Hospital Medical Equipment Parts Mindray D3 defibrillator Rechargeable Li-ion battery Model No.LI24I001A
► Brief Introduction
The equipment is designed to operate on battery power when external
power supply is not available. The battery is charged whenever the
equipment is connected to AC mains or the DC power supply through
an external DC/AC adapter, regardless of whether or not the
equipment is currently turned on. In case of power failure, the
equipment will automatically run power from internal batteries. So
we recommend you always install a fully charged battery in the
equipment.
The equipment is configured with one smart lithium ion batteries
which are free of maintenance.
On-screen battery symbols indicate the current battery charge
status, for an example:
≤100%, but >80% of capacity
≤80%, but >60% of capacity
≤60%, but >40% of capacity
≤40%, but >20% of capacity
≤20% of capacity
Low battery and charging is required immediately
Battery is not installed
You can also check the battery's charge status by pressing the fuel
gauge button on the battery to illuminate the battery gauge. The
fuel gauge consisting of 5 LEDs, each LED represents a charge of
approximately 20% of capacity. If the battery charge is too low, a
technical alarm will be triggered and the "Low Battery" message
displayed in the Technical Alarm Area. At this moment, change the
battery or apply external power to the equipment.
NOTE
Always connect the equipment to AC mains whenever it is possible.
Always install a fully charged battery in the equipment.
After long term use, the power capacity indicated by the battery
symbol may be different from the actual capacity.
Always observe the alarm information displayed on the screen
► Battery Run time
Monitoring | ≥2.5 h | Without recording, typical ECG monitoring, LCD brightness set to 1 |
Defibrillation | ≥100 discharges | 360J discharges at intervals of one minute, without recording |
pacing | ≥2 h | 50 Ω load impedance, pacing rate: 80bpm,pacing output 60mA, without recording |
► Specifications
Battery type | 14.8V/3Ah, smart lithium ion battery, rechargeable and free of maintenance, one battery can be configured |
Charge time | Less than 2 hours to 80% and less than 3 hours to 100% with
equipment power off; Less than 3.5 hours to 80% and less than 4.5 hours to 100% with equipment power on. |
Battery fuel gauge | 5 LEDs indicating the current battery charge level |
Shutdown delay | At least 20 minutes of monitoring and six 360J discharges (after the low battery alarm occurs) |
► Charging batteries
The batteries can be charged only when they are installed in the
equipment or using an device approved by the equipment
manufacturer. With the equipment turned off and at a temperature of
25°C (77°F), a completely discharged battery charges to 80% of its
capacity in approximately 2 hours, and to 100% of its capacity in
approximately 3 hours. Batteries charge at a lower rate with the
equipment turned on.
Batteries should be charged at temperatures between 0°C (32°F) to
45°C (113°F). To optimize performance, a fully (or nearly fully)
discharged battery should be charged as soon as possible
► Low Battery Alarm
If the equipment is run on battery and the battery charge is low, a
technical alarm "Low Battery" will be triggered. In this case,
replace the battery with a fully charged on or connect the
equipment to external power supply immediately.
If the battery is almost depleted, a prompt "Battery depleted! System will shut down imminently. Connect to AC mains or replace battery" pops up. In monitoring, manual defibrillation, and pacing mode, additional alarm lights and alarm tones are provided. In this case, take appropriate actions immediately. This prompt will not disappear until the battery is replaced or the equipment is connected to the external power supply. The equipment automatically shuts down if no action is taken within a period of about 3 minutes.
NOTE
The Low Battery alarm means that the battery is beginning to weaken
and should be replaced at the first opportunity. At least 20
minutes of monitoring and six full energy shocks can be performed
when the Low Battery alarm is activated. Replace the battery or
connect the equipment to AC mains as soon as possible.