Hospital Medical Equipment Philipilip Avalon FM20 FM30 fetal monitor Display assy 5 wire Part 453564435191

 

 

Removing the Display Assembly

  1. Remove the top cover assembly
  2. Place the top cover assembly top down on a cloth or other soft surface.
  3. First release the cable lock by moving it in the direction of the arrows, then disconnect the white recorder ribbon cable from the recorder adapter board.
  4. Next, disconnect the silver display ribbon cable from the recorder adapter board after releasing the cable lock.
  5. Turn over the top cover assembly.
  6. Remove the two plastic hinge pins on either side of the display housing. Release them by turning the slotted head anti-clockwise with a small flat-bladed screwdriver.
  7. Squeeze the arms of the ratchet clip to remove it from the slots on the rear of the display housing and the top cover as illustrated.
  8. Gently pull the display ribbon cable through the cable guide, and remove the display ssembly

Replacing the Display Assembly

Replacing the display assembly is the reverse of the removal procedure.

  1. Wth the top cover assembly facing upwards, feed the display ribbon cable through the cable guide in the top cover assembly
  2. Carefully turn over the top cover assembly together with the display assembly.
  3. Gently pull the display ribbon cable fully through the cable guide.
  4. Reconnect the display ribbon cable to the recorder adapter board, ensuring that the cable lock is secure after fitting.
  5. Reconnect the white recorder ribbon cable to the recorder adapter board, ensuring the cable lock is secure after fitting.
  6. Refit the plastic hinge pins. Do not apply excessive force when refitting the hinge pins. Turn the head of the hinge pin gently with a small screwdriver. You will feel when the pin is seated correctly, the head should be flush with the surface, and the slot in the head of the pin should be pointing in the same plane as the side of the display assembly housing
  7. Refit the ratchet clip into the slots on the rear of the display housing and the top cover, as a reversal of the removal procedure

 

Specifications

Maternal Heart Rate1 bpm
Toco1 unit
Ultrasound1 bpm
IUP1 mmHg

 

 

► Adjusting the Display Angle

You can tilt the display on the FM20 and FM30 to one of five different positions, or you can fold it completely down. The tilt/fold mechanism works on a one-way ratchet system. You hear a click as each of the five positions is reached. The screen can be folded back down only after tilting the display forwards as far as it will go.

To tilt the display from the folded position:

  1. Unlock the display by releasing the catch.
  2. Lift the display forward. You will hear a click as the first position engages. If you want to tilt the display further, lift the display further forward until you reach the desired angle.
  3. To fold the display, pull the display forwards as far as it will go
  4. Then push the display all the way back until it clicks shut.

    If your monitor is wall-mounted, the display should be folded flat.

We seek for providing perfect service to the satisfaction of customers. Thank you for your time! If you are interested in Philipilip Avalon FM20 FM30 fetal monitor Display assy 5 wire Part 453564435191, welcome to place your orders. Early order early shipment!

Hot Products

Dash5000 GE Refurbished Used Patient Monitor For Hospital And Clinic Main features: ① Optimize the ...
M3016A Philipilip Mainstream CO2 MMS Module M3016A Measurement Server Expansion Working Principle ...
3D4-8ET Samsung Medison Ultrasound Probe for Accuvix V20. Accuvix V10. SonoAce R7 Live 3D. SonoAce ...
Mindray IMEC7 Patient Monitor With ECG, RESP, NIBP, SpO2, PR, TEMP Parameters Mindray iMEC 7 patient ...
Hospital Medical Equipment Used Patient Monitor Edan SE-2003 SE-2012 Holter System ► Brief ...
Hospital Medical Equipment Stryker Endoscopy X-6000 X6000 Xenon Light Source, 220-185-000 ►Brief ...