the display housing and the top cover as illustrated.
![](/Philipilipoto/patients-monitor/editor/20211129131559_74750.jpg)
- Gently pull the display ribbon cable through the cable guide, and
remove the display ssembly
![](/Philipilipoto/patients-monitor/editor/20211129131559_74989.jpg)
► Replacing the Display Assembly
Replacing the display assembly is the reverse of the removal
procedure.
- Wth the top cover assembly facing upwards, feed the display ribbon
cable through the cable guide in the top cover assembly
![](/Philipilipoto/patients-monitor/editor/20211129131600_37314.jpg)
- Carefully turn over the top cover assembly together with the
display assembly.
- Gently pull the display ribbon cable fully through the cable guide.
![](/Philipilipoto/patients-monitor/editor/20211129131600_72066.jpg)
- Reconnect the display ribbon cable to the recorder adapter board,
ensuring that the cable lock is secure after fitting.
![](/Philipilipoto/patients-monitor/editor/20211129131600_57221.jpg)
- Reconnect the white recorder ribbon cable to the recorder adapter
board, ensuring the cable lock is secure after fitting.
![](/Philipilipoto/patients-monitor/editor/20211129131600_75807.jpg)
- Refit the plastic hinge pins. Do not apply excessive force when
refitting the hinge pins. Turn the head of the hinge pin gently
with a small screwdriver. You will feel when the pin is seated
correctly, the head should be flush with the surface, and the slot
in the head of the pin should be pointing in the same plane as the
side of the display assembly housing
![](/Philipilipoto/patients-monitor/editor/20211129131600_70661.jpg)
- Refit the ratchet clip into the slots on the rear of the display
housing and the top cover, as a reversal of the removal procedure
► Specifications
Maternal Heart Rate | 1 bpm |
Toco | 1 unit |
Ultrasound | 1 bpm |
IUP | 1 mmHg |
► Adjusting the Display Angle
You can tilt the display on the FM20 and FM30 to one of five
different positions, or you can fold it completely down. The
tilt/fold mechanism works on a one-way ratchet system. You hear a
click as each of the five positions is reached. The screen can be
folded back down only after tilting the display forwards as far as
it will go.
To tilt the display from the folded position:
- Unlock the display by releasing the catch.
![](/Philipilipoto/patients-monitor/editor/20211129131600_22071.jpg)
- Lift the display forward. You will hear a click as the first
position engages. If you want to tilt the display further, lift the
display further forward until you reach the desired angle.
![](/Philipilipoto/patients-monitor/editor/20211129131600_20865.jpg)
- To fold the display, pull the display forwards as far as it will go
![](/Philipilipoto/patients-monitor/editor/20211129131601_50896.jpg)
- Then push the display all the way back until it clicks shut.
![](/Philipilipoto/patients-monitor/editor/20211129131601_16896.jpg)
If your monitor is wall-mounted, the display should be folded flat.
We seek for providing perfect service to the satisfaction of
customers. Thank you for your time! If you are interested in Philipilip Avalon FM20 FM30 fetal monitor Display assy 5 wire Part
453564435191, welcome to place your orders. Early order early shipment!